MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE
(Addendum to GFC7000E Manual PN 04584)
MODEL GFC7000EU
CO ANALYZER
2
(Addendum to GFC7000E Manual PN 04584)
Teledyne Analytical Instruments
16830 Chestnut St.
City of Industry, Ca. 91748
TEL: 626-961-9221
FAX: 626-961-2538
TEL: 626-934-1500
E-MAIL: ai-customerservice@teledyne.com
WEB SITE: http://www.teledyne-ai.com
1
05134 Rev -
Streszczenie treści zawartej na stronie nr. 2
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) OVERVIEW The model GFC7000EU is a close derivative of the Model GFC7000E CO Analyzer. The optical bench is longer (14 m), making the instrument more sensitive at low CO 2 levels. The instrument has a user selectable full scale range of 0-100 PPB to 0-100 PPM of CO . 2 The instrument does not come equipped with a sample flow rate sensor. Hence, disregard any reference to the flow sensor in the users manual.
Streszczenie treści zawartej na stronie nr. 3
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) Humidity Range 0-95% RH, Non-Condensing (6) Temp Coefficient < 0.05 % of reading per °C (5 ppb/°C minimum) Voltage Coefficient < 0.05 % of reading per V Dimensions 7" x 17" x 23.5" (178 mm x 432 mm x 597 mm) (HxWxD) Weight 50 lb (22.7 kg) AC Power 100V 50/60 Hz (3.25A), 115 V 60 Hz (3.0A), 220 – 240 V 50/60 Hz (2.5A) Environmental Installation Category (Over voltage Category) II Conditions Pollution Degree 2
Streszczenie treści zawartej na stronie nr. 4
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) There are 4 relay alarm outputs (AL1-AL4) on the rear panel. AL1 is for system okay, AL2 is for concentration limit 1 exceeded and AL3 is for concentration limit 2 exceeded. The relay alarm output AL1 is enabled all the time, whereas the AL2 and AL3 can be enabled/disabled by going into the diagnostics menu (with 929 password), then factory options, then turning conc. Alarm Relays ON or OFF. When the concent
Streszczenie treści zawartej na stronie nr. 5
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) SAMPLE WHEEL TEMP WARNING CO = XX.XX TEST deactivates Warning Messages until New warning(s) TEST CAL MSG CLR SETUP are activated MSG activates Warning SAMPLE RANGE=50.0 PPM CO = XX.XX Messages. keys replaced with < TST TST > CAL MSG CLR SETUP TEST key All Warning messages are hidden, but MSG button appears SAMPLE WH
Streszczenie treści zawartej na stronie nr. 6
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) Warning Fault Condition Possible Causes Message FRONT PANEL The CPU is unable to WARNING only appears on Serial I/O COM Port(s) WARN Communicate with Front Panel Display will be frozen, blank or will not the Front Panel respond. Display /Keyboard Failed Keyboard 2 I C Bus failure Loose Connector/Wiring Failed IR Photo-Detector PHOTO TEMP PHT DRIVE is Failed Sync/Demod Board WARNING >4750 mVDC or
Streszczenie treści zawartej na stronie nr. 7
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) Warning Fault Condition Possible Causes Message on wheel rotation. Failed Sync/Demod Board If Status LED’s on the Sync/Demod Board ARE flashing the cause is most likely an a problem with the analyzer’s main power: Intermittent loss of the power supply too short to cause a SYSTEM RESET warning; Frequency problem with the AC Mains. SYSTEM RESET The computer has This message occurs at power on. rebooted.
Streszczenie treści zawartej na stronie nr. 8
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) #N05ATI 009690000 Filter, TFE, 47 mm 5 um, Qty 100 009690100 Filter, TFE, 47 mm 5 um, Qty 25 009820200 Assy, Motor, Synchronous, 115V/60HZ, "E" 010790000 Input Mirror 010800000 Output Mirror 016290000 Window, Sample Filter, 47 mm 016910000 Akit, Catalyst, CO/CO2 Converter (1 OZ) 019340000 Sample Thermistor M400 (S/N >099) or M401 (S/N>0573) 024710000 Tubing: 6', 1/8"CLR 024720000 Tubing: 6', 1/8" BLK 024750
Streszczenie treści zawartej na stronie nr. 9
MODEL GFC7000EU CO ANALYZER WITH AUTO-REFERENCE (Addendum to GFC7000E Manual PN 04584) OR0000001 O-Ring, Flow Control OR0000058 O-Ring, Sample Filter, 47 mm, 2-228V OR0000034 O-Ring, Input/Output Mirror/Detector OR0000039 O-Ring, Window OR0000041 O-Ring, Field and Objective Mirror 043590100 Pump, 115V 50/60 Hz PU0000022 Pump Rebuild Kit, KNF Model #N05ATI PS0000025 Power Supply, +12V PS0000011 Power Supply, +5/+- 15V 042680000 Solenoid, Stainless Steel, 3-Way, 12V 042690000 Solenoid, Stainless