Manual Addendum
Model GFC 7000T
CO Analyzer
2
(Addendum to GFC 7000TA Manual, PN 07272)
P/N M07273
DATE 06/11/13
TELEDYNE ELECTRONIC TECHNOLOGIES
Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91748
Telephone: (626) 934-1500
Fax: (626) 961-2538
Web: www.teledyne-ai.com
Teledyne Analytical Instruments
Краткое содержание страницы № 2
Model GFC7000T Carbon Dioxide Analyzer Copyright © 2013 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language or computer language in whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne Analytical Instruments, 16830 Chestnut Street, City of
Краткое содержание страницы № 3
Model GFC7000T Carbon Dioxide Analyzer ABOUT THIS ADDENDUM This addendum, part number 07273, is to be used in conjunction with the Model GFC 7000TA operation manual, part number 07272. Where operation of the Model GFC 7000T diverges from that of the Model GFC 7000TA, this addendum takes precedence. Please note that the GFC 7000TA operation manual contains important SAFETY messages for this instrument. It is strongly recommended that you read that operation manual in its entirety as wel
Краткое содержание страницы № 4
Model GFC7000T Carbon Dioxide Analyzer This page intentionally left blank Teledyne Analytical Instruments iv
Краткое содержание страницы № 5
Model GFC7000T Carbon Dioxide Analyzer TABLE OF CONTENTS About This Addendum ............................................................................................................... iii Table of Contents ...................................................................................................................... v 1. Overview ............................................................................................................................ 1 2. Specifications an
Краткое содержание страницы № 6
Model GFC7000T Carbon Dioxide Analyzer This page intentionally left blank. Teledyne Analytical Instruments vi
Краткое содержание страницы № 7
Model GFC7000T Carbon Dioxide Analyzer 1. OVERVIEW The Model GFC 7000T is a close derivative of the Model GFC 7000TA CO Analyzer. The 2 main differences are as follows. The optical bench is longer (14 m), making the instrument more sensitive at low CO 2 levels. The instrument has a user-selectable full scale range of 0-100 PPB to 0-100 PPM of CO . 2 The rear panel is different. There is an extra pneumatic port for the exhaust of the purge gas to the GFC wheel. Also, there are relay ala
Краткое содержание страницы № 8
Model GFC7000T Carbon Dioxide Analyzer PARAMETER SPECIFICATION 220 – 240 V 50/60 Hz Environmental Installation Category (Over voltage Category) II Pollution Degree 2 Conditions 1 Ethernet: 10/100Base-T Standard I/O 2 RS-232 (300 – 115,200 baud) 2 USB device ports 8 opto-isolated digital status outputs 6 opto-isolated digital control inputs 4 analog outputs 1 USB com port Optional I/O 1 RS485 8 analog inputs (0-10V, 12-bit) 4 digital alarm outputs Multidrop RS232 3 4-20mA curr
Краткое содержание страницы № 9
Model GFC7000T Carbon Dioxide Analyzer 4. PNEUMATIC CONNECTIONS This section provides information on the basic pneumatic configuration and the zero/span valve option pneumatic configuration. First are diagrams to illustrate the configurations, followed by a description of the pneumatic ports, and then step-by-step instructions for attaching the lines. 4.1. BASIC AND ZERO/SPAN VALVE OPTION CONFIGURATIONS CAUTION Sample and calibration gases should only come into contact with PTFE (Teflo
Краткое содержание страницы № 10
Model GFC7000T Carbon Dioxide Analyzer Figure 4-1. Pneumatic Connections for Pressurized Sample Delivery (w/o Internal Pump) Figure 4-2. Pneumatic Connections–Configuration with Zero/Span Valve Option Teledyne Analytical Instruments 4
Краткое содержание страницы № 11
Model GFC7000T Carbon Dioxide Analyzer Table 4-1. Model GFC 7000T Pneumatic Connections Rear Panel Label Function Connect a gas line from the source of sample gas here. SAMPLE Calibration gasses are also inlet here on units without zero/span valve option installed. Connect an exhaust gas line of not more than 10 meters EXHAUST long here. On units with zero/span valve option installed, connect a PRESSURE SPAN gas line to the source of calibrated span gas here. VENT SPAN Not used
Краткое содержание страницы № 12
Model GFC7000T Carbon Dioxide Analyzer EXAMPLE: If the application is to measure between 0 ppm and 50 ppm, an appropriate Span Gas would be 40 ppm. If the application is to measure between 0 ppm and 100 ppm, an appropriate Span Gas would be 80 ppm. Zero Air is similar in chemical composition to the earths atmosphere but scrubbed of all components that might affect the analyzer’s readings. In the case of CO 2 measurements this means CO less than 0.1 ppm of CO and Water Vapor. 2 2 Ze
Краткое содержание страницы № 13
Model GFC7000T Carbon Dioxide Analyzer 6. RELAY ALARM OUTPUTS (AIR PRODUCTS CONFIGURATION) There are 4 relay alarm outputs (AL1-AL4) on the rear panel. AL1 is for “system okay”, AL2 is for “high range status” and AL3 is for “zero calibration status” AL4 is not used The AL1 relay is energized when the system is okay and de-energized when the system has a fault. The AL2 relay is energized when the high auto-range is in use and and AL3 relays energize when the the instrument