Краткое содержание страницы № 1 
                    
                         
 
 
 
®
Victoreen 875 
High Range Containment Monitor 
 
 
 
 
 
 
 
 
Operators Manual 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
February 2005 
Manual No. 877-1-1 Rev. 9 
©2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. 
All product names are trademarks of their respective companies 
 
                                                                                                                                                                                            
                    
                    Краткое содержание страницы № 2 
                    
                                                                                                                    Fluke Biomedical  Radiation Management Services    6045 Cochran Road  Cleveland, Ohio 44139  440.498.2564    120 Andrews Road  Hicksville, New York 11801  516.870.0100    www.flukebiomedical.com/rms                                                                                                                                                                                                                  
                    
                    Краткое содержание страницы № 3 
                    
                        Table of Contents    Section 1: Introduction................................................................................................ 1-1  1.1 General Description ..................................................................................... 1-1  1.2 Specifications............................................................................................... 1-1  1.3 Equipment Overview.................................................................................... 1-3  1.4 Rece
                    
                    Краткое содержание страницы № 4 
                    
                        Appendix A: Calibration and Test Procedures..............................................................A-1  A.1 Calibration Procedures ................................................................................ A-1  A.2 Test Procedures .......................................................................................... A-1    Appendix B: Cable and Pull Box Procedures................................................................B-1  B.1 Cable and Pull Box Procedures ...............
                    
                    Краткое содержание страницы № 5 
                    
                        Introduction 1 General Description   Section 1  Introduction    1.1 General Description  Containment Monitor 875      NOTE   The 875 Containment Monitor has been qualified  for use in Nuclear Safety related applications.  Qualification Test Reports, 950.301, 950.308A, and  950.310A define the parameters that have been  verified by test and are available for purchase.    Containment Monitor 875 functions as an accident monitor for reactor containments, refer to Figure 1-1. It  consists of the 877
                    
                    Краткое содержание страницы № 6 
                    
                        Victoreen 875  Operators Manual    3 Hermetic Integrity   1E10-5 cm of helium per second @ one standard  atmosphere   Chamber Fill Gas  Nitrogen/helium (2%) at atmospheric pressure  Dimensions (W x H x D) with Mounting Bracket 9.25 x 12.5 x 10.06 in   (234.95 x 317.5 x 255.52 mm)  Weight Approximately 18 lb (8.17 kg)  Temperature, Storage  40° to 160°F (4° to 71°C)  Temperature, Operating 40° to 160°F (4° to 71°C)   375°F (180°C) peak, accident, 3 hours  Relative Humidity  100% (waterproof)  60 
                    
                    Краткое содержание страницы № 7 
                    
                        Introduction 1 Specifications   Relative Humidity (Storage): 0 to 95% (non- condensation)  Relative Humidity (Operating): 0 to 90% (non- condensation)  3  Irradiation: 3.5 to 1 x 10 Rads @ 60°C integrated over  40 years life.  Mounting 876-1-55, Rack Chassis    Available Options (consult factory) 876A-100: Readout, for use with digital systems   Outputs: 0 - 100 mV   0 - 50 mV   0 - 10 mV   0 - 5 V    Figure 1-2 illustrates the typical energy response for the detector, Figure 1-3 illustrates typ
                    
                    Краткое содержание страницы № 8 
                    
                        Victoreen 875  Operators Manual    Pull Box Assembly (878-12-5)  The 878-12-5 Pull Box Assembly is designed to allow for thermal expansion of the detector cables and to  provide a service loop. Various optional pull-box configurations are available per Table 1-1. For more  information refer to Section 2, Appendix B, and Appendix C.    Table 1-1.  Pull-box Variations  Outlet, Size and  Model Inlet, Size and Location Inlet/Outlet Orientation Location  878-12 1" 270° 2, ¾", 90° 180°  878-12-M 1 1" 
                    
                    Краткое содержание страницы № 9 
                    
                        Introduction 1 Equipment Overview                                                                           Figure 1-1. High-Range Containment Monitor 875    1-5                                                                                                                                                                                                                                                                                                                                                   
                    
                    Краткое содержание страницы № 10 
                    
                        Victoreen 875  Operators Manual                                                  Figure 1-2. Typical Energy Response Curve – Detector 877-1                                                                  1-6                                                                                                                                                                                                                                                                                                    
                    
                    Краткое содержание страницы № 11 
                    
                        Introduction 1 Equipment Overview   Figure 1-3. Typical Linearity Detector 877-1                                                                        Figure 1-4. Radiation Input vs. Meter Reading (Calculated Curve)    1-7                                                                                                                                                                                                                                                                                     
                    
                    Краткое содержание страницы № 12 
                    
                        Victoreen 875  Operators Manual      Figure 1-5. High-Range Detector, Recommended Installation    1.4 Receiving Inspection  Upon receipt of the unit:  1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier and  notify Fluke Biomedical, Radiation Management Services at 440.248.9300.  2. Remove the contents from the packing material.  3. Verify that all items listed on the packing list have been received and are in good condition.      NOTE   If any o
                    
                    Краткое содержание страницы № 13 
                    
                        Introduction 1 Storage   1.5 Storage  Storage of Victoreen instruments must comply with level B storage requirements as outlined in ANSI  N45.2.2 (1972) Section 6.1.2(.2). The storage area shall comply with ANSI N45.2.2 (1972) Section 6.2  Storage Area, paragraphs 6.2.1 through 6.2.5. Housekeeping shall conform to ANSI N45.2.3 (1972).  Level B components shall be stored within a fire resistant, tear resistant, weather tight enclosure in a well- ventilated building.  Storage of Victoreen instrume
                    
                    Краткое содержание страницы № 14 
                    
                        Victoreen 875  Operators Manual    (Blank page)                                                                                                                                                                                                                                                                                                                                                                                                                                                                     
                    
                    Краткое содержание страницы № 15 
                    
                        Installation 2 Installation   Section 2  Installation    2.1 Installation  Installation of the monitoring system consists of selecting suitable mounting sites for each component of  the system, mounting each of the components, and connecting the components into the system  configuration.  Installation of this system is as follows:    NOTE   Refer to the applicable drawings in Appendix C for  further installation Instructions.  Rack Chassis  The 876-1-55 Rack Chassis is a standard 19-inch chassis
                    
                    Краткое содержание страницы № 16 
                    
                        Victoreen 875  Operators Manual          NOTE   Under potential L.O.C.A. conditions of pressure and  temperature, the cable may expand as much as 11  inches per 100 feet.    The distance from the pull box to the detector will be determined by the amount of flex hose used to seal  the detector cables. Additional information is found in the paragraph below Cable Sealing and in CABLE- 877 and 878-12-3 procedures in Appendix B.  Once the cables have been pulled and tested, the pull box cover must be
                    
                    Краткое содержание страницы № 17 
                    
                        Installation 2 Installation   done until cable is pulled and tested. (Refer to the procedure Cable-877, in Appendix B, for more  information.)        NOTE   Minimum bend radius of the 878-1-9 cable is four  inches. When bending conduit or flex hose, take  this into consideration. Typical bend radius for P/N  878-12-30 is ¾ inch; stainless steel flex house is 12  inches.    2.2 Cable and Wiring Installation  Detector Cable Inside Containment  Detector cables used inside containment are to be inst
                    
                    Краткое содержание страницы № 18 
                    
                        Victoreen 875  Operators Manual    Thickness  0.015 inch (0.38 mm) nominal  Outer Diameter 0.217 inch (5.51 mm) nominal  Impedance  75 ohms nominal  Capacitance 22 pf/ft nominal    Ancillary Wiring    Ancillary wiring for computer, recorder, power and alarm contacts are installed according to drawing GEL  875-1 and Table 2-1 which indicates connector pin designations.  Table 2-1.  Installation    Computer & Recorder      Function   J3     Ground   D     Ground   F     Computer   E     Recorder  
                    
                    Краткое содержание страницы № 19 
                    
                        Operation 3 Operation   Section 3  Operation    3.1 Operation  Once installation is completed, operation is fully automatic. The 876A-1 Readout Module continuously  indicates the level of radioactivity measured at the detector site. When the radiation level exceeds an  alarm set point, an alarm is actuated.  The following steps explain how to operate the 875 monitor:  1. Turn the function switch to the TEST position, and press and release the Electronic Check Source  (ECS) push button. About fou
                    
                    Краткое содержание страницы № 20 
                    
                        Victoreen 875  Operators Manual    alarm level. Return the module to its proper position in the rack. Tighten the rear holding screws.  Tighten the knurled holding knob on the front. Return the module to service.  During normal operation, the radiation field is usually less than 1 R/h (the lower limit of detection of the  high range detector) and the analog meter display will be at the low end of the meter scale.