Краткое содержание страницы № 1
Operator’s Manual
™
Perfecto Series
2
®
Oxygen Concentrators without SensO ,
2
Oxygen Concentrators with SensO
2
®
HomeFill Compatible
Model IRC5PO2AW Model IRC5PAW
Dealer: This manual MUST be given to the end
user.
User: BEFORE using this product, read this
manual and save for future reference.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
0120
Краткое содержание страницы № 2
WARNING WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment ‐ otherwise,
Краткое содержание страницы № 3
TABLE OF CONTENTS ACCESSORIES ..................................................2 SPECIAL NOTES ...............................................4 LABEL LOCATION ............................................6 SECTION 1—GENERAL GUIDELINES .....................7 Radio Frequency Interference.......................................................... 9 SECTION 2—FEATURES ......................................10 SECTION 3—HANDLING ....................................11 Unpacking..........................
Краткое содержание страницы № 4
SPECIAL NOTES SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage. Refer to the following table for definitions of the signal words. SIGNAL WORD MEANING Danger indicates an imminently hazardous situation which, if not avoided, will result in death or serious DANGER injury. Warning indicates a potentially hazardous situation which, if not avoided, could result in death or WARNING serious injury
Краткое содержание страницы № 5
SPECIAL NOTES CAUTION “Caution: Statutory law can restrict this device to sale by or on order of a physician, or any other practitioner licensed by the law of the governmental agency in which he/she practices to use or order the use of this device.” Invacare recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition or mechanical failure. Consult your physician or equipment provider for the type of reserve system required. This equipment is to be
Краткое содержание страницы № 6
LABEL LOCATION LABEL LOCATION Perfecto ™ 2 FOR OPERATING INSTRUCTIONS or ALARMS (See Operator’s Manual) DANGER - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and away from areas where oxygen is being delivered. Textiles, oil and other combustiles are easily ignited and burn with great intensity in oxygen enriched air. DANGER - ELECTRIC SHOCK DO NOT remove cover. Refer servicing to qualified service personnel. Flowrates below 0.5 L/min w
Краткое содержание страницы № 7
SECTION 1—GENERAL GUIDELINES SECTION 1—GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Perfecto concentrator these instructions 2 MUST be followed. WARNING This section contains important information for the safe operation and use of this product. DANGER Risk of electric shock. DO NOT disassemble. Refer servicing to qualified service personnel. No user serviceable parts. TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE OR INJURY TO PERSONS. A
Краткое содержание страницы № 8
SECTION 1—GENERAL GUIDELINES Operating Information For optimum performance, Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time. Shorter periods of operation may reduce maximum product life. Keep the oxygen tubing, cord, and unit out from under such items as blankets, bed coverings, chair cushions, clothing and away from heated or hot surfaces, including space heaters, stoves and similar electrical appliances. DO NOT move or relocate concent
Краткое содержание страницы № 9
SECTION 1—GENERAL GUIDELINES Radio Frequency Interference This equipment has been tested and found to comply with EMC limits specified by IEC/EN 60601‐1‐2. These limits are designed to provide a reasonable protection against electromagnetic interference in a typical medical installation. Other devices may experience interference from even the low levels of electromagnetic emissions permitted by the above standards. To determine if the emissions from the Perfecto is 2 causing the interfere
Краткое содержание страницы № 10
SECTION 2—FEATURES SECTION 2—FEATURES Oxygen Outlet Oxygen Purity Indicator Lights / Fault and Power Indicator Lights Flowmeter Circuit Breaker Elapsed Time Power Switch Meter *NOTE: This outlet fitting is to be used only for filling oxygen cylinders with the REAR VIEW HomeFill home oxygen compressor. The outlet fitting does not affect concentrator performance. Cabinet Refer to the HomeFill Filter ownerʹs manual, part number 1116444, for connection and operating instructions. When
Краткое содержание страницы № 11
SECTION 3—HANDLING SECTION 3—HANDLING The concentrator should ALWAYS be kept in the upright position to prevent cabinet damage while being transported. If the concentrator is to be reshipped by common carrier, additional cartons are available from Invacare. Unpacking NOTE: For this procedure, refer to FIGURE 3.1. 1. Check for any obvious damage to the carton or its contents. If damage is evident, notify the carrier, or your local dealer. 2. Remove all loose packing from the carton. 3. Ca
Краткое содержание страницы № 12
SECTION 4—TECHNICAL DESCRIPTION SECTION 4—TECHNICAL DESCRIPTION The Invacare Perfecto concentrator is used by patients with 2 respiratory disorders who require supplemental oxygen. The device is not intended to sustain or support life. The oxygen concentration level of the output gas ranges from 87% to 95.6%. The oxygen is delivered to the patient through the use of a nasal cannula. The Invacare Perfecto concentrator uses a molecular sieve 2 and pressure swing adsorption methodology to
Краткое содержание страницы № 13
SECTION 5—TYPICAL PRODUCT PARAMETERS SECTION 5—TYPICAL PRODUCT PARAMETERS Alternating Current Type BF equipment Unit running Unit not running Attention - Consider Accompanying Documents DO NOT smoke Class II, Double Insulated Not Protected IPX0 Electrical Requirements: 230 VAC ± 10% (253 VAC/207 VAC), 50 Hz Rated Current Input: 1.4 A Sound Level: 39.5 dBA Average Altitude: Up to 1828 meters (6,000 ft) above sea level without degradation of concentration levels. Not recommended for use above
Краткое содержание страницы № 14
SECTION 5—TYPICAL PRODUCT PARAMETERS Flow Range: 0.5 to 5 L/min (maximum). For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter Accessory (IRCPF16AW). Low Flow Alarm: 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected). Average Power 300 W Consumption: 280 W @ 3 L/min Pressure Relief Mechanism 241 kPa ± 24.1 kPa (35 psi ± 3.5 psi) Operational at: Change in maximum 0.7 L/min recommended flow when back pressure of 7kPa is applied:
Краткое содержание страницы № 15
SECTION 5—TYPICAL PRODUCT PARAMETERS No AP/APG Not suitable for use in the presence of a flammable anaesthetic mixture. Standards and Regulatory IRC5PO2AW/IRC5PAW Listing: IEC/EN61000-3-2 IEC/EN61000-3-3, IEC/EN 60601-1, A1, A2 IEC/EN 60601-1-2 IRC5PO2AW Models Only ISO8359 MDD 93/42/EEC, Annex I and IX CE marked model IRC5PO2AW Electrical: No extension cords. Placement: No closer than 7.5 cm (3 in) from any wall, furniture, draperies, or furniture to assure sufficient air flow. Avoid deep p
Краткое содержание страницы № 16
SECTION 5—TYPICAL PRODUCT PARAMETERS IRC5PO2AW INDICATORS LABEL O PURITY INDICATOR LIGHTS 2 SYMBOL (LED) SYSTEM OKAY GREEN Indicator Light O over 85% (± 2%) O 2 2 O Between 73% (± 3%) YELLOW Indicator light 2 A. YELLOW Solid to 85% (±2%) B. YELLOW Flashing Sensor Failure Call a qualified technician. SYSTEM FAILURE RED Indicator Light O Below 73% (±3%) Continuous Audible Alarm 2 Sieve-GARD Compressor Shutdown IRC5PAW INDICATORS LABEL STATUS INDICATOR LIGHTS SYMBOL (LED) SYSTEM OKAY GREEN Ind
Краткое содержание страницы № 17
SECTION 6—OPERATING INSTRUCTIONS SECTION 6—OPERATING INSTRUCTIONS Introduction Your oxygen concentrator is intended for individual use. It is an electronically operated device that separates oxygen from room air. It provides high concentration of oxygen directly to you through a nasal cannula. Clinical studies have documented that oxygen concentrators are therapeutically equivalent to other types of oxygen delivery systems. Your provider will show you how to use your oxygen concentrator.
Краткое содержание страницы № 18
SECTION 6—OPERATING INSTRUCTIONS Your oxygen concentrator will perform best when operated under the conditions outlined in the Typical Product Parameters on page13. Usage in environments other than those described may result in the need for increased equipment maintenance. The air intake of the unit should be located in a well ventilated area to avoid airborne pollutants and/or fumes. Set Up 1. Plug in power cord to an electrical outlet. 2. Connect Humidifier (if so prescribed). WARNING
Краткое содержание страницы № 19
SECTION 6—OPERATING INSTRUCTIONS 3. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door off (FIGURE 6.2). 4. Pull up and remove the humidifier bottle adapter (FIGURE 6.2). 5. Replace the filter access door and install the air scoop onto the filter access door by inserting the six tabs on the air scoop into the six slots on the sides of the filter access door. Humidifier Bottle Insert Flat Adapter Head Screwdriv
Краткое содержание страницы № 20
SECTION 6—OPERATING INSTRUCTIONS 10.After assembly, ensure that oxygen is flowing through the cannula. DETAIL “B” DETAIL “A” Oxygen Humidifier Outlet Oxygen Bottle Connector Tubing Adapter Humidifier Humidifier Bottle Bottle Outlet Humidifier Bottle Humidifier Compartment FIGURE 6.3 Humidifier Compartment Power Switch NOTE: For this procedure, refer to On/Off FIGURE 6.4. (I/O) 1. Press power switch to On Power position. All the panel Switch lights and the audible alarm will com