Summary of the content on the page No. 1
MP1763C
Pulse Pattern Generator
Operation Manual
(Function/Operation)
Ninth Edition
For safety and warning information, please read this
manual before attempting to use the equipment.
Keep this manual with the equipment.
ANRITSU CORPORATION
Document No.: M-W1848AE-9.0
Summary of the content on the page No. 2
Safety Symbols To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Corporation uses the following safety symbols to indicate safety-related information. Ensure that you clearly understand the meanings of the symbols BEFORE using the equipment. Some or all of the following symbols may be used on all Anritsu equipment. In addition, there may be other labels attached to products that are not shown in the diagrams in this manual. Symbols used in manual This in
Summary of the content on the page No. 3
For Safety WARNING 1. ALWAYS refer to the operation manual when working near locations at which the alert mark shown on the left is attached. If the advice in the operation manual is not followed there is a risk of personal injury or reduced equipment performance. The alert mark shown on the left may also be used with other marks and descriptions to indicate other dangers. 2. IEC 61010 Standard The IEC 61010 standard specifies four categories to ensure that an instrument is used only at locat
Summary of the content on the page No. 4
For Safety WARNING 4. This equipment cannot be repaired by the operator. DO NOT attempt Repair to remove the equipment covers or unit covers or to disassemble internal components. Only qualified service personnel with a knowledge of electrical fire and shock hazards should service this equipment. There are high-voltage parts in this equipment presenting a risk of severe injury or fatal electric shock to untrained personnel. In addition, there is a risk of damage to precision components. 5.
Summary of the content on the page No. 5
For Safety CAUTION 1. Always remove the mains power cable from the power outlet before replacing blown fuses. There is a risk of electric shock if fuses are Fuse Replacement replaced with the power cable connected. Always use new fuses of the type and rating specified on the rear panel of the instrument. There is a risk of fire if a fuse of a different rating is used. T6.3A indicates a time-lag fuse. 2. Keep the power supply and cooling fan free of dust. Clean the power inlet regularly. If d
Summary of the content on the page No. 6
For Safety CAUTION Replacing Memory This equipment uses a Poly-carbomonofluoride lithium battery to backup Back-up Battery the memory. This battery must be replaced by service personnel when it has reached the end of its useful life; contact the Anritsu sales section or your nearest representative. Note: The battery used in this equipment has a maximum useful life of 7 years. It should be replaced before this period has elapsed. External This equipment uses memory cards as external storage me
Summary of the content on the page No. 7
Equipment Certificate Anritsu Corporation certifies that this equipment was tested before shipment using calibrated measuring instruments with direct traceability to public testing organizations recognized by national research laboratories, including the National Institute of Advanced Industrial Science and Technology, and the National Institute of Information and Communications Technology, and was found to meet the published specifications. Anritsu Warranty Anritsu Corporation will repair this
Summary of the content on the page No. 8
Notes On Export Management This product and its manuals may require an Export License/Approval by the Government of the product's country of origin for re-export from your country. Before re-exporting the product or manuals, please contact us to confirm whether they are export-controlled items or not. When you dispose of export-controlled items, the products/manuals need to be broken/shredded so as not to be unlawfully used for military purpose. Lifetime of Parts The life span of certain parts u
Summary of the content on the page No. 9
Crossed-out Wheeled Bin Symbol Equipment marked with the Crossed-out Wheeled Bin Symbol complies with council directive 2002/96/EC (the “WEEE Directive”) in European Union. For Products placed on the EU market after August 13, 2005, please contact your local Anritsu representative at the end of the product's useful life to arrange disposal in accordance with your initial contract and the local law. ix
Summary of the content on the page No. 10
CE Conformity Marking Anritsu affixes the CE Conformity marking on the following product(s) in accordance with the Council Directive 93/68/EEC to indicate that they conform to the EMC and LVD directive of the European Union (EU). CE marking 1. Product Model Model: MP1763C Pulse Pattern Generator 2. Applied Directive EMC: Council Directive 89/336/EEC LVD: Council Directive 73/23/EEC 3. Applied Standards • EMC:Emission: EN 61326: 1997 + A1: 1998 + A2: 2001 + A3: 2003 (Class A) Immunity:EN 61326: 1
Summary of the content on the page No. 11
C-tick Conformity Marking Anritsu affixes the C-tick marking on the following product(s) in accordance with the regulation to indicate that they conform to the EMC framework of Australia/New Zealand. C-tick marking 1. Product Model Model: MP1763C Pulse Pattern Generator 2. Applied Standards EMC: Emission: EN 61326: 1997 A1: 1998 A2: 2001 A3: 2003 + + + (ISM, Group 1, Class A equipment) xi
Summary of the content on the page No. 12
Power Line Fuse Protection For safety, Anritsu products have either one or two fuses in the AC power lines as requested by the customer when ordering. Single fuse: A fuse is inserted in one of the AC power lines. Double fuse: A fuse is inserted in each of the AC power lines. Example 1: An example of the single fuse is shown below: Fuse Holder Example 2: An example of the double fuse is shown below: Fuse Holders xii
Summary of the content on the page No. 13
Composition of MP1763C Operation Manuals The MP1763C Pulse Pattern Generator operation manuals are composed of the follow- ing two documents. Use them properly according to the usage purpose. Function and Operation Part Composition of MP1763C Operation Manuals GPIB Programming Function and Operation Part: These outline the MP1763C, and describes the prepara- tions before use, the panels, specifications, perfor- mances, functions, and operation procedures. GPIB Programming: The MP1763C GPIB confo
Summary of the content on the page No. 14
Title of Contents For Safety ....................................................................................................................... iii SECTION 1 GENERAL ................................................................................................... 1-1 1.1 Features......................................................................................................................... 1-1 1.2 Specifications ..............................................................
Summary of the content on the page No. 15
4.4.1 File save........................................................................................................... 4-23 4.4.2 File recall ......................................................................................................... 4-24 4.4.3 Disk formatting ............................................................................................... 4-24 4.4.4 File deletion ........................................................................................
Summary of the content on the page No. 16
IV.
Summary of the content on the page No. 17
SECTION 1 GENERAL 1.1 Features The MP1763C Pulse Pattern Generator has two data output channels (DATA and DATA) and three clock output channels (CLOCK1, CLOCK1, CLOCK2). The MP1763C is used with the MP1764A/C Error Detector to test high-speed digital communication systems and high-speed semiconductors. It operates over the 50 MHz to 12.5 GHz frequency range and generates four pulse patterns: alternate, programmable, zero substitution, and pseudorandom. Programmable DATA pat- tern can generate
Summary of the content on the page No. 18
Section 1 GENERAL 1.2 Specifications Internal Clock (OPTION 01) 0.05 to 12.5 GHz Operation frequency range External Clock 0.05 to 12.5 GHz N Pattern Pattern length PRBS 2 -1 (N=7,9,11,15,20,23,31) generation 1/2,1/4,1/8,0/8 Mark ratio (1/2,3/4,7/8,8/8 also possible by logic inversion) Number of “AND 1 bit or 3 bits bit” shifts when setting (Selectable using rear panel DIP switch) mark ratio Zero substitution Continuous 0 pattern can be inserted up to pattern length -1. 7 9
Summary of the content on the page No. 19
1.2 Specifications -n Internal Error ratio 1 10 or single (n = 4 , 5, 6, 7 , 8 , 9) Pattern Error generation insertion Insertion possible at any one of 32 channels Insertion (Rear panel swith) position External Error insertion by rising edge of external signal input Eror injection Insertion possible at any one of 32 channels Insertion (Rear panel switch) position DISABLE Error insertion when external signal input level is “H” function -n Error ratio or single 1 10 (n = 4 , 5, 6, 7 , 8 , 9
Summary of the content on the page No. 20
Section 1 GENERAL DATA output Output waveform NRZ Number of outputs DATA, DATA 2 systems independence 0.25 to 2.0 Vp-p/Step 2 mV Amplitude Setting error: 15 % or 100 mV, whichever is larger –2.0 to 2.0 V (V ) /Step 1 mV Offset voltage Setting error: 15 % or 15 % of Amplitude , and 100 mV, whichever is larger Amplitude and offset of DATA and DATA can be set to the same DATA/DATA value. Tracking Rise / fall time 5 GHz, amplitude 1 to 2 Vp-p 35