Краткое содержание страницы № 1
User’s and Service Guide
Agilent Technologies 85036B/E
75Ω Type-N Calibration Kits
This manual applies directly to Agilent 85036B calibration kits with serial
number prefix 3514A, and Agilent 85036E calibration kits with serial number
prefix 3142A. The calibration devices in this kit are individually serialized.
Record the device serial numbers in the table provided in this manual. (See
“Recording the Device Serial Numbers” in Chapter 1.)
Manufacturing Part Number: 85036-90016
Printed in U
Краткое содержание страницы № 2
Documentation Warranty THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR
Краткое содержание страницы № 3
Contents 1. General Information Calibration Kit Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Calibration Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Equipme
Краткое содержание страницы № 4
Contents 6. Replaceable Parts Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2 A. Standard Definitions Standard Class Assignments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2 Blank Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Краткое содержание страницы № 5
1 General Information 1-1
Краткое содержание страницы № 6
General Information Calibration Kit Overview Calibration Kit Overview The Agilent 85036B and 85036E type-N calibration kits are used to calibrate Agilent network analyzers up to 3 GHz for measurements of components with 75Ω type-N connectors. Kit Contents The 85036B calibration kit contains the following: • one male and one female 75Ω type-N open termination • one male and one female 75Ω type-N short termination • one male and one female 75Ω type-N load • three 75Ω type-N to 75Ω type-N adapter
Краткое содержание страницы № 7
General Information Calibration Kit Overview Adapters Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance. The adapters utilize a dual-beaded connector structure to ensure stable, repeatable connections. The beads are designed to minimize return loss and are separated far enough so that interaction between the beads is minimized. The adapters are designed so that their nominal electrical lengths are the same, which allows t
Краткое содержание страницы № 8
General Information Incoming Inspection Incoming Inspection Verify that the shipment is complete by referring to Figure 1-1 on page 1-5. Check for damage. The foam-lined storage case provides protection during shipping. If the case or any device appears damaged, or if the shipment is incomplete, contact Agilent. See Table 5-1 on page 5-3. Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company. When you
Краткое содержание страницы № 9
General Information Incoming Inspection Figure 1-1 Calibration Kit Contents 85036B/E 1-5
Краткое содержание страницы № 10
General Information Recording the Device Serial Numbers Recording the Device Serial Numbers In addition to the kit serial number, the devices in this kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in Table 1-1 for the 85036B and Table 1-2 for the 85036E. Recording the serial numbers will prevent confusing the devices in this kit with similar devices in other kits. Table 1-1 Serial Number Record for 85036B Device Serial Nu
Краткое содержание страницы № 11
General Information Calibration Kits Documented in this Manual Calibration Kits Documented in this Manual This manual applies to any 85036B or 85036E calibration kit whose serial prefix is listed on the title page. If your calibration kit has a different serial number prefix than the one listed on the title page, refer to “Calibration Kit History” below for information on how this manual applies. Calibration Kit History This section describes calibration kits with serial number prefixes diffe
Краткое содержание страницы № 12
General Information Clarifying the Sex of a Connector Clarifying the Sex of a Connector In this manual, the sex of calibration devices and adapters are referred to in terms of their connector interface. For example, a male open has a male connector. However, during a measurement calibration, the network analyzer softkey menus label a type-N calibration device with reference to the sex of the analyzer’s test port connector—not the calibration device connector. For example, the label SHORT(F) o
Краткое содержание страницы № 13
2 Specifications 2-1
Краткое содержание страницы № 14
Specifications Environmental Requirements Environmental Requirements Table 2-1 Environmental Requirements Parameter Limits a +18 °C to +28 °C (+64 °F to +82 °F) Operating temperature b ±1 °C of measurement calibration temperature Error-corrected temperature range Storage temperature - 40 °C to +75 °C (- 40 °F to +167 °F) Altitude Operation < 4,500 meters (≈15,000 feet) Storage < 15,000 meters (≈50,000 feet) Relative humidity Always non-condensing Operation 0 to 80% (26 °C maximum dry bulb)
Краткое содержание страницы № 15
Specifications Mechanical Specifications Mechanical Specifications The mechanical specifications in Table 2-2 apply to the devices in the 85036B and the 85036E 75 Ω type-N calibration kits. Table 2-2 Mechanical Specifications Device Specification Type-N Male Open Inside diameter of outer conductor: 6.985 to 7.010 Type-N Male Short Inside diameter of outer conductor: 7 ±0.015 mm Distance from reference plane to shorting plane: 5.29 ±0.013 mm Supplemental Mechanical Characteristics Supplemental c
Краткое содержание страницы № 16
Specifications Electrical Specifications Electrical Specifications The electrical specifications in Table 2-4 apply to the devices in your calibration kit when connected with an Agilent precision interface. Table 2-4 Electrical Specifications Device Specification Frequency (GHz) a Return loss ≥ 46 dB (ρ ≤ 0.00501) DC to ≤ 2 Loads Return loss ≥ 40 dB (ρ ≤ 0.01000) > 2 to ≤ 3 a. 23° +5 °C; typical resistance change: ±300 ppm/°C Certification Agilent Technologies certifies that this product met
Краткое содержание страницы № 17
3 Use, Maintenance, and Care of the Devices 3-1
Краткое содержание страницы № 18
Use, Maintenance, and Care of the Devices Electrostatic Discharge Electrostatic Discharge Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets). Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration
Краткое содержание страницы № 19
Use, Maintenance, and Care of the Devices Visual Inspection Visual Inspection Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles from the connector threads may fall into the connector when it is disconnected. One connection made with a dirty or damaged connector can damage both connectors beyond repair. In some cases, magnification is necessary to see damage on a connector; a magnifying device with a magnification of ≥10× is recommen
Краткое содержание страницы № 20
Use, Maintenance, and Care of the Devices Cleaning Connectors Inspect the Precision Slotless Connectors (female) Precision slotless female connectors are used to improve accuracy. The slotless contacts are not affected by the slight variations in male contact pin diameter. However, it is still advisable to inspect them regularly for damage. NOTE Inspection is particularly important when mating nonprecision to precision devices. Cleaning Connectors Clean connectors are essential for ensuring