Resumo do conteúdo contido na página número 1
OPERATING GUIDE
GPIB DC POWER SUPPLIES
Agilent Technologies Models
664xA, 665xA, 667xA, 668xA, and 669xA
sA
Agilent Part No. 5964-8267 Printed in Malaysia
Microfiche Part No. 5964-8268 January, 2005
Resumo do conteúdo contido na página número 2
Certification Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members. Warranty This Agilent Technologies hardware pr
Resumo do conteúdo contido na página número 3
Safety Summary The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer's failure to comply with these requirements. GENERAL Any LEDs used in this product are Class 1 LEDs as per
Resumo do conteúdo contido na página número 4
Safety Symbol Definitions Symbol Description Symbol Description Direct current Terminal for Line conductor on permanently installed equipment Alternating current Caution, risk of electric shock Both direct and alternating current Caution, hot surface Three-phase alternating current Caution (refer to accompanying documents) Earth (ground) terminal In position of a bi-stable push control Protective earth (ground) terminal Out position of a bi-stable push control On (
Resumo do conteúdo contido na página número 5
DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 Manufacturer’s Name and Address Responsible Party Alternate Manufacturing Site Agilent Technologies, Inc. Agilent Technologies (Malaysia) Sdn. Bhd 550 Clark Drive, Suite 101 Malaysia Manufacturing Budd Lake, New Jersey 07828 Bayan Lepas Free Industrial Zone, PH III USA 11900 Penang, Malaysia Declares under sole responsibility that the product as originally delivered Product Names a) Single
Resumo do conteúdo contido na página número 6
DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 Manufacturer’s Name and Address Responsible Party Alternate Manufacturing Site Agilent Technologies, Inc. Agilent Technologies (Malaysia) Sdn. Bhd 550 Clark Drive, Suite 101 Malaysia Manufacturing Budd Lake, New Jersey 07828 Bayan Lepas Free Industrial Zone, PH III USA 11900 Penang, Malaysia Declares under sole responsibility that the product as originally delivered Product Names a) Single
Resumo do conteúdo contido na página número 7
DECLARATION OF CONFORMITY According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 Manufacturer’s Name and Address Responsible Party Alternate Manufacturing Site Agilent Technologies, Inc. Agilent Technologies (Malaysia) Sdn. Bhd 550 Clark Drive, Suite 101 Malaysia Manufacturing Budd Lake, New Jersey 07828 Bayan Lepas Free Industrial Zone, PH III USA 11900 Penang, Malaysia Declares under sole responsibility that the product as originally delivered Product Names a) Single
Resumo do conteúdo contido na página número 8
Table of Contents Certification 2 Safety Summary 3 Printing History 4 Declarations 5 Table of Contents 8 GENERAL INFORMATION 11 Introduction 11 Safety Considerations 12 Instrument Identification 12 Options 12 Accessories 13 Description 13 Front Panel Programming 14 Remote Programming 14 Analog Programming 14 Output Characteristic 15 General 15 Downprogramming 15 Specifications and Supplemental Characteristics 15 INSTALLATION 43 Inspection 43 Damage 43 Packaging Material 43
Resumo do conteúdo contido na página número 9
Checksum Errors. 58 Runtime Error Messages 58 USER CONNECTIONS 59 Rear Panel Connections (All Models) 59 Load Wire Selection (All Models) 59 Analog Connector (All Models) 60 Digital Connector (All Models) 60 Connecting Series 664xA and 665xA Power Supplies to the Load 61 Output Isolation 61 Load Considerations 61 Local Voltage Sensing 62 Remote Voltage Sensing 63 Connecting One Supply to the Load 64 Connecting Supplies in Auto-Parallel 65 Connecting Supplies in Series 67 External
Resumo do conteúdo contido na página número 10
Introduction 95 Equipment Required 95 General Procedure 95 Parameters Calibrated 95 Test Setup 96 Front Panel Calibration 96 Entering the Calibration Values 96 Saving the Calibration Constants 96 Disabling the Calibration Mode 96 Changing the Calibration Password 96 Recovering From Calibration Problems 99 Calibration Error Messages 99 Calibration Over the GPIB 100 Calibration Language Dictionary 100 CAL:CURR 100 CAL:CURR:LEV 100 CAL:CURR:MON (Series 668xA/669xA only) 101 CAL:PAS
Resumo do conteúdo contido na página número 11
1 General Information Introduction Two guides are shipped with your power supply - an Operating Guide (this document) and a Programming Guide. You will find information on the following tasks in these guides: 1 Quick Document Orientation Topic Location Calibrating the power supply Appendix A - this guide Compatibility programming language Appendix B - Programming Guide Configuring the digital port Appendix D - this guide Line voltage: Connecting ac power source Chapter 2 - t
Resumo do conteúdo contido na página número 12
Safety Considerations This power supply is a Safety Class 1 instrument, which means it has a protective earth terminal. That terminal must be connected to earth ground through a power source equipped with a 3-wire ground receptacle. Refer to the Safety Summary page at the beginning of this guide for general safety information. Before installation or operation, check the power supply and review this guide for safety warnings and instructions. Safety warnings for specific procedures are loc
Resumo do conteúdo contido na página número 13
List of Options (continued) Option Description Used with Agilent Series 664xA 665xA 667xA 668xA 669xA Rack mount kit (Agilent 5062-3974) x 908 Rack mount kit (Agilent 5062-3977) x x Support rails (E3663AC) are required. Rack mount kit (Agilent 5062-3977 & 5062-3974) x x Support rails (E3663AC) are required. 909 Rack mount kit with handles (Agilent 5062-3975) x Rack mount kit with handles (Agilent 5062-3983) x x Support rails (E3663AC) are required. Rack mou
Resumo do conteúdo contido na página número 14
Operational features include: Constant voltage (CV) or constant current (CC) output over the rated output range. Built-in overvoltage (OV), overcurrent (OC), and overtemperature (OT) protection. Automatic turn-on selftest. Pushbutton nonvolatile storage and recall of up to 5 operating states (4 in Series 668xA/669xA supplies). Local or remote sensing of output voltage. Auto-parallel operation for increased total current. Series operation for increased total voltage. Ana
Resumo do conteúdo contido na página número 15
Output Characteristic General The power supply can operate in either CV (constant voltage) or CC (constant current) over its voltage and current ratings (see Table 1-l). The operating locus is shown by the Output Characteristic Curve in Table 1-2. The operating point is determined by the voltage setting (V ), the current setting (I ), and the load impedance. Two operating points are shown. s s Point 1 is defined by the load line cutting the operating locus in the constant-voltage regio
Resumo do conteúdo contido na página número 16
1 Table 1-1a. Performance Specifications for Series 664xA Parameter Agilent Model Number 6641A 6642A 6643A 6644A 6645A Output Ratings 0 - 8 V 0 - 20 V 0 - 35 V 0 - 60 V 0 - 120 V Voltage: 0 - 20 A 0 - 10 A 0 - 6 A 0 - 3.5 A 0 - 1.5 A Current:@ 40° ° ° °C 0 - 18 A 0 - 9 A 0 - 5.4 A 0 - 3.2 A 0 - 1.4 A Current:@ 50° ° ° °C Current:@ 55° ° ° °C 0 - 17 A 0 - 8.5 A 0 - 5.1 A 0 - 3.0 A 0 -1.4 A Programming Accuracy (@ 25 ± 5 °C) 2 Voltage:
Resumo do conteúdo contido na página número 17
1 Table 1-1b. Supplemental Characteristics for Series 664xA Parameter Agilent Model Number 6641A 6642A 6643A 6644A 6645A Output Programming Range (maximum programmable values) Voltage: 8.190 V 20.475 V 35.831 V 61.425 V 122.85 V Current: 20.475 A 10.237 A 6.142 A 3.583 A 1.535 A Overvoltage Protection (OVP): 8.8 V 22.0 V 38.5 V 66.0 V 132.0 V Average Resolution Voltage: 2 mV 5 mV 10 mV 15 mV 30 mV Current: 6 mA 3 mA 2 mA 1.2 mA 0.5 mA Overvoltage Protection (OVP)
Resumo do conteúdo contido na página número 18
1 Table 1-lb. Supplemental Characteristics for Series 664xA (continued) Parameter Agilent Model Number 6641A 6642A 6643A 6644A 6645A Maximum AC Line Current Ratings 100 Vac nominal: 4.4 A rms 120 Vac nominal: 3.8 A rms 220 Vac nominal: 2.2 A rms 230 Vac nominal: 2.1 A rms 240 Vac nominal: 2.0 A rms Maximum Reverse Bias Current: With AC input power applied and the dc output reverse biased by an external dc source, the supply will continuously withstand without damag
Resumo do conteúdo contido na página número 19
1 Table 1-1b. Supplemental Characteristics for Series 664xA (continued) Parameter All Models Digital Port Characteristics (see Table 1-6) GPIB Interface Capabilities (see Table 1-6) Serial Link Capabilities (see Table 1-6) Recommended Calibration Interval: 1 year Safety Compliance Complies with: CSA 22.2 No.231,IEC 348 Designed to comply with: UL 1244 RFI Suppression (complies with): CISPR-ll, Group 1, Class B Dimensions Width: 425.5 mm (16.75 in) Heigh
Resumo do conteúdo contido na página número 20
1 Table 1-1b. Supplemental Characteristics for Series 664xA (continued) Parameter All Models Output Impedance Curves (Typical): l Notes: For Performance Specifications, see Table l-la. 20 General Information