Resumo do conteúdo contido na página número 1
Instruction Manual
PN 51-6081P/rev.A
November 2008
Model 6081-P
Wireless pH/ORP Transmitter
Resumo do conteúdo contido na página número 2
ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Ana
Resumo do conteúdo contido na página número 3
CAUTION ® If a Model 375 Universal Hart Communicator is used with these transmitters, the software within the Model 375 may require modification. If a software modification is required, please contact your local Emerson Process Management Service Group or National Response Center at 1-800-654-7768. About This Document This manual contains instructions for installation and operation of the Model 6081-P Wireless pH/ORP Transmitter. The following list provides notes concerning all revisions of thi
Resumo do conteúdo contido na página número 4
QUICK START GUIDE FOR MODEL 6081 WIRELESS pH TRANSMITTER 1. Refer to Section 2.0 Installation for installation instructions. 2 . Wire the pH or ORP sensor to the transmitter. Refer to the sensor instruction sheet for details. 3. Once the connections are secure and verified, install the Power Module to power to the transmitter. 4. When the transmitter is powered up for the first time, Quick Start screens appear. Using Quick Start is easy. a. A blinking field shows the position of the curs
Resumo do conteúdo contido na página número 5
Quick Start Menu English Francais Deutsch Italiano Espanol >> Portuguese >> Measure? pH Redox ORP Use Preamp in? Xmtr Sensor Diagnostics? On Off Measure update rate: 1min Temperature in? °C °F Setup Wireless Network? Yes No Wireless Network HART PollAddress Network ID: 00000 Network Join Key 1 of 4: 00000000 Network Join Key 2 of 4: 00000000 Network Join Key 3 of 4: 00000000 Network Join Key 4 of 4: 00000000 Wireless Network HART
Resumo do conteúdo contido na página número 6
F FIGURE 1-1. MENU TREE FOR MODEL 6081 pH Wireless TRANSMITTER
Resumo do conteúdo contido na página número 7
MODEL 6081-P pH/ORP TABLE OF CONTENTS MODEL 6081-P pH/ORP WIRELESS TRANSMITTER TABLE OF CONTENTS Section Title Page 1.0 DESCRIPTION AND SPECIFICATIONS ................................................................ 1 1.1 Features and Applications........................................................................................ 1 1.2 Specifications – General ......................................................................................... 1 1.3 Specifications –Wireless.............
Resumo do conteúdo contido na página número 8
MODEL 6081-P pH/ORP TABLE OF CONTENTS TABLE OF CONTENTS CONT’D 8.0 PROGRAMMING THE TRANSMITTER.................................................................. 25 8.1 General .................................................................................................................... 25 8.2 Changing Start-up Settings ...................................................................................... 25 8.3 Choosing and Configuring the Analytical Measurement ...........................
Resumo do conteúdo contido na página número 9
MODEL 6081-P pH/ORP TABLE OF CONTENTS LIST OF FIGURES Number Title Page 2-1 Wall Mounting Insatallation Model 6081................................................................... 8 2-2 Pipe Mounting Insatallation Model 6081 .................................................................. 9 2-3 Removing Rear Cover.............................................................................................. 10 2-4 Power Module Warning Label ....................................................
Resumo do conteúdo contido na página número 10
MODEL 6081-P pH/ORP TABLE OF CONTENT This page left blank intentionally iv
Resumo do conteúdo contido na página número 11
MODEL 6081-P pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS SECTION 1.0 DESCRIPTION AND SPECIFICATIONS • HIGH ACCURACY AND RELIABILITY for monitoring applications • SELF-ORGANIZING NETWORK for high data reliability and network stability • INDUSTRY LEADING WIRELESS SECURITY • COMPATIBLE WITH 1420 WIRELESS GATEWAY and Emerson Process Management TM WirelessHART networks • EASY TO READ two-line display with easy to use menus • WirelessHART 7 Digital Communications • SMART Sensor Enabled • CONTINU
Resumo do conteúdo contido na página número 12
MODEL 6081-P pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS Automatic Temperature Compensation: External 3-wire 1.3 SPECIFICATIONS - WIRELESS Pt100 RTD or Pt1000 RTD located in the sensor, Output: WirelessHART V7 compensates the pH reading for temperature Transmit Rate: User selectable, 1/sec. to 1/60 min fluctuations. Compensation covers the range -10 (via 1420 Wireless Gateway or AMS) to 150°C (14 to 302°F). Manual temperature Measurement update rate: 1/sec. to 1/10 min compensation is als
Resumo do conteúdo contido na página número 13
MODEL 6081-P pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS FCC and IC 1.5 Product Certifications This device complies with Part 15 of the FCC Rules. Telecommunication Compliance Operation is subject to the following conditions: This All wireless devices require certification to ensure that device may not cause harmful interference, this device they adhere to regulations regarding the use of the RF must accept any interference received, including inter- spectrum. Nearly every country require
Resumo do conteúdo contido na página número 14
MODEL 6081-P pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICATIONS 1.7 ASSET MANAGEMENT SOLUTIONS Asset Management Solutions (AMS) is software that helps plant personnel better monitor the performance of ana- lytical instruments, pressure and temperature transmitters, and control valves. Continuous monitoring means maintenance personnel can anticipate equipment failures and plan preventative measures before costly break- down maintenance is required. AMS uses remote monitoring. The operator, sittin
Resumo do conteúdo contido na página número 15
MODEL 6081-P pH/ORP SECTION 2.0 INSTALLATION SECTION 2.0 INSTALLATION 2.1 Considerations 2.2 Unpacking and Inspection 2.3 Pre-Installation Set Up 2.4 Mechanical Installation 2.5 Ground the Transmitter 2.6 Power Module Installation 2.1 CONSIDERATIONS The transmitter can be commissioned before or after installation. It may be useful to commission it on the bench, before installation, to ensure proper operation and to become familiar with its functionality. When applicable, make sure the instru- me
Resumo do conteúdo contido na página número 16
MODEL 6081-P pH/ORP SECTION 2.0 INSTALLATION 2.3 PRE-INSTALLATION SETUP 2.3.1 Temperature Element The Model 6081-P pH/ORP transmitter is compatible with sensors having Pt 100 and Pt 1000. Sensors from other manufacturers may have a Pt 1000 RTD. For Rosemount Analytical sensors, the type of temperature element in the sensor is printed on the tag attached to the sensor cable. For the majority of sensors manufactured by Rosemount Analytical, the RTD IN lead is red and the RTD RTN lead is white. The
Resumo do conteúdo contido na página número 17
MODEL 6081-P pH/ORP SECTION 2.0 INSTALLATION NOTE: Always use facility recommended wiring practices. WARNING Failure to follow these installation guidelines could result in death or serious injury. • Make sure only qualified personnel perform the installation. Explosions could result in death or serious injury. • Before connecting a 375 Field Communicator in an explosive atmosphere, make sure the instruments are installed in accordance with intrinsically safe or non-incendive field wiring practi
Resumo do conteúdo contido na página número 18
MODEL 6081-P pH/ORP SECTION 2.0 INSTALLATION 2.4.2 Mounting on a Flat Surface. Figure 2-1. Wall Mounting Installation Model 6081. Use Pipe/Wall Mounting Bracket Kit, PN 23820-00 Note: PN 23820-00 mounting bracket kit includes mounting hardware for pipe mounting only. Wall mounting hardware to be provided by customer. Only use suitable fasteners and hardware to securely fasten the bracket and transmitter to the wall surface. 8
Resumo do conteúdo contido na página número 19
MODEL 6081-P pH/ORP SECTION 2.0 INSTALLATION 2.4.3 Pipe Mounting. The pipe mounting kit (PN 23820-00/01) accommodates 1-1/2 to 2 in. pipe. Figure 2-2. Pipe Mounting Installation Model 6081. Use Pipe/Wall Mounting Bracket Kit, PN 23820-00 9
Resumo do conteúdo contido na página número 20
MODEL 6081-P pH/ORP SECTION 2.0 INSTALLATION 2.6 POWER MODULE INSTALLATION The section describes the procedure for installation of the power module (part number 00753-9220-0001). The power module should stored in a safe place with a controlled environment until the Model 6081 is ready for live operation. For first time installation of the power module, follow these steps: 1. Unscrew the two long machine screws to remove the rear cover of the Model 6081. Separate the rear cover from the