Summary of the content on the page No. 1
OPERATING MANUAL
150 WATT ELECTRONIC LOAD MODULE
Agilent Model 60501B
FOR MODULES WITH SERIAL NUMBERS:
3118A-00101 AND ABOVE
Agilent Part No. 60501-90009 Printed in U.S.A.
Microfiche No. 60501-90010
Summary of the content on the page No. 2
DECLARATION OF CONFORMITY according to ISO/IEC Guide 22 and EN 45014 Manufacturer’s Name: Agilent Technologies Manufacturer’s Address: New Jersey Division 140 Green Pond Road Rockaway, NJ 07866 U.S.A. declares that the product Product Name: Load mainframe and modules Model Number(s): Agilent 6050A, 6051A mainframes with modules Agilent 60501A/B, 60502A/B, 60503A/B, 60504A/B, 60507A/B conform(s) to the following Product Specifications: 1 Safety: IEC 348:1978 / HD401 S1:1981
Summary of the content on the page No. 3
150-Watt Module About This Manual This manual provides information for the Agilent 60501B 150-Watt Electronic Load Module. It is designed as a supplement to the Agilent 6050A/6051A Multiple Input Mainframe Electronic Load Operating Manual (part number 06050- 90001). Four tables provide the following module-specific information: Table 60501-1 lists both the specifications and supplemental characteristics of the module. Specifications indicate warranted performance in the 25 °C ± 5 °C region of
Summary of the content on the page No. 4
Table 60501-1. Specification and Supplemental Characteristics SPECIFICATIONS DC Input Rating: Current: 0 to 30 A Voltage: 3 to 60 V (minimum dc operation from 0 to 2 V for 0 to 30 A) Power: 150 W at 40 °C (derated to 112 W at 55 °C) A. OPERATING CHARACTERISTICS B. DERATED CURRENT DETAIL Constant Current Mode: Ranges: 0 to 3 A; and 0 to 30 A Accuracy: (after 30 second wait): ± 0.1% ± 40 mA (both ranges) Resolution: 0.8 mA (3 A range); 8 mA (30 A range) Regu
Summary of the content on the page No. 5
Table 60501-1 Specifications and Supplemental Characteristics (continued) Transient Operation: Continuous Mode Frequency Range: 0.25 Hz to 10 kHz Frequency Resolution: 4% Frequency Accuracy: 3% Duty Cycle Range: 3% to 97% (0.25 Hz to 1 kHz); 6% to 94% (1 kHz to 10 kHz) Duty Cycle Resolution: 4% Duty Cycle Accuracy: 6% of setting ± 2% Pulsed Mode Pulse Width: 50 μs ± 3% minimum; 4 s ± 3% maximum Transient Current Level (0 to 3 A and 0 to 30 A ranges): Resolution: 13 mA (3 A
Summary of the content on the page No. 6
Table 60501-1 Specifications and Supplemental Characteristics (continued) External Analog Programming 0 to 10 V (dc or ac): Bandwidth: 10 kHz (3 db frequency) Accuracy: ± 4.5% ± 40 mA (0 to 3 A range) ± 4.5% ± 130 mA (0 to 30 A range) ± 0.8% ± 200 mV (0 to 60 V range) ° ° Temperature Coefficient: 100 ppm/ C ± 3 mA/ C (current ranges) 100 ppm/°C ± 1 mV/°C (voltage range) External Current Monitor (0 to 10 V): Accuracy: ± 4% ± 40 mA (referenced to analog common) Temperature Coefficient: 60 ppm/°C ±
Summary of the content on the page No. 7
Table 60501-1 Specifications and Supplemental Characteristics (continued) Current Slew Rate:* Rate # 30 A Range Step 3 A Range Step Transition Time 1 0.5 A/ms 0.05 A/s 8.0 ms 2 1.2 A/ms 0.12 A/s 3.2 ms 3 2.5 A/ms 0.25 A/ms 1.6 ms 4 5 A/ms 0.5 A/ms 800 μs 5 12 A/ms 1.2 A/ms 320 μs 6 25 A/ms 2.5 A/ms 160 μs 7 5 A/ms 0.05 A/μs 80 μs 8 12 A/ms 0.12 A/μs 32 μs 9 25 A/ms 0.25 A/μs 16 μs 10 0.5 A/μs 0.05 A/μs 12 μs 11 1.2 A/μs 0.12 A/μs 12 μs 12 2.5 A/μs 0.25
Summary of the content on the page No. 8
Table 60501-1 Specifications and Supplemental Characteristics (continued) Source Turn-On Current Overshoot: Less than 10% of final value (in CC and CR modes when connected to power supplies with voltage rise times of greater than 500μs). Programmable Short Circuit: 0.066 Ω (0.04 Ω typical) Programmable Open Circuit: 20 kΩ (typical) Drift Stability (over an 8 hour interval): Current: ± 0.03% ± 5 mA Voltage: ± 0.01% ± 10 mV Reverse Current Capacity: 50 A when unit is on; 20 A when unit is off Weig
Summary of the content on the page No. 9
Table 60501-2. Programming Ranges (continued) Transient Operation Set Frequency FREQ value "TRAN:FREQ value" 0.25 Hz to 10 kHz Set Duty Cycle DCYCLE value "TRAN:DCYC value" 3-97% (0.25 Hz-1 kHz) (shift) 6-94% (1 kHz-10 kHz) *Set Pulse Width "TRAN:TWID value" 0.00005 to 4 s Trigger Operation *Set Trigger Period "TRIG:TIM value" 0.000008 to 4 s Current Protection *Set Current Level "CURR:PROT value" 0 to 30.6 A *Set Delay Time "CURR:PROT:DEL value" 0 to 60 s *Can only be programmed remotely via
Summary of the content on the page No. 10
Table 60501-4. Calibration Information Ranges and Variables Variables Power Supply Current Calibration Points Value Settings Shunt High Current Range Hi_curr_rng 30 5 V/31 A 100 A High Current Offset Hi_curr_offset 0.013 Low Current Range Lo_curr_rng 3 5 V/10 A 15 A Low Current Offset Lo_curr_offset 0.013 Voltage Range N/A N/A 61 V/2 A N/A Voltage Hi point Volt_hipt 60 Voltage Lo point Volt_lopt 2.7 Low Resistance Range Lo_res_rng 2 15 V/5.5 A 15 A Low Resistance Hi point Lo_res_hipt 1.9 Low Re
Summary of the content on the page No. 11
Agilent Sales and Support Office For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir You can also contact one of the following centers and ask for a test and measurement sales representative. United States: Latin America: Agilent Technologies Agilent Technologies Test and Measurement Call Center Latin American Region Headquarters P.O. Box 4026 5200 Blu