Streszczenie treści zawartej na stronie nr. 1
Manual Addendum
Model T360U
CO Analyzer
2
(Addendum to T360 Manual, PN 07272)
© TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TAPI)
9480 CARROLL PARK DRIVE
SAN DIEGO, CA 92121-5201
USA
Toll-free Phone: 800-324-5190
Phone: 858-657-9800
Fax: 858-657-9816
Email: api-sales@teledyne.com
Website: http://www.teledyne-api.com/
Copyright 2013 07273A DCN6707
Teledyne Advanced Pollution Instrumentation 09 July 2013
Streszczenie treści zawartej na stronie nr. 2
Streszczenie treści zawartej na stronie nr. 3
ABOUT THIS ADDENDUM This addendum, part number 07273, is to be used in conjunction with the Model T360 operation manual, part number 07272. Where operation of the Model T360U diverges from that of the Model T360, this addendum takes precedence. Please note that the T360 operation manual contains important SAFETY messages and the WARRANTY information for this instrument. It is strongly recommended that you read the T360 operation manual in its entirety as well as this addendum, before ope
Streszczenie treści zawartej na stronie nr. 4
This page intentionally left blank ii
Streszczenie treści zawartej na stronie nr. 5
TABLE OF CONTENTS About This Addendum ................................................................................................................ i Revision History.......................................................................................................................... i Table of Contents ......................................................................................................................iii 1. Overview ................................................
Streszczenie treści zawartej na stronie nr. 6
This page intentionally left blank. iv
Streszczenie treści zawartej na stronie nr. 7
1. OVERVIEW The Model T360U is a close derivative of the Model T360 CO Analyzer. The main differences are as follows: The optical bench is longer (14 m), making the instrument more sensitive at low CO 2 levels. The instrument has a user-selectable full scale range of 0-100 PPB to 0-100 PPM of CO . 2 The rear panel has an extra pneumatic port for the exhaust of the purge gas to the GFC wheel. Because the T360U is a close derivative of T360, the Model T360 Operation Manual cont
Streszczenie treści zawartej na stronie nr. 8
PARAMETER SPECIFICATION AC Power 100 – 120V 50/60 Hz (120W) 220 – 240 V 50/60 Hz (144W) Environmental Installation Category (Over voltage Category) II Pollution Degree 2 Conditions 1 Ethernet: 10/100Base-T Standard I/O 2 RS-232 (300 – 115,200 baud) 2 USB device ports 8 opto-isolated digital status outputs 6 opto-isolated digital control inputs 4 analog outputs 1 USB com port Optional I/O 1 RS485 8 analog inputs (0-10V, 12-bit) 4 digital alarm outputs Multidrop RS232 3 4-20m
Streszczenie treści zawartej na stronie nr. 9
4. PNEUMATIC CONNECTIONS This section provides information on the pneumatic configurations. Figure 4-1 shows the basic pneumatic flow of the T360U. Figure 4-1. T360U Pneumatic Flow Diagram 7
Streszczenie treści zawartej na stronie nr. 10
4.1. BASIC AND ZERO/SPAN VALVE OPTION CONFIGURATIONS CAUTION Sample and calibration gases should only come into contact with PTFE (Teflon), FEP, glass, stainless steel or brass. Do NOT operate without first removing dust plugs from rear panel fixtures. NOTE To prevent dust from entering the gas flow channels, your analyzer was shipped with small plugs inserted into each of the pneumatic fittings on the back panel. Remove these dust plugs and store for future use before proceeding.
Streszczenie treści zawartej na stronie nr. 11
Pressure regulator, set Needle valve, Sample output to 5-10 set flow to 800 Sample psig cc/min (+/- 10%) gas Chassis Exhaust Flowmeter Vent Span Needle valve, set flow to 800 Pressure cc/min (+/- Regulator (5-10 Pressure Span Span gas 10%) psig output) Needle valve, Zero Pressure set flow to 800 CO2 free Regular (5-10 gas cc/min (+/- psig output) 10%) From Purge To Purge 10-30 psig CO2 free gas Figure 4-3. Pneumatic Connections–Configuration with Zero/Span
Streszczenie treści zawartej na stronie nr. 12
4.2. MAKING THE PNEUMATIC CONNECTIONS CAUTION Venting should be outside the shelter or immediate area surrounding the instrument. 1. Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more than 2 meters long. 2. Attach sources of zero air and span gas 3. Span Gas is a gas specifically mixed to match the chemical composition of the type of gas being measured at near full scale of the desired measurement range. In the case of CO measureme
Streszczenie treści zawartej na stronie nr. 13
5. RELAY ALARM OUTPUTS (STANDARD CONFIGURATION) There are 4 relay alarm outputs (AL1-AL4) on the rear panel. AL1 is for system okay, AL2 is for concentration limit 1 exceeded and AL3 is for concentration limit 2 exceeded. AL4 is not used The relay alarm output AL1 is enabled all the time, whereas the AL2 and AL3 can be enabled/disabled by going into the diagnostics menu (with 929 password), then factory options, then turning conc. Alarm Relays ON or OFF. When the concentr
Streszczenie treści zawartej na stronie nr. 14
This page intentionally left blank. 12
Streszczenie treści zawartej na stronie nr. 15
T360U SPARE PARTS LIST Ref PN 07460, 05 Apr 2013, 14:24 PARTNUMBER DESCRIPTION 037600000 AKIT, EXP KIT, SODA LIME CH47 037860000 ORING, TEFLON, RETAINING RING, 47MM (KB) 039060100 ASSY, PRESSURE TRANSDUCER 039260501 DETECTOR, CO2, w/BANDPASS FILTER* 040010000 ASSY, FAN REAR PANEL (B/F) 040030300 PCA, PRESS SENSORS w/SW49 040370000 ASSY, CO SCRUBBER 041350000 PCA, RELAY BOARD, CO(KB) 042680000 ASSY, VALVE (SS) 042690000 ASSY, VALVE , 2-WAY, 12V 042890100 ASSY, PUMP CONF
Streszczenie treści zawartej na stronie nr. 16
T360U SPARE PARTS LIST Ref PN 07460, 05 Apr 2013, 14:24 009450300 ASSY, ZERO/SPAN VALVES, CO 006110200 ASSY, MOTOR WHEEL HEATER 003291500 ASSY, THERMISTOR, BENCH/WHEEL 003290500 ASSY, THERMISTOR, BENCH 001761300 ASSY, FLOW CTRL, .010, 1/8", SS 001760400 ASSY, FLOW CTL, 800CC, 1/4" CONN-B 000941000 CD, ORIFICE, .013 BLUE/GREEN (KB) 000940600 CD, ORIFICE, .010 BROWN (KB) 077490300 ASSY, PMP, INT, 100V, 9 OC CN0000073 POWER ENTRY, 120/60 (KB) CN0000458 PLUG, 12, MC 1.5/
Streszczenie treści zawartej na stronie nr. 17
Interconnect List, T360U Reference: 073640100A FROM TO Cable PN Signal Assembly PN J/P Pin Assembly PN J/P Pin 03995 CBL, MOTOR TO RELAY PCA GFC Drive - A Relay PCA 041350000 J6 1 GFC Motor 052380200 P1 1 GFC Drive - B Relay PCA 041350000 J6 2 GFC Motor 052380200 P1 2 Motor Return Relay PCA 041350000 J6 3 GFC Motor 052380200 P1 3 Chassis Gnd Relay PCA 041350000 J6 4 GFC Motor 052380200 P1 4 040230100CBL, I2C, RELAY BOARD TO MOTHERBOARD I2C Serial Clock Motherboard 058021100 P107 3 Relay PCA 0413
Streszczenie treści zawartej na stronie nr. 18
Interconnect List, T360U Reference: 073640100A FROM TO Cable PN Signal Assembly PN J/P Pin Assembly PN J/P Pin 06746 CBL, MB TO 06154 CPU GND Motherboard 058021100 P12 2 Shield RX0 Motherboard 058021100 P12 14 CPU PCA 067240000 COM1 1 RTS0 Motherboard 058021100 P12 13 CPU PCA 067240000 COM1 8 TX0 Motherboard 058021100 P12 12 CPU PCA 067240000 COM1 4 CTS0 Motherboard 058021100 P12 11 CPU PCA 067240000 COM1 7 RS-GND0 Motherboard 058021100 P12 10 CPU PCA 067240000 COM1 6 RTS1 Motherboard 058021100
Streszczenie treści zawartej na stronie nr. 19
Interconnect List, T360U Reference: 073640100A FROM TO Cable PN Signal Assembly PN J/P Pin Assembly PN J/P Pin 06814 CBL ASSY, SYNC DEMOD , DC POWER DGND Relay PCA 041350000 J15 1 Sync Demod 032960000 J2 1 +5V Relay PCA 041350000 J15 2 Sync Demod 032960000 J2 2 AGND Relay PCA 041350000 J15 3 Sync Demod 032960000 J2 3 +15V Relay PCA 041350000 J15 4 Sync Demod 032960000 J2 4 AGND Relay PCA 041350000 J15 5 Sync Demod 032960000 J2 5 -15V Relay PCA 041350000 J15 6 Sync Demod 032960000 J2 6 +12V ret R
Streszczenie treści zawartej na stronie nr. 20
This page intentionally left blank. 07273A DCN6707 18