Resumo do conteúdo contido na página número 1
Installation and Quick Start Guide
Agilent Technologies
8719ET/ES
8720ET/ES
8722ET/ES
Network Analyzers
Part Number: 08720-90391
Printed in USA
Print Date: February 2001
Supersedes: May 2000
Resumo do conteúdo contido na página número 2
Notice The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. © Copyright 1999-2001 Agilent Technolo
Resumo do conteúdo contido na página número 3
Certification Agilent Technologies Company certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members. Regulatory and Warranty Information The regulatory
Resumo do conteúdo contido na página número 4
General Safety Considerations WARNING For continued protection against fire hazard replace line fuse only with same type and rating (115V operation: T 5A 125V UL/ 230V operation: T 4A H 250V IEC). The use of other fuses or material is prohibited. WARNING This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective co
Resumo do conteúdo contido na página número 5
Documentation Map The Installation and Quick Start Guide provides procedures for installing, configuring, and verifying the operation of the analyzer. It also will help you familiarize yourself with the basic operation of the analyzer. The User’s Guide shows how to make measurements, explains commonly-used features, and tells you how to get the most performance from your analyzer. The Reference Guide provides reference information, such as specifications, menu maps, and key definitions. The Program
Resumo do conteúdo contido na página número 6
Contents 1. Installing Your Analyzer Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2 STEP 1. Verify the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3 STEP 2. Familiarize Yourself with the Analyzer Front and Rear Panels . . . . . . . . . . . . . . .1-5 Analyzer Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Resumo do conteúdo contido na página número 7
1 Installing Your Analyzer 1-1
Resumo do conteúdo contido na página número 8
Installing Your Analyzer Introduction Introduction This chapter shows you how to install your analyzer and confirm the correct operation, by following the steps below: • STEP 1. Verify the Shipment on page 1-3 • STEP 2. Familiarize Yourself with the Analyzer Front and Rear Panels on page 1-5 • STEP 3. Meet Electrical and Environmental Requirements on page 1-7 • STEP 4. Configure the Analyzer on page 1-9 • STEP 5. Verify the Analyzer Operation on page 1-20 • STEP 6. Back Up the EEPROM Disk on page
Resumo do conteúdo contido na página número 9
Installing Your Analyzer STEP 1. Verify the Shipment STEP 1. Verify the Shipment 1. Unpack the contents of all the shipping containers. WARNING The analyzer weighs approximately 54 pounds (25 kilograms). Use correct lifting techniques. 2. Carefully inspect the analyzer to ensure that it was not damaged during shipment. NOTE If your analyzer was damaged during shipment, contact your nearest Agilent Technologies office or sales representative. A list of Agilent Technologies sales and service offices
Resumo do conteúdo contido na página número 10
Installing Your Analyzer STEP 1. Verify the Shipment 3. Verify that all the accessories have been included with the analyzer. ItemNumber Part Number Description 1 08720-90391 Installation and Quick Start Guide 2 08720-90392 User's Guide 3 08720-90393 Reference Guide 4 08753-90475 Programmer’s Guide 5 08720-90418 CD-ROM 6 unique to instrument EEPROM Backup Disk 7 unique to country AC power cable 8 5062-3978 Rack Flange Kit (Option 1CM only) 8 5062-4072 Rack Flange Kit with Handles (Option 1CP onl
Resumo do conteúdo contido na página número 11
Installing Your Analyzer STEP 2. Familiarize Yourself with the Analyzer Front and Rear Panels STEP 2. Familiarize Yourself with the Analyzer Front and Rear Panels Analyzer Front Panel CAUTION Do not mistake the line switch for the disk eject button. See the figure below. If the line switch is mistakenly pushed, the instrument will be turned off, losing all settings and data that have not been saved. 1 LINE (power on/off) switch 8 RESPONSE function block 2 Display 9 ACTIVE CHANNEL keys 3 Disk
Resumo do conteúdo contido na página número 12
Installing Your Analyzer STEP 2. Familiarize Yourself with the Analyzer Front and Rear Panels Analyzer Rear Panel a b 1 10 MHZ REFERENCE ADJUST 11 RF IN/OUT 2 10 MHZ PRECISION REFERENCE 12 Serial number plate a OUTPUT 13 BIAS INPUTS and FUSES 3 Fan 14 TEST SET I/O INTERCONNECT 4 Line voltage selector switch 15 MEASURE RESTART 5 Power cord receptacle, with fuse 16 LIMIT TEST 6 KEYBOARD input (mini-DIN) 17 TEST SEQUENCE 7 RS-232 interface 18 EXTERNAL TRIGGER connector
Resumo do conteúdo contido na página número 13
Installing Your Analyzer STEP 3. Meet Electrical and Environmental Requirements STEP 3. Meet Electrical and Environmental Requirements 1. Set the line-voltage selector to the 2. Ensure the available AC power source position that corresponds to the AC power meets the following requirements: source. • 90–132 VAC • 47–66 Hz / 400 Hz (single phase) -or- • 198–265 VAC • 47–66 Hz (single phase) The analyzer power consumption is 350 VA maximum. 3. Ensure the operating environment meets 4. Verify that t
Resumo do conteúdo contido na página número 14
Installing Your Analyzer STEP 3. Meet Electrical and Environmental Requirements 5. Ensure there are at least six inches of clearance between the sides and back of either the stand-alone analyzer or the system cabinet. CAUTION The environmental temperature must be 4 °C less than the maximum operating temperature of the analyzer for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is >800 watts, then you must provide forced convection. 6. Set up a static-safe
Resumo do conteúdo contido na página número 15
Installing Your Analyzer STEP 4. Configure the Analyzer STEP 4. Configure the Analyzer This step shows you how to set up your particular analyzer configuration. • standard configuration • Option 1D5 configuration - high stability frequency reference • printer or plotter configuration • rack-mount configuration Chapter 1 1-9
Resumo do conteúdo contido na página número 16
Installing Your Analyzer STEP 4. Configure the Analyzer To Configure the Standard Analyzer Connect test port cables and optional adapters if you are using other connector types. To Configure an Analyzer with a High Stability Frequency Reference (Option 1D5) Connect the jumper cable on the analyzer rear panel as shown. 1-10 Chapter1
Resumo do conteúdo contido na página número 17
Installing Your Analyzer STEP 4. Configure the Analyzer To Configure the Analyzer with Printers or Plotters 1. Connect your printer or plotter to the corresponding interface. 2. If you are using the parallel interface, press and toggle PARALLEL until your Local choice of COPY or GPIO appears. If you choose: COPY the parallel port is dedicated for normal copy device use (printers or plotters). GPIO the parallel port is dedicated for general purpose I/O. The analyzer controls the data inpu
Resumo do conteúdo contido na página número 18
Installing Your Analyzer STEP 4. Configure the Analyzer 3. Press SET ADDRESSES and then choose either PRINTER PORT or PLOTTER PORT , depending on your printer/plotter device. Or, if you are plotting your files to disk, press SET ADDRESSES PLOTTER PORT DISK . 4. Press the key that corresponds to your printer or plotter interface: GPIB , PARALLEL (parallel port), or SERIAL (serial port). NOTE The plotter menu is shown as an example. It will only appear if you select PLOTTER PORT . Similar interfa
Resumo do conteúdo contido na página número 19
Installing Your Analyzer STEP 4. Configure the Analyzer 5. If you will be using the serial port, adjust the analyzer's baud rate until it is equal to the baud rate set on the peripheral by pressing PLOTTER BAUD RATE or PRINTER BAUD RATE and the and front panel keys. NOTE The plotter menu is shown as an example. It will only appear if you select PLOTTER PORT . You can set the analyzer to the following baud rates: • 1200 • 2400 • 4800 • 9600 • 19200 6. Also, if you will be using the serial port,
Resumo do conteúdo contido na página número 20
Installing Your Analyzer STEP 4. Configure the Analyzer 7. If you will be creating a plot of the data, toggle PLTR TYPE until your choice of PLOTTER or HPGL PRT appears. • Choose PLOTTER for a pen plotter. HPGL PRT • Choose for a PCL5 a compatible printer. a 8. If you will be using a printer, toggle PRNTR TYPE until your printer choice appears. • Choose your printer type from these Hewlett-Packard printers: ❏ THINKJET ❏ DESKJET (except for HP DeskJet 540 and Deskjet 850C) ❏ LASERJET ❏ PAINTJET