Resumo do conteúdo contido na página número 1
6081-P
Wireless pH/ORP Transmitter
Instruction Sheet 6081-P
LIQ_MAN_ABR_6081-P October 2013
Figure 1. 6081-P
Essential Instructions – Read this before proceeding
Rosemount Analytical designs, manufactures, and tests its products to meet many national and international standards.
Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to
ensure they continue
Resumo do conteúdo contido na página número 2
6081-P Instruction Sheet October 2013 LIQ_MAN_ABR_6081-P Essential Instructions (continued) • Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources. • To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product. • When replacement parts are
Resumo do conteúdo contido na página número 3
Instruction Sheet 6081-P LIQ_MAN_ABR_6081-P October 2013 Quick Start Guide - 6081 Wireless pH Transmitter 1. Install the Power Module inside the rear enclosure. Follow the installation instructions on p.9 “Power Module Installation”. 2. Wire the pH or ORP sensor to the transmitter. Refer to the sensor instruction sheet for details. 3. Once the connections are secure and verified, install the Pow
Resumo do conteúdo contido na página número 4
6081-P Instruction Sheet October 2013 LIQ_MAN_ABR_6081-P Figure 2. Menu Tree for 6081 pH Wireless Transmitter 4
Resumo do conteúdo contido na página número 5
Instruction Sheet 6081-P LIQ_MAN_ABR_6081-P October 2013 Product Description When used with appropriate sensors, the 6081-P can measure pH or ORP (oxidation reduction potential) of a liquid and transmit data wirelessly with a radio transceiver which uses HART 7 communication protocol. The instrument has a local operator interface consisting of a keyboard and LCD display which can be used to observe process paramete
Resumo do conteúdo contido na página número 6
6081-P Instruction Sheet October 2013 LIQ_MAN_ABR_6081-P EEPROM Warning Sense Line Open Warning Factory Cal Warning Keyboard Warning Once a fault or warning is detected, the display will show a message describing the problem. Sensor Temperature Range: -10 to 150°C (PT100 and PT1000) Display: 2-line, 16 character display supports display of pH and mV units. Display shows temperature. Approvals: RFI/EMI: EN-61326 EN 301 489-1 V1.2 2002 EN 301 489-17: V1.4.1 2002
Resumo do conteúdo contido na página número 7
Instruction Sheet 6081-P LIQ_MAN_ABR_6081-P October 2013 pH Calibration Procedure – Auto Buffer Calibration 1. Obtain two buffer solutions. Ideally, the buffer values should bracket the range of pH values to be measured. 2. Remove the pH sensor from the process liquid. If the process and buffer temperatures are appreciably different place the sensor in a container of tap water at the buffer temperature. Do not start
Resumo do conteúdo contido na página número 8
6081-P Instruction Sheet October 2013 LIQ_MAN_ABR_6081-P Sensor Wiring Note For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or the Manual DVD enclosed with each product. 1056, 1057, 56, 5081, 6081, 54e, and XMT : http://www3.emersonprocess.com/raihome/sp/liquid/wiring/XMT/ 1066 and sensors with SMART preamps: http://www2.emersonprocess.com/en-US/brands/rosemount
Resumo do conteúdo contido na página número 9
Instruction Sheet 6081-P LIQ_MAN_ABR_6081-P October 2013 Power Module Installation / Replacement The expected life of the power module (PN 701PBKKF) is a minimum of four years at reference conditions. This section describes the procedure for initial installation and replacement of the power module (PN 701PBKKF). The new power module should be stored in a safe place with a controlled environment until the 6081 is rea
Resumo do conteúdo contido na página número 10
6081-P Instruction Sheet October 2013 LIQ_MAN_ABR_6081-P Mounting Mounting on a Flat Surface Figure 4. Wall Mounting Installation for 6081. Use Pipe/Wall Mounting Bracket Kit, PN23820-00 Note: PN 23820-00 mounting bracket kit includes mounting hardware for pipe mounting only. Wall mounting hardware to be pro- vided by customer. Only use suitable fasteners and hardware to securely fasten the bracket and transmitter to the wall surface 10
Resumo do conteúdo contido na página número 11
Instruction Sheet 6081-P LIQ_MAN_ABR_6081-P October 2013 Pipe Mounting The pipe mounting kit (PN 23820-00/01) accommodates 1–½ in. pipe Figure 5. Pipe Mounting Installation for 6081. Use Pipe/Wall Mounting Bracket Kit, PN23820-00 11
Resumo do conteúdo contido na página número 12
6081-P Instruction Sheet October 2013 LIQ_MAN_ABR_6081-P Figure 6. FM IS Installation D 1700672 12 8 7 6 3 5 4 2 1 This document contains information proprietary to REVISION Rosemount Analytical, and is not to be made available to those who may compete with Rosemount Analytical. THIS DOCUMENT IS LTR ECO DESCRIPTION BY DATE CHECKED/APPROVED APPROVED MODELS CERTIFIED BY D LQD10215 SEE ECO CH 5-18-11 JP/ SP NOTES: UNLESS OTHERWISE SPECIFIED MARK
Resumo do conteúdo contido na página número 13
Instruction Sheet 6081-P LIQ_MAN_ABR_6081-P October 2013 Figure 7. CSA IS Installation D 1700673 13 8 7 6 5 3 4 2 1 This document contains information proprietary to REVISION Rosemount Analytical, and is not to be made available to those who may compete with Rosemount Analytical. THIS DOCUMENT IS LTR ECO DESCRIPTION BY DATE CHECKED/APPROVED APPROVED MODELS CERTIFIED BY D LQD10215 SEE ECO CH 5-18-11 J
Resumo do conteúdo contido na página número 14
6081-P Instruction Sheet October 2013 LIQ_MAN_ABR_6081-P Figure 8. ATEX IS Installation D 1700674 14 8 7 6 3 5 4 2 1 This document contains information proprietary to REVISION Rosemount Analytical, and is not to be made available to those who may compete with Rosemount Analytical. LTR ECO DESCRIPTION BY DATE CHECKED/APPROVED THIS DOCUMENT IS APPROVED MODELS CERTIFIED BY NOTES: UNLESS OTHERWISE SPECIFIED 6081-P-73-XXX REV 1. PCB MATERIAL
Resumo do conteúdo contido na página número 15
Instruction Sheet 6081-P LIQ_MAN_ABR_6081-P October 2013 15
Resumo do conteúdo contido na página número 16
6081-P Instruction Sheet October 2013 LIQ_MAN_ABR_6081-P facebook.com/EmersonRosemountAnalytical AnalyticExpert.com Credit Cards for U.S. Purchases Only. twitter.com/RAIhome youtube.com/user/RosemountAnalytical ©2013 Rosemount Analytical, Inc. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their