Summary of the content on the page No. 2
Dear Customer, RECOMMENDATIONS FOR USERS Thank you for choosing a Hanna Instruments product. Before using these products, make sure they are entirely suitable for Please read this instruction manual carefully before using the instruments. the environment in which they are used. This manual will provide you with the necessary information for correct use of the instruments, as well as a precise idea of their versatility. Operation of these instruments in residential areas could cause If you need a
Summary of the content on the page No. 3
HI 1413B PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION PRELIMINARY EXAMINATION Glass-body, single junction, flat tip, Viscolene, non-refillable, combination pH electrode. Use: surface measurement. Remove the instrument from the packing material and examine it to 12 mm 0.5" make sure that no damage has occurred during shipping. If there is any damage, notify your dealer or the nearest Hanna HI 1413 Customer Service Center. 110 mm 4.3" Each meter i
Summary of the content on the page No. 4
FC 100B FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION Plastic-body (PVDF), double junction, refillable, combination pH elec- trode. Use: general purpose for food industry. 12 mm 0.5" FC 100 120 mm 4.7" FC 200B Plastic-body (PVDF), open junction, conic, Viscolene, non-refillable, combination pH electrode. Use: meat & cheese. 6 mm 0.25" FC 200 75 mm 2.95" FC 210B Glass-body, double junction, conic, Viscolene, non-refillable, comb
Summary of the content on the page No. 5
HI 1330B SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS Glass-body, semimicro, single junction, refillable, combination pH electrode. -2.00 to 16.00 pH Use: laboratory, vials. 5mm DIA 5mm ±699.9 mV / ±1999 mV 0.2" 0.2" RANGE (HI 9125 only) HI 1330 -20.0 to 120.0 °C 120 mm (-4.0 to 248.0 °F) 4.7" "S" VERSION 0.01 pH HI 1331B 0.1 mV / 1 mV RESOLUTION Glass-body, semimicro, single junction, refillable, combination pH electrode. (HI 9125 only) Use: flasks. 0.1 °C (0.1 °F)
Summary of the content on the page No. 6
pH ELECTRODES OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE All electrodes part numbers ending in B are supplied with a BNC connector and 1 m (3.3') cable, as shown below: INITIAL PREPARATION The meter is supplied with rechargeable batteries (located inside the meter - see page 14 for details). To prepare the instrument for use, connect the pH electrode and the temperature probe to the BNC and temperature sockets on the top of HI 1043B the instrument.
Summary of the content on the page No. 7
The LCD will show the pH measurement and the temperature of the ACCESSORIES ACCESSORIES ACCESSORIES ACCESSORIES ACCESSORIES sample. pH CALIBRATION SOLUTIONS HI 70004P pH 4.01 Buffer Solution, 20 mL sachet, 25 pcs HI 70007P pH 7.01 Buffer Solution, 20 mL sachet, 25 pcs HI 70010P pH 10.01 Buffer Solution, 20 mL sachet, 25 pcs HI 7004L pH 4.01 Buffer Solution, 500 mL bottle HI 7004M pH 4.01 Buffer Solution, 230 mL bottle HI 7006L pH 6.86 Buffer Solution, 500 mL bottle HI 7006M pH 6.86 Buffer Soluti
Summary of the content on the page No. 8
Submerge the ORP electrode 4cm (1½") into the sample to be tested TROUBLESHOOTING GUIDE TROUBLESHOOTING GUIDE TROUBLESHOOTING GUIDE TROUBLESHOOTING GUIDE TROUBLESHOOTING GUIDE and wait a few minutes for the reading to stabilize. Measurements within the ±699.9 mV range are displayed with 0.1 SM YMPTOMSPN ROBLE SOLUTIO mV resolution, while outside this range the resolution automatically Slow reponse/excessiveDe irty pH electrode. The electrode needs to b switches to 1 mV. drift. cleaned. For accur
Summary of the content on the page No. 9
For refillable electrodes, if the refill solution (electrolyte) is more than 2½ p p p p pH CALIBRATION H CALIBRATION H CALIBRATION H CALIBRATION H CALIBRATION cm (1") below the fill hole, add the appropriate Electrolyte Solution. It is recommended to calibrate the instrument frequently, especially if MEASUREMENT high accuracy is required. Rinse the electrode tip with distilled water, immerse it 4 cm (1½") in The pH range should be recalibrated: the sample and stir gently for a few seconds. • Whe
Summary of the content on the page No. 10
• The LCD will flash the "WAIT NOT READY" message. E E E E ELECTRODE CONDITIONING & MAINTENANCE LECTRODE CONDITIONING & MAINTENANCE LECTRODE CONDITIONING & MAINTENANCE LECTRODE CONDITIONING & MAINTENANCE LECTRODE CONDITIONING & MAINTENANCE • Once the reading is stable, if it is not close to the selected buffer, "WRONG " and "WRONG " will blink alternatively; if it is close to the selected buffer the display will change to "READY" and blinking "CFM". • Press the CFM key to confirm the calibrati
Summary of the content on the page No. 11
• If the reading is not close to the selected buffer, "WRONG " TEMPERATURE CORRELATION TEMPERATURE CORRELATION TEMPERATURE CORRELATION TEMPERATURE CORRELATION TEMPERATURE CORRELATION and "WRONG " will blink alternatively; FOR FOR FOR FOR FOR p p p p pH SENSITIVE GLASS H SENSITIVE GLASS H SENSITIVE GLASS H SENSITIVE GLASS H SENSITIVE GLASS • If the reading is close to the selected buffer and the reading is The resistance of glass electrodes partially depends on the temperature. stable, the “READY
Summary of the content on the page No. 12
p p p p pH BUFFER TEMPERATURE H BUFFER TEMPERATURE H BUFFER TEMPERATURE H BUFFER TEMPERATURE H BUFFER TEMPERATURE LCD MESSAGE GUIDE LCD MESSAGE GUIDE LCD MESSAGE GUIDE LCD MESSAGE GUIDE LCD MESSAGE GUIDE DEPENDENCE DEPENDENCE DEPENDENCE DEPENDENCE DEPENDENCE TAGS & SYMBOLS The temperature has an effect on pH. The calibration buffer solutions mode tags are affected by temperature changes to a lesser degree than normal solutions. During calibration the instrument will automatically calibrate pH or
Summary of the content on the page No. 13
Replace the rechargeable batteries only if necessary. To replace the SETUP MENU SETUP MENU SETUP MENU SETUP MENU SETUP MENU rechargeable batteries, follow the next steps: • Turn OFF the instrument. The instrument allows the user to configure several parameters • Unscrew the screws from the bottom sides of the instrument. through the Setup Menu. To enter the Menu mode, while in measurement mode, press and hold the SETUP key for about 5 seconds. Once the menu is entered, each parameter can be chan
Summary of the content on the page No. 14
It is recommended to recharge the rechargeable batteries as soon as m m m m mV CALIBRATION (HI 9125 V CALIBRATION (HI 9125 V CALIBRATION (HI 9125 V CALIBRATION (HI 9125 V CALIBRATION (HI 9125 only only only only only) ) ) ) ) the display will flash the battery symbol. HI 9125 has been precalibrated for mV range at the factory. The meter is also provided with the BEPS (Battery Error Prevention System) For optimum accuracy, it is recommended to recalibrate the meter for feature which automatically