9105 / 9107
Dry-well Calibrator
User’s Guide
Rev. 2A2801
Краткое содержание страницы № 2
© Copyright,1991–2002 Hart Scientific, Inc. 799 E. Utah Valley Drive American Fork, Utah 84003-9775 Telephone: (801) 763-1600 Fax: (801) 763-1010 Internet: http://www.hartscientific.com E-mail: support@hartscientific.com Rev.2A2801
Краткое содержание страницы № 3
Table of Contents 1 Before You Start ............................1 1.1 Symbols Used .......................................1 1.2 Safety Information.....................................2 1.2.1 Warnings..................................................2 1.2.2 Cautions ..................................................4 1.3 Hart Scientific Authorized Service Centers .........................5 2 Introduction ..............................7 3 Specifications and Environmental Conditions...............
Краткое содержание страницы № 4
8 Controller Operation..........................23 8.1 Well Temperature.....................................23 8.2 Reset Cut-out.......................................23 8.3 Temperature Set-point ..................................25 8.3.1 Programmable Set-points.........................................25 8.3.2 Set-point Value..............................................26 8.3.3 Temperature Scale Units .........................................26 8.4 Scan .........................................
Краткое содержание страницы № 5
8.11.5.3 SCO................................................42 9 Digital Communication Interface ....................43 9.1 Serial Communications..................................43 9.1.1 Wiring ..................................................43 9.1.2 Setup...................................................43 9.1.2.1 BAUD Rate ............................................44 9.1.2.2 Sample Period...........................................44 9.1.2.3 Duplex Mode ...................................
Краткое содержание страницы № 6
Figures Figure 1 Top View............................................3 Figure 2 Rear Panel..........................................15 Figure 3 Front Panel..........................................16 Figure 4 Well and Insert........................................17 Figure 5 Well Insulator Top View....................................19 Figure 6 Controller Function Flowchart.................................24 Figure 7 Switch Test Data .......................................31 Figure 8 Well temper
Краткое содержание страницы № 7
Tables Table 1 International Electrical Symbols ................................1 Table 2 Communications Commands .................................47 Table 2 Communications Commands Continued............................48 Table 2 Communications Commands Continued............................49 v
Краткое содержание страницы № 8
1 Before You Start 1 Before You Start 1.1 Symbols Used Table 1 lists the symbols used on the instrument or in this manual and the meaning of each symbol. Table 1 International Electrical Symbols Symbol Description AC (Alternating Current) AC-DC Battery Complies with European Union directives DC Double Insulated Electric Shock Fuse PE Ground Hot Surface (Burn Hazard) Read the User’s Manual (Important Information) Off On 9105/9107 1
Краткое содержание страницы № 9
1 Before You Start Symbol Description Canadian Standards Association OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 refers to the CAT II level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATE- GORY II is energy-consuming equipment to be supplied from the fixed installation. Exam- ples include household, office, and laboratory appliances. C-TIC Australian EMC 1.2 Safety Information Use the instrument only as specified in this manual. Otherw
Краткое содержание страницы № 10
1 Before You Start 3 1 WARNING: THIS AREA IS HOT 2 Set Down Up Exit 9105 Figure 1 Top View 9105/9107 3
Краткое содержание страницы № 11
1 Before You Start • DO NOT REMOVE INSERTS AT HIGH TEMPERATURES. Inserts will be the same temperature as the display temperature. Use extreme care when removing hot inserts. • DO NOT operate this unit without a properly grounded, properly polar- ized power cord. • DO NOT connect this unit to a non-grounded, non-polarized outlet. • HIGH VOLTAGE is used in the operation of this equipment. SEVERE INJURY OR DEATH may result if personnel fail to observe safety pre- cautions. Before working inside the
Краткое содержание страницы № 12
1 Before You Start • Components and heater lifetime can be shortened by continuous high temperature operation. • Most probes have handle temperature limits. Be sure that the probe handle temperature limit is not exceeded in the air above the unit. • (9107 only) Always use the well insulator, see Section 5.4. • Allow for test probe expansion inside the well as the dry-well heats. • DO NOT use fluids to clean out the well. • Never introduce any foreign material into the probe hole of the insert. F
Краткое содержание страницы № 13
1 Before You Start Telefax: +31-402-675321 E-mail: ServiceDesk@fluke.nl Fluke Int'l Corporation Service Center - Instrimpex Room 2301 Sciteck Tower 22 Jianguomenwai Dajie Chao Yang District Beijing 100004, PRC CHINA Phone: +86-10-6-512-3436 Telefax: +86-10-6-512-3437 E-mail: xingye.han@fluke.com.cn Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Office Service Center 83 Clemenceau Avenue #15-15/06 Ue Square 239920 SINGAPORE Phone: +65-737-2922 Telefax: +65-737-5155 E-mail: antng@singa.fluke.
Краткое содержание страницы № 14
2 Introduction 2 Introduction The Hart Scientific Model 9105/9107 dry-well calibrator may be used as a por - table instrument for calibration of temperature probes. The 9105 operates over the range of –25°C and 140°C. The 9107 operates over the range of –45°C to 140°C. These dry-well calibrators feature an interchangeable aluminum probe sleeve along with 4 fixed wells. The temperature is accurately controlled by Hart's precision temperature controller. The calibrator controller uses a precision
Краткое содержание страницы № 15
3 Specifications and Environmental Conditions 3 Specifications and Environmental Conditions 3.1 Specifications The following table lists the specifications for this instrument. Accuracy speci- fications are applicable for a one-year calibration interval. In line with normal prudent metrology practices, Hart Scientific recommends a short-cycle interval of six months for new units during the first year. 9105 9107 Power 115 VAC (±10%), 50/60 Hz, 230 VAC (±10%), 350 watts Ambient Temperature 5–50°C
Краткое содержание страницы № 16
3 Specifications and Environmental Conditions • temperature range 5 - 50°C (41 - 122°F) • ambient relative humidity 15 - 50% • pressure - 75kPa - 106kPa • mains voltage within ± 10% of nominal • vibrations in the calibration environment should be minimized • altitude less than 2,000 meters 3.3 Warranty Hart Scientific, Inc. (Hart) warrants this product to be free from defects in ma- terial and workmanship under normal use and service for a period as stated in our current product catalog from the
Краткое содержание страницы № 17
4 Safety Guidelines 4 Safety Guidelines • Operate the instrument in room temperatures between 5–50°C (41–122°F). Allow sufficient air circulation by leaving at least 6 inches of space between the instrument and nearby objects. DO NOT place under a cabinet or other structure. Allow for overhead clearance. • The dry-well is a precision instrument. Although it has been designed for optimum durability and trouble free operation, it must be handled with care. Always carry the unit in an upright posit
Краткое содержание страницы № 18
4 Safety Guidelines the instrument. Wait until the power has stabilized before re-energizing the instrument. 12 Hart Scientific
Краткое содержание страницы № 19
5 Quick Start 5 Quick Start 5.1 Unpacking Unpack the dry-well carefully and inspect it for any damage that may have oc - curred during shipment. If there is shipping damage, notify the carrier immediately. Verify that the following components are present: • 9105 or 9107 Dry-well • 2173 Insert, 1/4” Aluminum • Power Cord • Manual • Two Well Insulators (9107 only) • Tongs (insert removal tool) 5.2 Set-up Place the calibrator on a flat surface with at least 6 inches of free space around the instrum
Краткое содержание страницы № 20
5 Quick Start “POWER” switch. The dry-well will turn on and begin to heat to the previously programmed temperature set-point. The front panel LED display will indicate the actual dry-well temperature. The 9107 is field switchable between 115 V and 230 V. Refer to Section 7.2, Switching to 230 V Operation, for information on switching the voltage. 5.4 Setting the Temperature Section 8.3 explains in detail how to set the temperature set-point on the cali- brator using the front panel keys. The pro