Energizer PORTABLE GAS DETECTION INSTRUMENT user manual

User manual for the device Energizer PORTABLE GAS DETECTION INSTRUMENT

Device: Energizer PORTABLE GAS DETECTION INSTRUMENT
Category: Smoke Alarm
Manufacturer: Energizer
Size: 0.37 MB
Added : 10/6/2014
Number of pages: 50
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Energizer PORTABLE GAS DETECTION INSTRUMENT. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Energizer PORTABLE GAS DETECTION INSTRUMENT.

For your convenience

If looking through the Energizer PORTABLE GAS DETECTION INSTRUMENT user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Energizer PORTABLE GAS DETECTION INSTRUMENT on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Energizer PORTABLE GAS DETECTION INSTRUMENT to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Energizer PORTABLE GAS DETECTION INSTRUMENT User manual - Online PDF
Advertisement
« Page 1 of 50 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Energizer PORTABLE GAS DETECTION INSTRUMENT but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Energizer PORTABLE GAS DETECTION INSTRUMENT. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1

MINI-SA PORTABLE
GAS DETECTION
INSTRUMENT
1

Summary of the content on the page No. 2

INTRINSICALLY SAFE GAS LEVEL DETECTOR R C US 39NO CERTIFIED BY UNDERWRITERS LABORATORIES INC. ONLY AS TO INTRINSIC SAFETY FOR USE IN HAZARDOUS LOCATIONS CLASS I, DIV, 1, GROUPS A, B, C, & D. CERTIFIED TO CANADIAN CSA STANDARD C22.2 NO. 157-92 WARNING: SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY. USE ONLY WITH APPROVED BATTERIES. TEMPERATURE CODE T3C. CAUTION: FOR SAFETY REASONS, THIS EQUIPMENT MUST BE OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY. READ AND UNDERSTAND MANUAL BEFORE

Summary of the content on the page No. 3

SCOTT INSTRUMENTS MINI-SA PORTABLE GAS DETECTION INSTRUMENT Contents 1.0 INTRODUCTION................................................................. 4 2.0 OPERATIONAL ENVIRONMENT ........................................ 4 3.0 GENERAL INSTRUCTIONS ............................................... 5 3.1 UNPACKING .......................................................... 5 3.2 REGISTRATION ..................................................... 5 3.3 BATTERY INSTALLATION ............................

Summary of the content on the page No. 4

SCOTT INSTRUMENTS MINI-SA PORTABLE GAS DETECTOR 1.0 INTRODUCTION The SCOTT INSTRUMENTS MINI-SA Portable Gas Detector is a self contained instrument designed to detect and provide audible, visual, and tactile indication of the presence of oxygen or toxic gas in the air. The toxic gas concentration is monitored and shown in the graphic display in parts per million (ppm) of the atmosphere. The oxygen concentration is shown in the graphic display as a percent (%). When the gas level exceeds the pr

Summary of the content on the page No. 5

3.0 GENERAL INSTRUCTIONS 3.1 UNPACKING Remove the SCOTT INSTRUMENTS MINI-SA instrument from its packaging and examine it for external damage. If any damage is found, notify your SCOTT INSTRUMENTS distributor, sales representative, or the SCOTT IN- STRUMENTS Service Department. 3.2 REGISTRATION Fill out and mail the postage paid Registration Card provided. Registering your SCOTT INSTRUMENTS MINI-SA instrument assures that you will re- ceive notification of any future enhancements and/or upgrades

Summary of the content on the page No. 6

4.0 INSTRUMENT DESCRIPTION THE SCOTT INSTRUMENTS MINI-SA INSTRUMENT is designed for use as a personal monitor. The MINI-SA is equipped with a snap mounting clip for ease of use. A break-away neck strap and rubber boot are also available. All operations of the instrument are controlled by two push-button actuators which are large enough to accommodate use with gloved hands. In each menu screen, “PROMPTS” over each push-button define their use. The SCOTT INSTRUMENTS MINI-SA INSTRUMENT case is cons

Summary of the content on the page No. 7

4.1 ALARM LEGENDS The following alarm legends appear in the graphic display: “ALARM” flashing legend indicates one of the following: a) Gas concentration which exceeds the preset level b) A Time Weighted Average (TWA) Alarm c) A Short Term Exposure Level (STEL) Alarm. For toxic gas, the ppm, TWA, or STEL level detected and the “ALARM” legend will flash alternately. A graphic battery level indicator appears continuously on the display screen. The indicator shows battery capacity as ten (10) lines

Summary of the content on the page No. 8

4.2 AUDIBLE ALARMS The SCOTT INSTRUMENTS MINI-SA has an audible alarm which is acti- vated when any of the following conditions occurs: - The preset gas alarm level is exceeded (ALARM) - The Time Weighted Average (TWA) is exceeded - The Short Term Exposure Limit (STEL) is exceeded - The battery is exhausted - The sensor is missing or incompatible In addition, each time the instrument is turned ON, the audible alarm will pulse three times. The following describes the function of the Audible Alar

Summary of the content on the page No. 9

4.3 VIBRATION ALARM An internal vibrating motor activates whenever the Audible Alarm sounds. The Vibration Alarm operates as a secondary alarm feature in high noise environ- ments where the audible alarm may be difficult to hear. The Vibration Alarm is reset the same as the Audible Alarm as described in Section 4.2. The Vibration Alarm will also pulse briefly whenever the MINI-SA instrument is turned ON or OFF. WARNING THE VIBRATION ALARM DOES NOT FUNCTION FOR THE LOW BAT- TERY INDICATION. 4.4

Summary of the content on the page No. 10

5.0 SWITCH FUNCTIONS AND DISPLAYS All operations on the MINI-SA Portable Detector are controlled by two push- buttons located on the bottom of the unit. The operations vary depending on the operating mode and the graphic display appearing above the two switches. The unit sounds a brief audible “beep” whenever either pushbutton is de- pressed while the instrument is on. Whenever the instrument is turned ON or OFF, three successive audible tones will sound and the Red Alarm LED will flash. The Gre

Summary of the content on the page No. 11

6.0 OPERATING INSTRUCTIONS Before using this instrument, read and understand ALL instructions for ALL operating modes described in this Section. Prior to use, the instrument should be checked for proper calibration as described in this Section. WARNING BEFORE EACH DAY’S USAGE, THE SENSITIVITY OF THE GAS SENSOR MUST BE TESTED ON A KNOWN CONCENTRATION OF CALIBRATION GAS. WARNING ACTUAL CALIBRATION SHOULD ONLY BE PERFORMED BY QUALI- FIED INSTRUMENT MAINTENANCE PERSONNEL. WARNING ANY INDICATION OR R

Summary of the content on the page No. 12

6.1 INSTRUMENT TURN ON / OFF To turn the instrument ON, depress either pushbutton. Three successive “beeps” will sound. The Red LED will flash and the vibration alarm will pulse for each “beep.” The graphic display will show one of the following screens while the instrument initializes. The sensor installed in the unit defines which display is shown: SCOTT INSTRUMENTS For Toxic Gases MINI-SA VERSION X.X

Summary of the content on the page No. 13

sensor type has been changed or if the previous calibration was not suc- cessful, the bottom line of the display will show “CALIBRATION REQUIRED”. ppm For Toxic Gases 0 MENU For O 20.9% O 2 2 MENU In the normal Operating Mode, the following

Summary of the content on the page No. 14

6.2 MENU SELECTION From the Operating Mode screen, there are six menus that can be selected by the user. They are “ZERO CALIBRATION,” “SPAN CALIBRATION,” “PEAK/ HOLD,” “ALARM SETPOINT,” “TWA/STEL,” and “CONFIGURATION.” De- press the pushbutton below the MENU prompt on the display and then one of the following screens will appear: ZERO CALIBRATION For Toxic Gases NEXT AUTO ZERO AUTO CALIBRATION For O 2 NEXT CALIBRATE The AUTO ZERO or CALIBRATE command and “NEX

Summary of the content on the page No. 15

PEAK / HOLD NEXT SELECT To go to the PEAK/HOLD menu, depress the “SELECT” pushbutton. To go to the next menu, depress the left pushbutton below the “NEXT” com- mand and one of the following screens will be displayed: ALARM SETPOINT For Toxic Gases NEXT SELECT O ALARM SETPOINT 2 For O 2 NEXT SELECT To go to the ALARM SETPOINT menu, depress the “SELECT” pushbutton. To go to the next menu, depress the left pushbutton below the “NEXT” c

Summary of the content on the page No. 16

CONFIGURATION NEXT SELECT To go to the CONFIGURATION menu, press the “SELECT” pushbutton. To go to the next menu, depress the left pushbutton below the “NEXT” com- mand and the ZERO CALIBRATION for toxic gas or AUTO CALIBRATION for oxygen screen will again be displayed. The MINI-SA will scroll repeatedly through these Menus each time the “NEXT” pushbutton is depressed. Each time either pushbutton is depressed, the unit will sound a single “beep”. If no menu action is taken for

Summary of the content on the page No. 17

6.3 ZERO CALIBRATION MENU The MINI-SA instrument can be Auto Zeroed at any time. An Auto Zero should only be performed in an area known to be free of toxic gas or any known cross sensitive materials. See Section 8.0 for SCOTT INSTRUMENTS sup- plied calibration gas cylinders. When performing an auto-zero with an oxy- gen unit, be sure the area contains 20.9% oxygen. To perform Zero Calibration of the sensor as described in Section 6.2, de- press the “AUTO ZERO” pushbutton when the ZERO CALIBRATIO

Summary of the content on the page No. 18

6.4 SPAN CALIBRATION MENU FOR TOXIC GAS ONLY Note: The MINI-SA software requires that an “AUTO ZERO” be performed as described in Section 6.3 before the instrument will allow a “SPAN CALIBRA- TION”. Failure to do so will display the following screen when attempting a “SPAN CALIBRATION” of the instrument: AUTO ZERO REQUIRED BEFORE SPAN CALIBRATION ACKNOWLEDGE Depress both pushbuttons simultaneously to acknowledge this message and “AUTO ZERO” the instrument pe

Summary of the content on the page No. 19

Calibration cylinders supplied by SCOTT INSTRUMENTS contain 25 PPM H S or 50 PPM CO. To use another certified concentration, depress the left 2 pushbutton below the “CHANGE” command. The one of the following screens will be displayed: SENSOR For Toxic Gases SPAN CALIBRATION GAS = <_> ppm UP DOWN Factory preset span calibration gas values: CO = 50 ppm Cl = 5.0 ppm 2 ClO = 5.0 ppm Cl 2 2 HCN = 10.0 ppm H S = 25 ppm 2 NH = 25 ppm 3 NO = 5.0 ppm

Summary of the content on the page No. 20

This screen will remain on until the instrument detects calibration gas. If no calibration gas is detected after thirty (30) seconds, the instrument will sound a single beep and return to the operating mode. Depressing the “EXIT” pushbutton will also return the instrument to normal operation with no calibration having occurred. The last successful calibration will remain valid. To continue “SPAN CALIBRATION”, attach the 40014936 Calibration Adapter over the sensor cap as shown in Figure 4. TO


Alternative user manuals
# User manual Category Download
1 Energizer PORTABLE GAS DETECTION INSTRUMENT User manual Smoke detectors 0
2 Adobe CS3 User manual Smoke Alarm 1
3 ADT Security Services ADT-LCD40 User manual Smoke Alarm 6
4 ADT Security Services Safewatch Pro 2000 User manual Smoke Alarm 5
5 ADT Security Services SAVS20ADT-1 User manual Smoke Alarm 3
6 ADT Security Services Unimode 9600 User manual Smoke Alarm 7
7 ADT Security Services SAVS20ADTG User manual Smoke Alarm 4
8 Aico Ei166RC User manual Smoke Alarm 12
9 Aico Carbon Monoxide Alarm Ei 225EN User manual Smoke Alarm 40
10 Aico Ei161RC User manual Smoke Alarm 5
11 Aico Ei169RF User manual Smoke Alarm 1
12 Aico Ei168RC User manual Smoke Alarm 14
13 Aico Carbon Monoxide Alarms 260 Series User manual Smoke Alarm 0
14 Aico Carbon Monoxide Alarm Ei 220EN User manual Smoke Alarm 0
15 Aico Ei164RC User manual Smoke Alarm 5