Summary of the content on the page No. 1
Victoreen® 948A-40 and 948A-50
Remote Alarm and Alarm/Meter
Instruction Manual
Summary of the content on the page No. 2
(Blank Page)
Summary of the content on the page No. 3
Contents Section 1: Introduction.................................................................................................................... 1-1 1.1 General Description................................................................................................................................................1-1 1.2 Application............................................................................................................................................................
Summary of the content on the page No. 4
(Blank Page)
Summary of the content on the page No. 5
Section 1 Introduction 1.1 General Description Remote Alarm and Alarm/Meter (948A-40 & 948A-50) The 948A-50 Remote Alarm/Meter continuously indicates the level of radiation at the detector site and provides a visual and audible indication of a high radiation level on an analog meter, light, and Sonalert. The meter of the 7 Remote Alarm/Meter is an analog meter scale that indicates in the range of 0.1 to 10 mR/h. The light and Sonalert are actuated when the radiation level exceeds the h
Summary of the content on the page No. 6
1.4 Receiving Inspection Upon receipt of the unit: 1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier and notify Cardinal Health at 440.248.9300. 2. Remove the contents from the packing material. 3. Verify that all items listed on the packing list have been received and are in good condition. CAUTION If any of the listed items are missing or damaged, notify Cardinal Health at 440.248.9300. 1.5 Storage Storage of Victoreen instru
Summary of the content on the page No. 7
Indicated battery and other operational tests must be performed prior to each use to assure that the instrument is functioning properly. If applicable, failure to conduct periodic performance tests in accordance with ANSI N323-1978 (R1983) Radiation Protection Instrumentation Test and Calibration, paragraphs 4.6 and 5.4, and to keep records thereof in accordance with paragraph 4.5 of the same standard, could result in erroneous readings or potential danger. ANSI N323-1978 becomes, by this re
Summary of the content on the page No. 8
(Blank page)
Summary of the content on the page No. 9
Section 2 Installation 2.1 Installation CAUTION Ensure all power is removed prior to installing the 948A- 40 or 948A-50 Remote Alarm, Alarm/Meter. Installation of the 948A-50 Remote Alarm/Meter consists of finding a suitable location for mounting, such as a wall or post, and providing an electrical interface. Both the 948A-40 and 948A-50 Remote Alarm, Remote Alarm/Meter can be mounted up to 1500 feet from the detector and UDR. There are two possible wiring configurations (referenc
Summary of the content on the page No. 10
(Blank page)
Summary of the content on the page No. 11
Section 3 Operation 3.1 Operation Since the operation and theory for the 948A-40 is identical to the 948A-50 with the exception that the 948A-40 has no meter, only the 948A-50 will be discussed. Once installation has been completed, operation of the VICTOREEN 948A-50 Remote Alarm/Meter is automatic and does not require operator attention. 7 The analog meter continuously indicates the level of radioactivity measured at the detector site from 0.1 to 10 mR/h. The light and Sonalert are ac
Summary of the content on the page No. 12
(Blank Page)
Summary of the content on the page No. 13
Section 4 Maintenance, Calibration, and Troubleshooting 4.1 Maintenance The 948A-40 Remote Alarm and 948A-50 Remote Alarm/Meter require no routine maintenance other than periodic calibration for the 948A-50. Refer to calibration for further information. 4.2 Calibration The 948A-50 requires calibration before it is initially placed into service. In addition, it should be recalibrated at regular intervals during routine service. The length of time between calibration intervals should
Summary of the content on the page No. 14
Troubleshooting consists of checking the field wiring; refer to loop drawing 945-3 in Appendix A. If a problem develops with the remote alarm/meter, refer to the applicable drawings located in Appendix A and to TP948A for the factory test procedure. Should it be necessary to return the remote alarm/meter during the warranty period, contact Cardinal Health at 440.248.9300. NOTE If a problem cannot be resolved by applying the troubleshooting procedures described above, contact Cardinal
Summary of the content on the page No. 15
Appendix A Applicable Drawings and Bill of Materials A.1 Drawings Drawing Number Description GEL948A-50 Dimensional Outline 948A-40-5 Assembly Drawing w/o Meter 948A-50-5 948A-40/948A-50 Assembly Drawing 948A-50-3 Schematic Diagram 945-3 Loop Diagram 945A-3 Loop Diagram WL948A-40-5 Wire List WL948A-50-5 Wire List TP948A Test Procedure A.2 Bill of Materials Drawing Number Description 948A-40 Remote Alarm 948A-50 Remote Alarm/Meter 948A-40-5 Remote Alarm 948A-50-5 Remot
Summary of the content on the page No. 16
Radiation Management Services For additional information, please contact Radiation Management Services business of Cardinal Health at 440.248.9300. © 2004 Cardinal Health, Inc. or one of its subsidiaries All rights reserved. Manual No. 948A-50-1 Rev. 2 04 aug 04 Cardinal Health 6045 Cochran Road Cleveland, Ohio 44139 www.cardinal.com/rms