Fluke 80T-150UA user manual

User manual for the device Fluke 80T-150UA

Device: Fluke 80T-150UA
Category: Thermometer
Manufacturer: Fluke
Size: 0.22 MB
Added : 10/21/2014
Number of pages: 6
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Fluke 80T-150UA. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Fluke 80T-150UA.

For your convenience

If looking through the Fluke 80T-150UA user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Fluke 80T-150UA on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Fluke 80T-150UA to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Fluke 80T-150UA User manual - Online PDF
Advertisement
« Page 1 of 6 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Fluke 80T-150UA but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Fluke 80T-150UA. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1

®

80T-150UA
Universal Temperature Probe
Instruction Sheet

Introduction
The Model 80T-150UA Universal Temperature Probe is a self-contained
temperature-to-voltage converter. The probe is designed to provide a
direct temperature reading when it is connected to any high impedance
DMM that is capable of 1 mV resolution and at least a 300-count full
scale readout capability. Output is 1 mV per degree (Celsius or
Fahrenheit). Two switch-selected temperature output scalings are
provided:

Summary of the content on the page No. 2

Specifications The 80T-150UA will achieve rated accuracy when it is used with any 0.25 % DMM that has an input impedance of ≥1 MΩ. Electrical Measurement Range: (See Probe Limitations) -50 to +150 °C -58 to +302 °F Accuracy: (Applies for one year after purchase or calibration) AMBIENT °C ACCURACY +15 to +35 °C ±1 °C from 0 to +100 °C, decreasing linearly to ±3 °C at -50 and +150 °C 0 to +15 °C and ±2 °C from 0 to +100 °C, +35 to +50 °C decreasing linearly to ±4 °C at -50

Summary of the content on the page No. 3

General Weight: 150 g (5.3 oz) Overall Length: 53.8 inch (1.36 m) Battery: Standard 9 V battery (NEDA #1604,6F22,006P) Battery Life: 600+ hours, typical (Alkaline Battery), 6.5 V minimum Output Termination: Standard 0.75-inch spaced double banana plug Probe Material: Glass-filled valox Probe Size: 0.6 inch maximum diameter Tip Material: Nickle Plated Copper Tip Size: 0.31 cm (0.12 inch diameter) Operating Notes The following paragraphs are intended to familiarize the operator with t

Summary of the content on the page No. 4

F1.eps Figure 1. Initial Case Temperature Above Ambient vs Meter Reading Above Ambient Operation Use the following procedure to operate the 80T-150UA probe: 1. Connect the banana plugs on the 80T-150UA to the input terminals of a high impedance DMM. Observe polarity. 2. Select a dc voltage range that will provide at least 1 mV resolution (1 mV/degree) and a full scale readout that will encompass the expected temperature. The 2 V range of a 3 1/2-digit DMM is adequate. Ignore readin

Summary of the content on the page No. 5

Battery Replacement XWWarning To avoid electrical shock, remove the probe from the measurement surface before opening the case. Totally reassemble the instrument before attempting to use it. 1. Set the power switch to the OFF position. 2. Disconnect the 80T-150UA from the DMM. 3. Turn the 80T-150UA so the power switch is facing down. Remove the single screw located between the banana plugs. 4. Grasp one case half in each hand. Pull the two halves apart, beginning at the end with the ba

Summary of the content on the page No. 6

Sensing Probe and Cable Assembly VR2 VR4 VR3 VR1 MP101 MP102 F4.eps Figure 2. 80T-150UA Temperature Probe Test Equipment Requirements Minimum Use Recommended Instrument Type Specifications Model Mercury Thermometer 0.1 °C Resolution Princo Model SAMA- CP45 Dewar Flask and Cap 1-Pint Capacity (for Ice Bath) Thermos Bottle Metal or Glass Container 1-Pint Capacity Suitable for Boiling Water Digital Voltmeter 100 mV Range with 10 μV Fluke Model 8845A Resolution 1000 mV Range with 10


Alternative user manuals
# User manual Category Download
1 Fluke 114 User manual Thermometer 24
2 Fluke 117 User manual Thermometer 414
3 Fluke 115 User manual Thermometer 126
4 Fluke 1522 User manual Thermometer 16
5 Fluke 1521 User manual Thermometer 33
6 Fluke 1529 Chub-E4 User manual Thermometer 39
7 Fluke 116 User manual Thermometer 34
8 Fluke 1523 User manual Thermometer 12
9 Fluke 1594A User manual Thermometer 4
10 Fluke 1524 User manual Thermometer 22
11 Fluke 1595A User manual Thermometer 5
12 Fluke 2180A User manual Thermometer 6
13 Fluke 233 User manual Thermometer 48
14 Fluke 2190A User manual Thermometer 15
15 Fluke 50S User manual Thermometer 20
16 A&D DT-103 User manual Thermometer 40
17 A&D DT-102 User manual Thermometer 24
18 A&D UT-302 User manual Thermometer 15
19 Acu-Rite 416 User manual Thermometer 7
20 Acu-Rite #00609SBDIA1 User manual Thermometer 3