User’s and Service Guide
Agilent Technologies 85054B
50Ω Type-N Calibration Kit
This manual applies directly to 85054B calibration kits with serial number
prefix 3101A. The calibration devices in this kit are individually serialized.
Record the device serial numbers in the table provided in this manual (see
“Recording the Device Serial Numbers” in Chapter 1.)
Manufacturing Part Number: 85054-90049
Printed in USA
Print Date: June 2002
Supersedes: Januray 1997
© Copyright 1993, 1997, 2002 Ag
Resumo do conteúdo contido na página número 2
Documentation Warranty THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR
Resumo do conteúdo contido na página número 3
Contents 1. General Information Calibration Kit Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2 Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2 Calibration Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3 Incoming Inspection . . . . . . . . . . . . . . . . . . .
Resumo do conteúdo contido na página número 4
Contents 5. Troubleshooting Troubleshooting Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2 Returning a Kit or Device to Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 Where to Look for More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3 Contacting Agilent. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Resumo do conteúdo contido na página número 5
1 General Information 85054B 1-1
Resumo do conteúdo contido na página número 6
General Information Calibration Kit Overview Calibration Kit Overview The Agilent 85054B type-N calibration kit is used to calibrate Agilent network analyzers up to 18 GHz for measurements of components with 50Ω type-N connectors. The standards in this calibration kit allow you to perform simple 1- or 2-port and TRM (thru–reflect–match) calibrations. This manual describes the 85054B calibration kit and provides replacement part numbers, specifications, and procedures for using, maintaining, a
Resumo do conteúdo contido na página número 7
General Information Calibration Kit Overview Lowband Loads The lowband loads are metrology-grade, 50 ohm terminations which have been optimized for lowband performance up to 2 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss. Sliding Loads The sliding loads in this kit are designed to provide excellent performance from 3 GHz to 18 GHz. The inner and outer conductors of the a
Resumo do conteúdo contido na página número 8
General Information Incoming Inspection Incoming Inspection Refer to “Kit Contents” on page 1-2 to verify a complete shipment. Use Table 1-1 on page 1-5 to record the serial numbers of all serialized devices in your kit. Check for damage. The foam-lined storage case provides protection during shipping. If the case or any device appears damaged, or if the shipment is incomplete, refer to “Contacting Agilent” on page 5-3. Agilent will arrange for repair or replacement of incomplete or damaged
Resumo do conteúdo contido na página número 9
General Information Serial Numbers Serial Numbers A serial number is attached to this calibration kit. The first four digits followed by a letter comprise the serial number prefix; the last five digits are the suffix, unique to each calibration kit. Recording the Device Serial Numbers In addition to the kit serial number, the devices in the kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in Table 1-1. Recording the serial
Resumo do conteúdo contido na página número 10
General Information Calibration Kits Documented in This Manual Calibration Kits Documented in This Manual This manual applies to any 85054B calibration kit whose serial number prefix is listed on the title page. If your calibration kit has a different serial number prefix, refer to the “Calibration Kit History” section below for information on how this manual applies. Calibration Kit History This section describes calibration kits with serial number prefixes lower that the ones listed on the
Resumo do conteúdo contido na página número 11
General Information Precision Slotless Connectors Precision Slotless Connectors The female type-N connectors in this calibration kit are metrology-grade, precision slotless connectors (PSC). A characteristic of metrology-grade connectors is direct traceability to national measurement standards through their well-defined mechanical dimensions. Conventional female center conductors are slotted. When mated, the female center conductor is flared by the male pin. Because physical dimensions determ
Resumo do conteúdo contido na página número 12
General Information Clarifying the Sex of a Connector Clarifying the Sex of a Connector In this manual, the sex of calibration devices and adapters are referred to in terms of their connector interface. For example, a male open has a male connector. However, during a measurement calibration, the network analyzer softkey menus label a type-N calibration device with reference to the sex of the analyzer’s test port connector—not the calibration device connector. For example, the label SHORT(F) o
Resumo do conteúdo contido na página número 13
2 Specifications 2-1
Resumo do conteúdo contido na página número 14
Specifications Environmental Requirements Environmental Requirements Table 2-1 Environmental Requirements Parameter Limits a +20 °C to +26 °C (+68 °F to +79 °F) Operating temperature b ±1 °C of measurement calibration temperature Error-corrected temperature range Storage temperature - 40 °C to +75 °C (- 40 °F to +167 °F) Altitude Operation < 4,500 meters (≈15,000 feet) Storage < 15,000 meters (≈50,000 feet) Relative humidity Always non-condensing Operation 0 to 80% (26 °C maximum dry bulb)
Resumo do conteúdo contido na página número 15
Specifications Mechanical Characteristics Mechanical Characteristics Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Agilent Technologies verifies the mechanical characteristics of the devices in this kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any improper pin depth
Resumo do conteúdo contido na página número 16
Specifications Mechanical Characteristics NOTE The gages for measuring type-N connectors compensate for the designed offset of 5.26 mm (0.207 inch), therefore, protrusion and recession readings are in relation to a zero reference plane (as if the inner and outer conductor planes were intended to be flush). Gage readings can be directly compared with the observed values listed in Table 2-2. The pin depth value of each calibration device in this kit is not specified, but is an important mecha
Resumo do conteúdo contido na página número 17
Specifications Electrical Specifications Electrical Specifications The electrical specifications in Table 2-3 apply to the devices in your calibration kit when connected with an Agilent precision interface. Table 2-3 Electrical Specifications Device Frequency (GHz) Parameter Specification Lowband loads DC to ≤2 Return Loss ≥48 dB (≤0.00398ρ) a > 2 to ≤18 Return Loss ≥42 dB (≤0.00794ρ) Sliding loads ≥34 dB (≤0.0200ρ) Adapters DC to ≤8 Return Loss ≥28 dB (≤0.0398ρ) (both styles) > 8 to ≤18 Ret
Resumo do conteúdo contido na página número 18
Specifications Electrical Specifications 2-6 85054B
Resumo do conteúdo contido na página número 19
3 Use, Maintenance, and Care of the Devices 3-1
Resumo do conteúdo contido na página número 20
Use, Maintenance, and Care of the Devices Electrostatic Discharge Electrostatic Discharge Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets). Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration