Summary of the content on the page No. 1
Agilent E5070BU/E5071BU Option 018
Hard Disk Drive Kit
User’s Guide
Second Edition
Agilent Part No. E5070-90119
June 2005
Printed in Japan
Summary of the content on the page No. 2
Notices The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies. © Copyright 2005 Agilent Technologies Manual Printing History The manual’s printing date and part number indicate its current edition. The printing date
Summary of the content on the page No. 3
1 General Information This chapter describes the Kit, contents of the kit, and required tools to install the Option 018 into the Agilent E5070B/E5071B. 3
Summary of the content on the page No. 4
General Information Option 018 Hard Disk Drive Kit Option 018 Hard Disk Drive Kit The purpose of this kit is to install the option 018 Hard Disk Drive into the Agilent E5070B/E5071B. Contents of the Agilent E5070BU/E5071BU Option 018 Hard Disk Drive Kit The kit includes the parts listed in Table 1-1. If anything is missing, contact the nearest Agilent Technologies office. Table 1-1 Contents of the Agilent E5070BU/E5071BU Option 018 Hard Disk Drive Kit Description Qty. Upgrade Manual (This ma
Summary of the content on the page No. 5
2 Option Installation This chapter describes the procedure to install the option 018 into the Agilent E5070B/E5071B. 5
Summary of the content on the page No. 6
Option Installation Introduction Introduction This chapter describes the procedure to install Option 018 into the E5070B/E5071B. The procedure consists of the following items. • Installation Requirement • Removing Removable Hard Disk • Mounting Removable Hard Disk • Writing System Correction Data of Backup Memory to Removable HDD • Calibration of the Touch Screen • Caution for replacing the spare HDD with the original HDD Update Procedure Installation Requirement • TORX Screwdriver T10 Removing
Summary of the content on the page No. 7
Option Installation Update Procedure Figure 2-1 Mounting/Removing Removable Hard Disk -1 Figure 2-2 Mounting/Removing Removable Hard Disk -2 Chapter 2 7
Summary of the content on the page No. 8
Option Installation Update Procedure Writing System Correction Data of Backup Memory to Removable HDD The following is a description of the procedures to write the system correction data stored in the backup memory to the mounted removable HDD. Step 1. Turn on the E5070B/E5071B. Step 2. Press . Step 3. Press Load Project... on the softkey menu. Step 4. A dialog box appears to allow you to select a program to be loaded. Select RestoreSysCorFile.vba from the D:\Agilent\Service folder, and then
Summary of the content on the page No. 9
Option Installation Update Procedure Initial Registration of E5070B/E5071B When you start up the E5070B/E5071B for the first time or after executing system recovery, you need to perform the initial registration of the Windows 2000 operating system of the E5070B/E5071B. NOTE You cannot use the front panel keys during the initial registration of the E5070B/E5071B therefore connect the mouse and the keyboard before turning on the power. NOTE If you perform the following procedure incorrectly, a
Summary of the content on the page No. 10
Option Installation Update Procedure Step 3. In the Windows 2000 Professional Setup dialog box, select the I accept this agreement box and click the Next >button (Figure 2-5). Then, the Windows 2000 operating system is restarted automatically. Figure 2-5 Windows 2000 Professional Setup dialog box Step 4. The Network Identification wizard appears. Click the Next > button (Figure 2-6). Figure 2-6 Network Identification wizard 10 Chapter 2
Summary of the content on the page No. 11
Option Installation Update Procedure Step 5. In the Network Identification Wizard dialog box (1/2), select the Windows always assumes the following user has logged on to this computer box and check that agena is in the User Name box. If not, type in agena. Finally, click the Next> button (Figure 2-7). Figure 2-7 Network Identification Wizard dialog box (1/2) Step 6. In the Network Identification Wizard dialog box (2/2), click the Finish button to finish the initial registration of the E5070B/
Summary of the content on the page No. 12
Option Installation Update Procedure Calibration of the Touch Screen When you have replaced the HDD on an E5070B/E5071B equipped with an Option 016 touch screen, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen. Step 1.Press . Step 2. Press Service Menu. Step 3. Press Test Menu. Step 4. Press Adjust Touch Screen. The touch screen calibration screen (Figure 2-9) appears. Figure 2-9 Touch Panel Calibration Screen Step 5. Touch the x mar