Teledyne 311-D user manual

User manual for the device Teledyne 311-D

Device: Teledyne 311-D
Category: Oxygen Equipment
Manufacturer: Teledyne
Size: 0.77 MB
Added : 8/27/2013
Number of pages: 18
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Teledyne 311-D. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Teledyne 311-D.

For your convenience

If looking through the Teledyne 311-D user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Teledyne 311-D on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Teledyne 311-D to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Teledyne 311-D User manual - Online PDF
Advertisement
« Page 1 of 18 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Teledyne 311-D but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Teledyne 311-D. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1

INSTRUCTION MANUAL

FOR

TRACE OXYGEN ANALYZER

MODEL 311-D






RANGES: 0-100 PPM; 0-1000 PPM; 1% AND 100%











TELEDYNE ANALYTICAL INSTRUMENTS
16830 CHESTNUT STREET
CITY OF INDUSTRY, CALIF. 91749
TELEPHONE: (626) 934-1500
FAX: (626) 961-2538
TWX: (910) 584-1887 TDYANYL COID
EASYLINK: 628-31862
Web: www.teledyne-ai.com

Summary of the content on the page No. 2

TABLE OF CONTENTS MODEL 311D 1. INTRODUCTION 1 1.1 Description 1 1.2 Method of Analysis 1 1.3 Outstanding Features 1 1.3.1 Micro-Fuel Cell 1 1.3.2 Reliable Calibration 1 1.3.3 Integral Power Supply 2 1.3.4 Accuracy and Response 2 1.3.5 Compact Packaging 3 1.3.6 Circuit Description 3 2. SUPPORTING EQUIPMENT AND SERVICES 4 2.1 Sampling Equipment 4 2.2 Power Service 4 3. OPERAT

Summary of the content on the page No. 3

1. INTRODUCTION 1.1 Description: The Teledyne Analytical Instruments (TAI) Model 311-D is a portable trace oxygen analyzer which can be operated without an external power source, and reliably calibrated without the use of cumbersome, questionable, “certified” calibration gases. The instrument provides for trace oxygen analysis in decade steps ranging from 0-100 to 0-1000 ppm (full scale) plus a 0-1% and 0-100% calibration range that encompasses the known oxygen concentration of atmosp

Summary of the content on the page No. 4

After the electronics have been properly zeroed (a one-time factory operation), the instrument cannot produce an output indication in the absence of oxygen; therefore, the need for a “zero” standardization gas is obviated. 1.3.3 Integral Power Supply: The differential power and digital panel meter (DPM) requirements of the instrument amplifier are furnished by three internally-mounted 750 milliampere-hour nickel-cadmium batteries. Fully charged, these batteries will provide enough p

Summary of the content on the page No. 5

Further disassembly may be accomplished by removing the backplate assembly from its four mounting standoffs, and laying the two separated assemblies out as illustrated on the “Analyzer Wiring Diagram”. The diagram is included among the drawings at the rear of the manual. 1.3.6 Circuit Description: A Micro-Fuel Cell (Class B-2C) is used as the transducer; its output current is in proportion to the concentration of the oxygen measured. A current-to-voltage converter, A3, supplies an out

Summary of the content on the page No. 6

necessary or desirable, it must have a metallic diaphragm. Regulators with organic or plastic diaphragms are permeable to oxygen, and if used in the sampling system, will lead to high oxygen readings. For atmospheric pressure sampling, connect a pump and flow control valve downstream from the analyzer and draw (rather than push) the sample through the instrument. TAI supplies three male disconnect fittings with the instrument; one for installation of the customer’s sample line, one to

Summary of the content on the page No. 7

dictated by the customer’s application. If the instrument were being used to certify the oxygen content of a product for delivery, then a calibration prior to certification would certainly be in order. If, on the other hand, the instrument is being used to monitor or guard a sample, the evidence provided by the analyzer will determine when a calibration check is in order. DO NOT CALIBRATE THE INSTRUMENT UNLESS THERE IS A TRACE OXYGEN GAS READILY AVAILABLE FOR PURGING IMMEDIATELY FOLLOWI

Summary of the content on the page No. 8

If, on the other hand, the instrument is not to be used immediately after calibration, and a low ppm oxygen gas is being employed as a purge, allow the manifold to be purged overnight, and then disconnect both male fittings. ALWAYS DISCONNECT THE SOURCE FITTING FIRST, AND IMMEDIATELY THEREAFTER, THE VENT FITTING. 3.3 Positive Pressure Sampling: When connecting the instrument to a positive pressure sample source, ALWAYS proceed as follows: 1. Before making ANY connections to the in

Summary of the content on the page No. 9

every two (2) days of continuous use. To recharge the batteries, place the range switch in the “OFF” position and connect the power cord to a convenience outlet. The integral charging circuit will automatically energize and regulate battery charging current when the switch is in the “OFF” position and the AC cord is plugged into the power line. Under no circumstances allow the instrument to remain “ON” when the “LOW BATT” warning is indicated. Charging the batteries more than this will

Summary of the content on the page No. 10

No tools are required to replace the cell in the instrument. Simply unscrew (ccw) the plug at the bottom of the analyzer, and the cell will drop out of the manifold cavity. Remove the new cell from its package, and carefully remove the shorting clip. Do not touch the silver-colored sensing surface of the cell -- as it is covered with a delicate Teflon membrane that can be ruptured in handling. Place the cell on the end of the manifold plug, so that the sensing surface of the cell is i

Summary of the content on the page No. 11

1. Disconnect cell. 2. Move range switch to “CAL” position. 3. Adjust P1 such that the output of A3, pin 6, measures between 0 and +0.5 mV, ideally +-.3 mV. 4. Adjust P2 for 0 ± 1 mV at output of A2, pin 6. 5. Verify that the offset is the same on all ranges. 6. Re-connect cell. 6. LEAK TESTING If a leak is suspected in the unit, DO NOT ATTEMPT TO TIGHTEN THE DISCONNECT FITTINGS. THE FITTINGS ARE POTTED IN EPOXY AND TIGHTENING THEM WILL BREAK THE SEAL! To check for leaks, TAI

Summary of the content on the page No. 12

OPERATING TEMPERATURE RANGE: 30 ° TO 125 °F. RESPONSE AND RECOVERY: AT THE SPECIFIED FLOWRATE (0.25 SCFH), THE SENSOR OUTPUT SHOULD BE, IN TEN SECONDS, WITHIN ±10% OF AN INTRODUCED GAS HAVING A CONCENTRATION OF 100 PPM OR MORE; FOR A CONCENTRATION OF LESS THAN 100 PPM, THE SENSOR OUTPUT SHOULD BE WITHIN ±10% IN SIXTY SECONDS (PROVIDED THAT THE SENSOR IS NOT ALREADY SUPERSATURATED WITH A VERY HIGHLY-CONCENTRATED GAS, AND THE INTRODUCED GAS IS NOT VERY LOW IN CONCENTRATION.

Summary of the content on the page No. 13

EFFECT OF CO2 ON B-2C CELL LIFE -11-

Summary of the content on the page No. 14

RECOMMENDED SPARE PARTS LIST MODEL 311D QTY. P/N DESCRIPTION 5 F-39 MICRO FUSE, 0.25 AMP., FAST-BLOW (FOR 210-240 VAC) 5 F-51 MICRO FUSE, 0.50 AMP., FAST-BLOW (FOR 100-125 VAC) 1 C-6689-A2C MICRO-FUEL CELL, CLASS A2C (FOR GAS SAMPLES WITH CO2) 1 C-6689-B2C MICRO-FUEL CELL, CLASS B-2C (STANDARD) 3 B-27296 BATTERY ASSEMBLY 1 B-30717 LCD METER ASSEMBLY 1 L-79 LAMP, NEON ASSEMBLY A MINIMUM CHARGE IS APPLICABLE TO SPARE PARTS ORDERS. IMPORTA

Summary of the content on the page No. 15

DRAWING LIST MODEL 311D B-26473 PICTORAL DIAGRAM C-28339 SCHEMATIC C-26606 WIRING DIAGRAM NOTE: The MSDS on this material is available upon request through the Teledyne Environmental Health and Safety Coordinator. Contact at (626) 934-1592. -13-

Summary of the content on the page No. 16

Summary of the content on the page No. 17

Date: 6/20/2006 Time : 01:45:07 PM Title: \\Indsrsvr01\drawings\MASTER\26000\C26606-14.dwg

Summary of the content on the page No. 18

Date: 7/20/04 Time : 10:56:54 AM Title: G:\MASTER\28000\C28339-8.DWG


Alternative user manuals
# User manual Category Download
1 Teledyne OT-3 User manual Oxygen Equipment 2
2 Teledyne 3000PA User manual Oxygen Equipment 3
3 Teledyne 3000ZA-3X User manual Oxygen Equipment 2
4 Teledyne 3000TA User manual Oxygen Equipment 24
5 Teledyne OXYGEN DETECTOR 319R User manual Oxygen Equipment 3
6 Teledyne 3000ZA User manual Oxygen Equipment 1
7 Teledyne 3000TA-XL-EU User manual Oxygen Equipment 8
8 Teledyne 3010MB User manual Oxygen Equipment 5
9 Teledyne 311PC User manual Oxygen Equipment 6
10 Teledyne 3010TBC User manual Oxygen Equipment 1
11 Teledyne 310 User manual Oxygen Equipment 2
12 Teledyne 307WA User manual Oxygen Equipment 0
13 Teledyne 3110 User manual Oxygen Equipment 75
14 Teledyne 315 User manual Oxygen Equipment 2
15 Teledyne 3300PA User manual Oxygen Equipment 3
16 3M 8110S User manual Oxygen Equipment 1
17 3M 7187 User manual Oxygen Equipment 1
18 3M 8210PLUS User manual Oxygen Equipment 1
19 3M 8214 User manual Oxygen Equipment 0
20 3M 8210 User manual Oxygen Equipment 6