Fluke 561 user manual

User manual for the device Fluke 561

Device: Fluke 561
Category: Thermometer
Manufacturer: Fluke
Size: 0.83 MB
Added : 10/21/2014
Number of pages: 22
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Fluke 561. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Fluke 561.

For your convenience

If looking through the Fluke 561 user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Fluke 561 on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Fluke 561 to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Fluke 561 User manual - Online PDF
Advertisement
« Page 1 of 22 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Fluke 561 but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Fluke 561. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1

®

561 HVACPro
Infrared Thermometer
Users Manual







PN 2562924
February 2006
© 2006 Fluke Corporation, All rights reserved. Printed in USA
All product names are trademarks of their respective companies.

Summary of the content on the page No. 2

LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for two years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to ex- tend any other warranty on Fluke’s behalf. To obtain service during the warranty period, contact your nearest Fluke auth

Summary of the content on the page No. 3

Table of Contents Title Page Introduction............................................................................................. 1 Contacting Fluke...................................................................................... 1 Safety Information................................................................................... 2 Features................................................................................................... 3 Display..........................

Summary of the content on the page No. 4

561 HVACPro Users Manual Measuring Grille, Register, or Diffuser Discharge Temperature ......... 15 Verifying Thermostat/Room Sensor Accuracy .................................... 15 Checking for Blockage in Air-To-Air Evaporators or Condensers........ 15 Checking Superheat on Fixed Restrictor or Capillary Tube Equipped Evaporators ......................................................................... 16 Checking Subcooling on Air-To-Air Systems With Expansion Valve Equipped Evaporat

Summary of the content on the page No. 5

561 HVAC PRO Infrared Thermometer Introduction The Fluke Model 561 HVACPro Infrared Thermometer (hereafter, “the Thermometer”) can determine the surface temperature by measuring the amount of infrared energy radiated by the target’s surface or by contact using a thermocouple probe. The Thermometer was designed specifically for use in heating, ventilating, and air conditioning (HVAC) applications. Contacting Fluke To contact Fluke, call one of the following telephone numbers: USA: 1-8

Summary of the content on the page No. 6

561 HVACPro Users Manual Safety Information WWarning A Warning identifies conditions and actions that pose hazards to the user. To avoid electrical shock or personal injury, follow these guidelines: • * Do not point laser directly at eye or indirectly off reflective surfaces. • Before using the Thermometer inspect the case. Do not use the Thermometer if it appears damaged. Look for cracks or missing plastic. • Replace the batteries as soon as the battery indicator ( B) appears. • Do

Summary of the content on the page No. 7

Infrared Thermometer Features AVOID EXPOSURE - LASER RADIATION IS EMITTED FROM THIS APERTURE CAUTION LASER RADIATION - DO NOT STARE INTO BEAM < OUTPUT 1mW WAVELENGTH 630 - 670nm CLASS 2 (II) LASER PRODUCT COMPLIES WITH FDA 21CFR 1040.10 AND 1040.11 COMPLIES WITH IEC 60825 efh010f.eps Figure 1. Symbols and Safety Markings Features The Thermometer includes: • Single-spot Laser Sighting • Backlit Display • Hard Case • Current Temperature Plus MIN, MAX, DIF Temperature Displays

Summary of the content on the page No. 8

561 HVACPro Users Manual Laser Display Trigger AA Batteries (2) Function Buttons Battery Cover efh007f.eps Figure 2. Infrared Thermometer Display The primary temperature display reports the current or last IR temperature read until the 7-second hold time elapses. The secondary temperature display reports current thermocouple temperature when a type-K thermocouple is attached. When a thermocouple is not connected, the small temperature display reports a choice of maximum, minimum, or

Summary of the content on the page No. 9

Infrared Thermometer Buttons and Connector 2 1 3 8 4 5 7 6 efh01f.eps A Laser “On” symbol B SCAN or HOLD C °C/°F symbol (Celsius/Fahrenheit) D Primary temperature display E Secondary temperature display F Emissivity LO, MED, HI Temperature values for the MIN, MAX, DIF, KTC. G KTC indicates the thermocouple temperature. Low Battery symbol. Appears when the battery charge H is < 25%. Figure 3. Thermometer Display Buttons and Connector Button/ Description Connector C Pre

Summary of the content on the page No. 10

561 HVACPro Users Manual How the Thermometer Works Infrared thermometers measure the surface temperature of an opaque object. The Thermometer’s optics sense infrared energy, which is collected and focused onto a detector. The Thermometer’s electronics then translate the information into a displayed temperature reading which appears on the display. The laser is used for aiming purposes only. Operating the Thermometer The Thermometer turns on when you press the trigger. The Thermometer tu

Summary of the content on the page No. 11

Infrared Thermometer Operating the Thermometer 132 mm @ 1500 mm 75 mm @ S 900 mm 38 mm @ 300 mm 5.3 " @ 3 " @ 60 " 36 " 1.5 " @ 12 " D D:S=12:1 efh005f.eps Figure 5. Distance and Spot Size Field of View Make sure that the target is larger than the spot size. The smaller the target, the closer you should be to it. See Figure 6. No Yes efh004f.eps Figure 6. Field of View Emissivity Emissivity describes the energy-emitting characteristics of materials. Most organic materials and pa

Summary of the content on the page No. 12

561 HVACPro Users Manual Table 2. Surface Emissivity Measured Measured Switch Setting Switch Setting Surface Surface Aluminum Iron, Cast Oxidized Low Oxidized High, Medium Alloy A3003 Unoxidized Low Oxidized Low Molten Low Roughened Low Iron, Wrought Brass Dull High Burnished Low Lead Oxidized Low Rough Low Copper Oxidized Low, Medium Oxidized Medium Molybdenum Electrical Medium Oxidized Low, Medium Terminal Blocks Haynes Nickel Alloy Medium Oxidized Low Inconel

Summary of the content on the page No. 13

Infrared Thermometer Operating the Thermometer Switch efh012f.eps Figure 7. Switching Between °C and °F Using the Contact Temperature Probe XW Warning To avoid electrical shock or personal injury, do not connect the optional external probe to live electrical circuits. Connect the probe to the input on the top of the Thermometer. The probe temperature and KTC appears in the secondary display. The live infrared temperature continues to show in the primary display. Connect the tempe

Summary of the content on the page No. 14

561 HVACPro Users Manual Table 3 lists recommended Fluke temperature probes for use with the Thermometer: Table 3. Recommended Temperature Probes Probe Usage The piercing probe is the most versatile option. Good for checking air temperature in ducts, surface temperature under carpets/pads, liquids, thermometer wells, vent 80PK-25 temperatures, and for penetrating pipe insulation. The general purpose bead probe is an alternative, for quick, accurate surface 80PK-1 temperatures and air

Summary of the content on the page No. 15

Infrared Thermometer Typical Measurements • The most likely conditions for condensation formation are low dry bulb temperature and high wet bulb temperature (low temperature, high relative humidity). • Use the Fluke 971 to measure the attic/crawlspace relative humidity and determine the dew point temperature. This is the temperature at which condensation will form on the duct wrap. • Use the thermocouple to measure the duct wrap surface temperature. The temperature must remain above th

Summary of the content on the page No. 16

561 HVACPro Users Manual Testing Contactors (Starters) 1. Press D to select emissivity. Press A to select LO for bright contacts, or MED for darkened contacts. 2. Press Cand the press A to select MAX. 3. Measure line and load side of one pole without releasing trigger. 4. A temperature difference between the line and load sides of a pole indicate increased resistance of one point and a contactor may be failing. Testing Enclosed Relays 1. Press D and then press A to set emissivity to LO

Summary of the content on the page No. 17

Infrared Thermometer Typical Measurements Testing Bearings WWarning To avoid injury when testing bearings: • Do not wear loose clothing, jewelry, or anything around neck when working around moving parts such as motors, belts, blower, and fans. • Make sure an electrical disconnect is within reach and operating correctly and freely. • Do not work alone. Note It works best to compare two similar motors operating similar loads. D 1. Press and then press A to select HI emissivity. 2

Summary of the content on the page No. 18

561 HVACPro Users Manual wall you should find parallel isothermal rows indicating the location of heat tubes below the surface. Perpendicular to the outside wall, you should find rising and falling temperatures at equal distances. High temperatures indicate you are scanning a heat tube beneath the floor surface, low falling temperatures indicate a space between the heat tubes. D 1. Press and then press A to select HI emissivity. 2. Press and then press A to select MIN. C 3. To locate r

Summary of the content on the page No. 19

Infrared Thermometer Typical Measurements • If system pressure has been increased above design settings, this can result in symptom of trap failure (stuck open). Check steam pressure. 1. Press D and then press A to select HI emissivity for black iron pipe and painted traps. 2. Press C and then press A to select DIF. 3. Aim the Thermometer at the steam pipe. 4. Scan steam pipe upstream of trap. 5. Scan the steam trap. Scan downstream of trap on condensate return side. 6. Press C and t

Summary of the content on the page No. 20

561 HVACPro Users Manual • All condenser return bend/hairpins should be at or slightly less than condenser saturation temperature. • If a group of return bends/hairpins do not conform to expected temperatures, that indicates a blocked or restricted distributor or distributor tube. Checking Superheat on Fixed Restrictor or Capillary Tube Equipped Evaporators Note Superheat is a critical temperature. 1. Ensure that filter and blower are clean and all registers are open and unobstructed.


Alternative user manuals
# User manual Category Download
1 Fluke 114 User manual Thermometer 24
2 Fluke 117 User manual Thermometer 414
3 Fluke 115 User manual Thermometer 126
4 Fluke 1522 User manual Thermometer 16
5 Fluke 1521 User manual Thermometer 33
6 Fluke 1529 Chub-E4 User manual Thermometer 39
7 Fluke 116 User manual Thermometer 34
8 Fluke 1523 User manual Thermometer 12
9 Fluke 1594A User manual Thermometer 4
10 Fluke 1524 User manual Thermometer 22
11 Fluke 1595A User manual Thermometer 5
12 Fluke 2180A User manual Thermometer 6
13 Fluke 233 User manual Thermometer 48
14 Fluke 2190A User manual Thermometer 15
15 Fluke 50S User manual Thermometer 20
16 A&D DT-103 User manual Thermometer 40
17 A&D DT-102 User manual Thermometer 24
18 A&D UT-302 User manual Thermometer 15
19 Acu-Rite 416 User manual Thermometer 7
20 Acu-Rite #00609SBDIA1 User manual Thermometer 3