TABLE OF CONTENTS CONTENTS 1.0 IMPORTANT INSTRUCTIONS 2.0 GENERAL WARNINGS AND PRECAUTIONS 3.0 MODEL 900EC 4.0 OPERATING THE CHAIR 4.1 OPERATION OF STEER/BRAKE 4.2 USING THE 4 - CASTER FREE FUNCTION 4.3 APPLYING THE BRAKES 4.4 ADJUSTING THE ARMREST 4.5 RAISING THE SURFACE 4.6 I.V. SOCKETS AND STORAGE 4.7 HANDSET OPERATION 5.0 DIAGNOSTICS & REPAIR 6.0 CLEANING & MAINTENANCE
Summary of the content on the page No. 3
1.0 IMPORTANT INSTRUCTIONS Safety First: The primary concern of ConvaQuip is that this equipment be operated and maintained with the safety of the user in mind. To assure safe and reliable operation. 1. Read this manual before operating your equipment. 2. It is the responsibility of the purchaser to assure that appropriate personnel are in formed on the contents of this manual. 3. Be sure that you understand the instructions contained in this manual before at tempting to ope
Summary of the content on the page No. 4
10. Take special care to lock the casters securely when transferring patient to or from the chair. 11. Do not allow the patient to sit on the chair when it is in the prone position. Never sit or lean on the headrest or the legrest when the chair is in the prone position. Caution: Failure to observe these safety precautions may result in personal injury or mechanical failure. Read all instructions carefully before using the chair. . Weight Limit: Never exceed the weig
Summary of the content on the page No. 5
3.0 MODEL 900EC The backrest is adjustable from 0 - 60°. This chair is equipped with central steering and braking system operable from any corner of the chair. This facilitates patient transport. High density foam seat, leg and back cushions are covered with low shear fabric. 4.0 OPERATING THE CHAIR Operation of all features and functions prior to usage is essential to ensure that chair was not damaged in transit. 4.1 OPERATION OF STEERING/BRAKING The steering l
Summary of the content on the page No. 6
4.3 APPLYING THE BRAKES The central braking system locks all four wheels. To engage the brakes, go to any corner of the chassis and press the red side of pedal down approximately 45° to lock. After applying the brakes, try to lightly push the chair, it should not roll. If it does, remove the patient and discontinue using the chair until maintenance can be performed. WARNING Failure to make sure the chair does not roll with the brakes engaged could result in p
Summary of the content on the page No. 7
4.5 To either raise or lower the chair it must be in the prone position. The backrest must be all the way down and both legrest must be all the way up. The bottom row of buttons provide this feature with pushing one button. 4.6 The IV sockets are located on each side of the head end. The IV storage is located under the seat on the left side. 4.7 Handset Operation 1st row will raise and lower unit when in the prone position 2nd row will raise and lower the backrest 3rd row will rai
Summary of the content on the page No. 8
5.0 DIAGNOSTICS & REPAIR PROBLEM CAUSE SOLUTION No functions operate No power to unit Check power cord Is it plugged into receptacle? Green Power-on light Electrical power cord not does not glow properly plugged into Check cord for breaks. control module Power cord plugged Check for power to receptacle in but no functions operate Bad handset Replace handset Bad control module Replace control module 6.0 CLEANING & MAINTENANCE The chair i
Summary of the content on the page No. 9
7.0 SPECIFICATIONS Product name: ConvaQuip EC TRANS. CHAIR Model: 900EC 850 LB. Capacity Dimensions: Overall Width 33 1/2” Overall Length chair position 49 1/2” stretcher position 75 1/2” Seat Width 28 1/2” Seat Depth 22” Back Angle 60 Degrees Chair Weight 280 LB
Summary of the content on the page No. 10
LIMITED WARRANTY ConvaQuip will warrant your chair to be free from defects in material and workmanship under normal use and service for a period of two (2) years on parts & labor from date of the delivery to the original purchaser, except: the effective time period of this warranty is limited to one (1) year for cushions, wheels, tires, IV pole and accessories. Defective parts will be repaired or replaced by the company or its appointed agent. The warranty does not cover damage caused b