Resumo do conteúdo contido na página número 1
Agilent E5052BU Option 018
Hard Disk Drive Kit
User’s Guide
First Edition
Manufacturing No. E5052-90202
May 2007
Resumo do conteúdo contido na página número 2
Notices The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies. © Copyright 2007 Agilent Technologies Manual Printing History The manual’s printing date and part number indicate its current edition. The printing date
Resumo do conteúdo contido na página número 3
1 General Information This chapter describes the Hard Disk Drive Kit, contents of the kit, and required tools to install the Option 018 into the Agilent E5052B. 3
Resumo do conteúdo contido na página número 4
General Information Option 018 Hard Disk Drive Kit Option 018 Hard Disk Drive Kit The purpose of this kit is to install the option 018 into the Agilent E5052B. Contents of the Agilent E5052BU Option 018 Hard Disk Drive Kit The kit includes the parts listed in Table 1-1. If anything is missing, contact the nearest Agilent Technologies office. Table 1-1 Contents of the Agilent E5052BU Option 018 Hard Disk Drive Kit Description Qty. Upgrade Manual (This manual) 1 HDD Assembly 1 4 Chapter 1
Resumo do conteúdo contido na página número 5
2 Option Installation This chapter describes the procedure to install the option 018 into the Agilent E5052B. 5
Resumo do conteúdo contido na página número 6
Option Installation Introduction Introduction This chapter describes the procedure to install Option 018 into the E5052B. The procedure consists of the following items. • Installation Requirement • Saving System calibration Data from Removable Hard Disk to USB Pen Drive • Removing Removable Hard Disk • Mounting Removable Hard Disk • Initial Registration of E5052B • Calibration of the Touch Screen • Recalling System Calibration Data from USB Pen drive to Removable Hard Disk Procedure Installatio
Resumo do conteúdo contido na página número 7
Option Installation Procedure Saving System calibration Data from Removable Hard Disk to USB Pen Drive The following is a description of the procedures to save the system calibration data stored in the removable hard disk to a USB Pen drive. Step 1. Exit the measurement window. System > Service Menu > Administorator Menu > Exit E5052B > OK Step 2. Insert a USB Pen drive into the USB port in front panel. Step 3. Open D:\Recovery drive. Step 4. Click .D:\syscal from the explorer. Step 5. Select
Resumo do conteúdo contido na página número 8
Option Installation Procedure Removing Removable Hard Disk Follow the steps below to remove a removable hard disk. CAUTION Before removing a hard disk, you must turn off the unit’s power and then disconnect the power cord from the outlet. Step 1. Turn off the power. Step 2. Insert the key in the lock and turn anti-clockwise. Step 3. Open the screw in clockwise direction. Step 4. Press the button under direction. Step 5. Remove the hard disk. Mounting Removable Hard Disk Follow the steps below t
Resumo do conteúdo contido na página número 9
Option Installation Procedure Initial Registration of E5052B When you start up the E5052B at the first time, you need to perform the initial registration of the Windows XP operating system of the E5052B. NOTE You cannot use the front panel keys during the initial registration of the E5052B therefore connect the mouse and the keyboard before turning on the power. NOTE If you perform the following procedure incorrectly, a message asking you whether to return to the previous registration screen
Resumo do conteúdo contido na página número 10
Option Installation Procedure Step 3. In the Windows XP Professional Setup dialog box, select the I accept this agreement box and click the Next >button (Figure 2-2). Figure 2-2 Windows XP Professional Setup dialog box (1/2) e5071cse1074 Step 4. In the next dialog box , input E5052 in the Name box. Then, click the Next> button (Figure 2-3). Figure 2-3 Windows XP Professional Setup dialog box (2/2) E5052 e5052bse1033 Step 5. After a while, the E5052B restart automatically. Then, the measuremen
Resumo do conteúdo contido na página número 11
Option Installation Procedure Calibration of the Touch Screen When you have replaced the HDD on an E5052B, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen. Step 1.Press . Step 2. Press Service Menu. Step 3. Press Test Menu. Step 4. Press Adjust Touch Screen. The touch screen calibration screen (Figure 2-4) appears. Figure 2-4 Touch Panel Calibration Screen Step 5. Touch the x mark on the upper left with your finger. The mark x appears
Resumo do conteúdo contido na página número 12
Option Installation Procedure Recalling System Calibration Data from USB Pen drive to Removable Hard Disk The following is a description of the procedures to copy the system calibration data stored in the USB Pen drive to the removable hard disk. Step 1. Exit the measurement window. System > Service Menu > Administorator Menu > Exit E5052B > OK Step 2. Insert a USB pen drive in to the USB port in front panel. Step 3. Open D:\Recovery drive. Step 4. Click syscal folder in the USB Pen drive fro
Resumo do conteúdo contido na página número 13
REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative. 21/01/2004 United States: Australia/New Zealand: Test and Measurement Call Center (tel) (61 3) 9210-5555 (Australia) (tel) 1 800 452-4844