Краткое содержание страницы № 1
Manual Supplement
Manual Title: Impulse 6000D/7000DP Users Supplement Issue: 3
Print Date: August 2007 Issue Date: 3/09
Revision/Date: 1, 1/08 Page Count: 10
This supplement contains information necessary to ensure the accuracy of the
above manual. This manual is distributed as an electronic manual on the
following CD-ROM:
CD Title: 6000D/7000DP
CD Rev. & Date: 1, 2/2008
CD PN: 3028681
© 2008-2009 Fluke Corporation. All rights reserved. Printed in the U.
Краткое содержание страницы № 2
Impulse 6000D/7000DP Users Manual Supplement Change #1, 46470, 46697, 47714, 48007, 49680 Replace pages 31 through 35, with the following: Defibrillator Analyzer Specifications Energy Output Measurement Compatible Defibrillator Waveshapes ................ Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic Note AC Pulsed Biphasic waveform has not been approved in the United States. Autoranged Measurement.................................... 0.1 to 600 J Accuracy 0.1 to 3
Краткое содержание страницы № 3
Manual Supplement Impulse 6000D/7000DP Users Accuracy............................................................. ±0.1 ms Voltage Range ................................................................. 20 to 5000 V Accuracy............................................................. ±(1 % of reading + 2 V) Current Range ................................................................. 0.4 to 100.0 A Accuracy............................................................. ±(1 % of reading
Краткое содержание страницы № 4
Impulse 6000D/7000DP Users Manual Supplement Scope Output Autorange ........................................................... 2000:1, 400:1 and 80:1: dependant on the range Waveform Playback Output ............................................................. BNC Output impedance........................................... 50 Ω Amplitude Accuracy ........................................ ±5 % Charge Time Measurement Range ..............................................................
Краткое содержание страницы № 5
Manual Supplement Impulse 6000D/7000DP Users Ventricular Fibrillation ......................................... Coarse and fine Monomorphic Ventricular Tachycardia ............... 120 to 300 (by 5) BPM Polymorphic Ventricular Tachycardia ................. 5 types Asystole .............................................................. Flat line ECG Waves ECG General Lead configuration .............................................. 12-lead simulation. RA, LL, LA, RL, V1-6 with independent
Краткое содержание страницы № 6
Impulse 6000D/7000DP Users Manual Supplement Normal Sinus waves Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 70 % 100 % 30 % 24 % 48 % 100 % 120 % 112 % 80 % Lead I reference Performance waves and R wave detection Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 100 % 150 % 50 % 100 % 100 % 100 % 100 % 100 % 100 % Normal Sinus waves Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 100 % 150 % 50 % 24 % 48 % 100 % 120 % 112 % 80 % ECG Normal Sinus Rates.......................................
Краткое содержание страницы № 7
Manual Supplement Impulse 6000D/7000DP Users Sine waves ......................................................... 0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz Pulse................................................................... 30 and 60 BPM, 60 ms pulse width R-Wave Detection Waveform ........................................................... Haver-triangle Amplitude............................................................ 0.05 to 0.45 (by 0.05) mV 0.5 to 5.0 (by 0.5) m
Краткое содержание страницы № 8
Impulse 6000D/7000DP Users Manual Supplement Accuracy ......................................................... ±(10 % of setting + 0.2 mV) Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent) Lead II reference Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 67 % 100 % 33 % 67 % 67 % 67 % 67 % 67 % 67 % Lead I reference Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 100 % 150 % 50 % 100 % 100 % 100 % 100 % 100 % 100 % Arrhyth
Краткое содержание страницы № 9
Manual Supplement Impulse 6000D/7000DP Users Change #2 On page 12, prior to Analyzing Pacemakers (7000DP only), add the following section: SCOPE OUTPUT Each time a defibrillator is fired into the Analyzer during an active defibrillator test, a delayed representation of the defibrillator pulse is sent to the Scope Output jack on the rear panel of the Analyzer. The Scope Output is an isolated signal and should not present any problems when connected to the input of an oscilloscope. To as
Краткое содержание страницы № 10
Impulse 6000D/7000DP Users Manual Supplement [1] Pulse +4 V, 1 ms pulse Start of defibrillator pulse playback playback [1] Pulse playback includes 0.4 ms before the pulse was detected and therefore displays any early transitional signals. Use the synchronizing marker pulses to display specific information from the Scope Output signal. To see when the defibrillator fires, set the oscilloscope trigger level to +2 V. To get the ranging information, set the oscilloscope to trigger on th
Краткое содержание страницы № 11
Manual Supplement Impulse 6000D/7000DP Users Change #3, 48038 On page 38, under Pacemaker Input, change: From: Accuracy …………..±1 % non-inductive (<2 μH) To: Accuracy……………±2 % non-inductive (<2 μH) 10 8/08