Teledyne Trace Oxygen Analyzer 3000TA user manual

User manual for the device Teledyne Trace Oxygen Analyzer 3000TA

Device: Teledyne Trace Oxygen Analyzer 3000TA
Category: Respiratory Product
Manufacturer: Teledyne
Size: 0.74 MB
Added : 5/22/2014
Number of pages: 19
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Teledyne Trace Oxygen Analyzer 3000TA. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Teledyne Trace Oxygen Analyzer 3000TA.

For your convenience

If looking through the Teledyne Trace Oxygen Analyzer 3000TA user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Teledyne Trace Oxygen Analyzer 3000TA on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Teledyne Trace Oxygen Analyzer 3000TA to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Teledyne Trace Oxygen Analyzer 3000TA User manual - Online PDF
Advertisement
« Page 1 of 19 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Teledyne Trace Oxygen Analyzer 3000TA but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Teledyne Trace Oxygen Analyzer 3000TA. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1


QUICKSTART GUIDE
MODEL 3000TA
Trace Oxygen Analyzer







Teledyne Analytical Instruments

Summary of the content on the page No. 2

Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments

Summary of the content on the page No. 3

Model 3000TA Quickstart Guide GETTING STARTED This Quickstart Guide is designed to get you set up and operating your Teledyne Analytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss. This Quickstart Guide should be used in conjunction with the Instruction Manual that shipped with your instrument. Only necessary features to get you operational are discussed in this guide. Many of the advanced features of this anal

Summary of the content on the page No. 4

Quickstart Guide Trace Oxygen Analyzer • Left & Right Arrows Select between functions currently displayed on the VFD screen. • Up & Down Arrows Increment or decrement values of functions currently displayed. • Enter Advances VFD display to the next screen in a series or returns to the Analyze screen if none remain. • Escape Backs VFD display to the previous screen in a series or returns to the Analyze screen if none remain. Digital Meter Display: The meter display is a L

Summary of the content on the page No. 5

Model 3000TA Quickstart Guide • Analog Output Connection: Terminals for current and voltage analog output signals. • Digital Input Terminals: Terminals for connecting digital input signals. • Alarm Connections: 2 concentration alarms and 1 system alarm. • Calibration Contact : To notify external equipment that instrument is being calibrated and readings are not monitoring sample. • Range ID Contacts: Four separate, dedicated, range relay contacts. Low, Medium, High, Cal. • Remote Span/

Summary of the content on the page No. 6

Quickstart Guide Trace Oxygen Analyzer SETUP AND INSTALL Mount the Analyzer The Model 3000TA is for indoor use in a general purpose area. It is NOT for hazardous environments of any type. It is designed for flush panel mounting. Use the four mounting holes— one in each corner of the rigid frame to panel mount the instrument. Gas Connections Note: Do not remove plastic caps on VCR fittings, unless you are ready to make connections. The analyzer is equipped with 1/4

Summary of the content on the page No. 7

Model 3000TA Quickstart Guide 4. If using remote calibration feature: • Connect zero gas to the ZERO IN port. • Connect span gas to the SPAN IN port. 5. Otherwise, calibration gases must be Tee'd into the sample inlet with appropriate valves. 6. Regulate the sample gas pressure between 2 and 50 psig sufficient to keep the front panel flowmeter reading in an acceptable range (0.1 to 2.4 SLPM). For non- pressurized sample or very low pressure, (less than 2 psig) vacuum service plumbing

Summary of the content on the page No. 8

Quickstart Guide Trace Oxygen Analyzer • Connect primary input power: Insert the power cord into the power cord receptacle. • Install the fuse: Fuses are not installed at the factory. Be sure to install the proper fuse as part of installation. Place small screwdriver in notch, and pry cover off, as shown. To change between American and European fuses, remove the single retaining screw, flip fuse block over 180 degrees, and replace screw. Replace fuse as shown below an

Summary of the content on the page No. 9

Model 3000TA Quickstart Guide • Connect Range ID Contacts: There are four dedicated Range ID relay contacts on the rear panel. The first three ranges are assigned to relays in ascending order. Range 1 ID— Low range Range 2 ID— Medium range Range 3 ID— High range Range 4 ID—Air Cal Range (25%). • Network I/O: A serial digital input/output for local network protocol. At this printing, this port is not yet functional. It is to be used in future options to the instrument. • Connect the

Summary of the content on the page No. 10

Quickstart Guide Trace Oxygen Analyzer 9. With the O-ring in place, align the guide pin with the hole on the cell holder. Then, with the holder, lift cell into the cell block. 10. Push the gate on the cell block down so that the slots on the side of the gate engage the locating screws on the side of the block. This forces the holder into position and forms a gas-tight seal. 11. Purge the system using sample or zero gas. 12. Power the system back up. If the installation is carried out

Summary of the content on the page No. 11

Model 3000TA Quickstart Guide BASIC OPERATION Once the analyzer has been installed, it can be configured for your application. Prior to using the instrument you must: • Calibrate the instrument • Define the analysis ranges and choose whether to use autoranging or a fixed range. • Set alarm setpoints and modes of alarm operation. Additional configuration parameters are available to tailor the instrument to your particular application. Refer to the Instruction Manual for specific informati

Summary of the content on the page No. 12

Quickstart Guide Trace Oxygen Analyzer Available Functions The six function buttons are used to control the analyzer. Some functions have submenus that allow additional choices as shown below. Any function can be selected at any time by pressing the appropriate button (unless password restrictions apply). Teledyne Analytical Instruments 12

Summary of the content on the page No. 13

Model 3000TA Quickstart Guide Password Protection If a password is assigned, then setting the following system parameters can be done only after the password is entered: span and zero settings, alarm setpoints, analysis range definitions, switching between autoranging and manual override, setting up an auto-cal, and assigning a new password. However, the instrument can still be used for analysis or for initiating a self-test without entering the password. If you have decided

Summary of the content on the page No. 14

Quickstart Guide Trace Oxygen Analyzer TRAK/HLD Auto—Cal PSWD Logout More • Use the ◄► arrow keys to scroll the blinking over to PSWD, • Press Enter to select the password function. Either the default TETAI password or AAAAA place holders for an existing password will appear on screen depending on whether or not a password has been previously installed. T E T A I Enter PWD —or— A A A A A Enter PWD • If you are not using password protection, press Enter to accept TETAI

Summary of the content on the page No. 15

Model 3000TA Quickstart Guide Auto Mode Zero Calibration • Press Zero to enter the zero function mode. • Use the ▲▼ arrow keys to toggle between AUTO and MAN zero settling. Stop when AUTO appears blinking. Zero: Settling: AUTO To Begin • Press Enter to begin zeroing. #### PPM Zero Slope=#### ppm/s . . . #### PPM Zero 4 Left=### ppm/s The zeroing process will automatically conclude when the output is within the acceptable range for a good zero. Then

Summary of the content on the page No. 16

Quickstart Guide Trace Oxygen Analyzer Span Calibration Make sure an appropriate span gas is connected to the sample port. Auto Mode Spanning: • Press Span to enter the span function. • Use the arrow keys to toggle between AUTO and MAN span settling. Stop when AUTO appears, blinking on the display. Span: Settling: AUTO For Next • Press Enter. Calib. Holding time Cal hold: 5 min • Press Enter to move to the next screen. Span Val: 000008.00 Span Mod # • Use

Summary of the content on the page No. 17

Model 3000TA Quickstart Guide Calib. Holding time Cal hold: 5 min • Press Enter key to continue. Span Val: 000008.00 Span Mod # • Press ▲ () to permit modification (Mod #) of span value. • Use the arrow keys to enter the oxygen concentration of the span gas you are using (209000.00 if you are using air). The ◄► arrows choose the digit, and the ▲▼ arrows choose the value of the digit. • Press Enter to enter the span value into the system and begin the span calibration.

Summary of the content on the page No. 18

Quickstart Guide Trace Oxygen Analyzer • Latching? Ltch–Y/N (Yes/No). • To change any of the alarm parameters, use the ◄► arrow keys to move the blinking over to the specific parameter. • Use the ▲▼ arrow keys to change the value of the parameter. Holding down the key speeds up the incrementing or decrementing. • Once the parameters for alarm 1 have been set, press Alarms again, and repeat this procedure for alarm 2 (AL–2). • To reset a latched alarm, go to Dft– and then press either

Summary of the content on the page No. 19

Model 3000TA Quickstart Guide • Use the ▲▼ arrow keys to switch from AUTO to FX/LO, FX/MED, or FX/HI to set the instrument on the desired fixed range (low, medium, or high). L—100 M—1000 H—10000 Mode—FX/LO —or— L—1 M—10 H—100 Mode—FX/MED —or— L—1 M—10 H—100 Mode—FX/HI Press Escape to re-enter the analyze mode using the fixed range. The Analyze Function Normally, all of the functions automatically switch back to the Analyze function when they have co


Alternative user manuals
# User manual Category Download
1 Teledyne TED 191 User manual Respiratory Product 2
2 Teledyne 3020T User manual Respiratory Product 21
3 Teledyne 3190Z User manual Respiratory Product 1
4 Teledyne Trace Oxygen Analyzer 3000TA-XL-EU User manual Respiratory Product 26
5 Teledyne MX300 User manual Respiratory Product 12
6 Teledyne 3300ZB User manual Respiratory Product 0
7 Teledyne T802 User manual Respiratory Product 0
8 Teledyne MX300-I User manual Respiratory Product 103
9 Agilent Technologies 54600-Series User manual Respiratory Product 6
10 3M 9210 User manual Respiratory Product 3
11 3M 7000 Series User manual Respiratory Product 1
12 3M FR-M40 User manual Respiratory Product 6
13 3M 9211 User manual Respiratory Product 3
14 AOSafety OMNI STAR User manual Respiratory Product 1
15 AOSafety Eye Gear Rotational Display A1655 User manual Respiratory Product 1