Краткое содержание страницы № 1
User’s Guide
Laser Diode Temperature Controller
LDX-5100
ILX Lightwave Corporation · P. O. Box 6310 · Bozeman, MT, U.S.A. 59771 · U.S. & Canada: 1-800-459-9459 · International Inquiries: 406-556-2481 · Fax 406-586-9405
ilx.custhelp.com · www.ilxlightwave.com
70014302 May 2007
Краткое содержание страницы № 2
Краткое содержание страницы № 3
TABLE OF CONTENTS TABLE OF CONTENTS Table of Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i Safety and Warranty Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii Safety Information and the Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii General Safety Considerations . . . . . . . . . . . .
Краткое содержание страницы № 4
TABLE OF CONTENTS TEC Gain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Measurement/Control (J2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Output Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Output Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Thermistor Monitor .
Краткое содержание страницы № 5
SAFETY AND WARRANTY INFORMATION The Safety and Warranty Information section provides details about cautionary symbols used in the manual, safety markings used on the instrument, and information about the Warranty including Customer Service contact information. Safety Information and the Manual Throughout this manual, you will see the words Caution and Warning indicating potentially dangero
Краткое содержание страницы № 6
SAFETY AND WARRANTY INFORMATION Safety Symbols SAFETY SYMBOLS This section describes the safety symbols and classifications. Technical specifications including electrical ratings and weight are included within the manual. See the Table of Contents to locate the specifications and other product information. The following classifications are standard across all ILX Lightwave products: • Indoor use only • Ordinary Protection: This product is NOT protected against the harmful ingress of moisture
Краткое содержание страницы № 7
SAFETY AND WARRANTY INFORMATION Warranty WARRANTY ILX LIGHTWAVE CORPORATION warrants this instrument to be free from defects in material and workmanship for a period of one year from date of shipment. During the warranty period, ILX will repair or replace the unit, at our option, without charge. Limitations This warranty does not apply to fuses, lamps, defects caused by abuse, modifications, or to use of the product for which it was not intended. This warranty is in lieu of all other warra
Краткое содержание страницы № 8
SAFETY AND WARRANTY INFORMATION Warranty Claims for Shipping Damage When you receive the instrument, inspect it immediately for any damage or shortages on the packing list. If the instrument is damaged, file a claim with the carrier. The factory will supply you with a quotation for estimated costs of repair. You must negotiate and settle with the carrier for the amount of damage. vi Safety and Warranty Information
Краткое содержание страницы № 9
SAFETY AND WARRANTY INFORMATION Warranty Comments, Suggestions, and Problems To ensure that you get the most out of your ILX Lightwave product, we ask that you direct any product operation or service related questions or comments to ILX Lightwave Customer Support. You may contact us in whatever way is most convenient: Phone . . . . . . . . . . . . . . . . . . . . . . . . . . . (800) 459-9459 or (406) 586-1244 Fax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Краткое содержание страницы № 10
SAFETY AND WARRANTY INFORMATION Warranty in a container with at least 3 inches (7.5 cm) of compressible packaging material on all sides. viii Safety and Warranty Information
Краткое содержание страницы № 11
CHAPTER1 INTRODUCTION This manual contains operation, calibration and specification information for the LDT-5100. It is recommended that the user become thoroughly familiar with this information prior to operating the LDT-5100. If any of the following symptoms exist, or are even suspected, remove the LDT-5100 Temperature Controller from service. Do not use until trained service personnel can
Краткое содержание страницы № 12
INTRODUCTION CHAPTER 1 Product Overview Product Overview The LDT-5100 Temperature Controller is a device which, when powered by the appropriate supply, measures the temperature dependent resistance of a thermistor. Using this resistance as feedback, it stabilizes temperature by driving current to a thermoelectric (TE) module. The TE drive for the LDT-5100 is bipolar (positive current to cool; negative current to heat). This allows for highly stable temperatures and fast settling times. Desi
Краткое содержание страницы № 13
INTRODUCTION CHAPTER 1 1 Specifications 1 Specifications INPUTS 2 Power Supply + 12 to 18 Volts at 75 mA 3 + 4.5 to 7 Volts at 2 A Control Output On (Default) when power supplied; user can control output on / off with TTL line (Low = output off) 4 Thermistor R Setting Jumper selectable internal/external R set Thermistor R driven to match reference precision resistor across External R set connector pins 5, 6 Internal R set 50 k Ω adjustable 10 T pot Thermistor Current Jumper selectable 10 µA /
Краткое содержание страницы № 14
INTRODUCTION CHAPTER 1 1 Specifications LED Indicators 6 Output On PCB pads for external LED (10 mA) and open collector output 6 TEC Open Error PCB pads for external LED (10 mA) and open collector output 6 Sensor Open Error PCB pads for external LED (10 mA) and open collector output GENERAL LDT-5100 Size 1.75” x 3.75” x 5.5” (44mm x 95mm x 139mm) Weight 8.75 oz o Operating Temperature 0 to 50 C o Storage Temperature -40 to 70 C 1. All values tested after a one hour warm up period. 2. Additional
Краткое содержание страницы № 15
CHAPTER2 OPERATION See Figure 2.1 for location and identification of control connectors and jumpers. Thermistor Resistance Set Point Two methods of thermistor R set point control are provided in the LDT-5100; internal resistor or external resistor. A jumper, “RESISTOR SEL”, (J6) is used to select which resistor is used as the set point reference for the control loop. (A jumper on the left an
Краткое содержание страницы № 16
OPERATION CHAPTER 2 Figure 2.1 LDT-5100 Control/Monitor Connections 6 LDT-5100
Краткое содержание страницы № 17
OPERATION CHAPTER 2 Thermistor Current Source Thermistor Current Source For a greater range of temperature sensing, two current sources are provided on the LDT-5100; 100 µA and 10 µA. Each source is selectable on the “THERMISTOR CURRENT” jumper (J5). The default setting from the factory is 100 µA. Note: If the jumper is removed, the 100 µA current source (default) is selected. The choice of current source will depend on the thermistor sensor used. Please refer to LX Lightwave Application No
Краткое содержание страницы № 18
OPERATION CHAPTER 2 Measurement/Control (J2) Output Monitors The LDT-5100 has two buffered output monitors present on the measurement / control connector (J2). These monitor the actual thermistor resistance and I TE (TE drive current). Thermistor Monitor The thermistor monitor (pin 10) provides an output voltage which is proportional to the thermistor sense resistor value. The transfer function with the 100 µA thermistor current range selected is 10k Ω/V and with the 10 µA thermistor current
Краткое содержание страницы № 19
OPERATION CHAPTER 2 LED Indicators supply voltage, this threshold is not a fixed value. Under normal operation (with a low impedance TE), the control line will be in a high impedance state. Output On/Off (pin 2) This control line indicates if the TE output is on (driving current to a TE load). A ground at this control line indicates that the output is off and no current is being driven through the TE modules. LED Indicators Pads are located on the LDT-5100 PCB to allow direct connection of in
Краткое содержание страницы № 20
OPERATION CHAPTER 2 DC Power-In Connector (J3) 10 LDT-5100