Resumo do conteúdo contido na página número 1
LF654 /LF620
Field Intelligent Device – Mount Anywhere Series - Flanged
LF654 /LF622
Electromagnetic Flowmeter
1/2" to 18" (15 to 450 mm)
Introduction
The electromagnetic flowmeter uses Faraday’s Law of
electromagnetic induction to measure the process flow.
The device consists of two units: a detector, through
which the fluid to be measured flows and in which
low-level signals proportional to flow rates are
obtained; and a converter, which supplies excitation
current to the
Resumo do conteúdo contido na página número 2
TIC-LF654L Drinking water approvals: Fluid conductivity: PU and HR liner: NSF certified 3µS/cm minimum Fluid temperature: Model LF654 Detector -4 to 248 °F (-20 to +120 °C): Mounting style: Flange connection type PFA lined flowmeter - Fluid pressure: -4 to 140 °F ( 20 to +60 °C): - - - 15 psi or 1.0 bar ( 0.1 MPa) to the nominal Polyurethane lined flowmeter - pressure of the connection flange. 23 to 176 °F ( 5to +80 °C): Note: Before shipping from the man
Resumo do conteúdo contido na página número 3
TIC-LF654L Pulse width: 0.5 to 500ms (but less than half of Model LF620 and LF622 converters the period for 100% flow rate) Input signals Note: The same and simultaneous pulse is not available Analog signal — the voltage signal from detector, between DO1 and DO2.) proportional to process flow rate (for LF622 • Multi-range selection outputs (Note 1) separate type converter). • High, High high, Low, and/or Low low alarm Digital input DI outputs (Note 2) Signal type: 20 to 30Vdc
Resumo do conteúdo contido na página número 4
TIC-LF654L LCD display: Surge protection: Arresters are installed in the power supply and a Full dot-matrix 128×128 dot LCD display current signal output circuit to help protect the meter (back–light provided) from lightning and improve personnel safety. A parameter change will rotate the display. Case: Aluminum alloy (equal to IP 67) Parameter settings — Parameters can be set as follows: Coating: • IR Switches: Three key switches are provided Acrylic resin-baked coating, pearl–
Resumo do conteúdo contido na página número 5
TIC-LF654L Installation Dimensions Unit: inch (mm) Note1: Eye bolts are provided at the top for flowmeters sized 8" (200mm) or above, and further, a roll-prevention base is provided for flowmeters sized 10" (250mm) or larger. Note2: Cable glands are not provided for LF654 of cFMus approved type. Refer to the cable connection port at converter. ASME B 16.5 class 150 dimensions: JIS B 2220 10K dimensions: Meter size L1 L2 L3 No. of Weight Meter size L1 L2 L3 No. of Weight (inch) (
Resumo do conteúdo contido na página número 6
TIC-LF654L Note1: Eye bolts are provided at the top for flowmeters sized 8" (200mm) or above, and further, a roll-prevention base is provided for flowmeters sized 10" (250mm) or larger. Note2: Cable glands are not provided for LF654 of cFMus approved type. Refer to the cable connection port at converter. ASME B 16.5 class 150 dimensions: JIS B 2220 10K dimensions: Unit: inch (mm) Meter size L1 L2 L3 No. of Weight Meter size L1 L2 L3 No. of Weight (inch) (inch) (inch) (inch) bolts (l
Resumo do conteúdo contido na página número 7
TIC-LF654L Note1: Eye bolts are provided at the top for flowmeters sized 8”(200mm) or above, and further, a roll-prevention base is provided for flowmeters 10”(250mm) or larger. Note2: Cable glands are not provided for LF654 of cFMus approved type. Refer to the cable connection d JIS B 2220 10K dimensions: ASME B 16.5 class 150 dimensions: Meter size L1 L2 L3 No. of Weight Meter size L1 (L2) L3 No. of Weight (inch) (inch) (in
Resumo do conteúdo contido na página número 8
TIC-LF654L Note1: Eye bolts are provided at the top for flowmeters sized 8" (200mm) or above, and further, a roll-prevention base is provided for flowmeters sized ASME B 16.5 class 150 dimensions: JIS B 2220 10K dimensions: Meter size L1 L2 L3 No. of Weight Meter size L1 (L2) L3 No. of Weight (inch) (inch) (inch) (inch) bolts (lbs) (mm) (mm) (mm) (mm) bolts (kg) 1/2 7.90 6.77 8.54 4 approx. 9.0 15 200 170 218 4 approx. 4.5 7.90 1 7.01
Resumo do conteúdo contido na página número 9
TIC-LF654L Unit: inch (mm) Weight: Approx. 7 lb (3.5 kg) Note: Cable glands are not provided for LF622F cFMus approved type. Refer to the part Cable connection port at detector. Note: 1 inch = 25.4 mm Figure 7. Separate type converter LF622 and LF622F External Connections ● Combined type LF650/LF620 flowmeter and LF654/LF620F flowmeters Instrument panel : Ordered separately Grounding with 100
Resumo do conteúdo contido na página número 10
TIC-LF654L ● Separate type LF650/LF622 flowmeter and LF654/LF622F flowmeter Instrument panel : Ordered separately Grounding with 100 or less 2 wire 5.5mm or more ground resistance Power switch Grounding with 100 or less (External double-pole power switch) ground resistance Power supply Thick walled steel conduit Current output (4 20mAdc) Signal cable or PROFIBUS (2-wire shielded hard-rubber sheathed cable) Digital input (20 30Vdc) Excitation cable or Modbus (3-wire shield
Resumo do conteúdo contido na página número 11
TIC-LF654L (6) The electromagnetic flowmeter is not equipped with Wiring Precautions terminating resistors. Use the terminating resistor (1) Explosion proof type flowmeters are not provided unit for PROFIBUS-PA or junction box, if with cable glands. necessary. Refer to the part Cable connection port at detector (7) Only one PROFIBUS-PA cable goes through a and converter. cable gland of the Electromagnetic Flowmeter. (2) Connect the grounding wire (IV wire 5.5mm² or Please use t
Resumo do conteúdo contido na página número 12
TIC-LF654L Table 3. Flow Rate and Flow velocity Ordering Information 1. When ordering the LF654 series flowmeters, refer (SI unit) 3 to Tables 4 and 5 (Type Specification Codes). Unit: m /h Flow rate An entry must be made for each of the columns in Size 3.0 12 each of these tables. 0.1 m/s 0.3 m/s 1.0 m/s (mm) m/s m/s 2. Fluid characteristics: 15 0.06362 0.1908 0.6361 1.908 7.634 (1) Type of fluid to be measured and its characteristics 25 0.1767 0.5301 1.767 5.301 21.21
Resumo do conteúdo contido na página número 13
TIC-LF654L Table 4. Specification Code (Flange type detector LF654 Series) Model Specification Code Detector category Description 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Gr.-A Gr.-B L F 6 5 Flange type electromagnetic flowmeter detector Usage 4 cFMus Class 1 Division 2 Hazardous Location √ √ Meter size D ½"(15mm) √ E 1"(25mm) √ S 1¼"(32mm) √ F 1½"(40mm) √ G 2"(50mm) √ W 2½"(65mm)
Resumo do conteúdo contido na página número 14
TIC-LF654L Table 5. Specification Code for converters Model Specification Code Contents LF620 LF622 1 2 3 4 5 6 7 8 9 1011 12 13 14 type type L F 6 2 Electromagnetic flowmeter converter 0 Combined (Integral) type – 2 Separate (Remote) type – Purpose F cFMus class I, Division 2 approved Shape A Standard type with case Converter mounting fitting A None – C Panel, Acce