Summary of the content on the page No. 1
Manual Addendum
Model T265
Chemiluminescence Ozone Analyzer
(Addendum to Model T200 Operation Manual PN 06858)
© TELEDYNE ADVANCED POLLUTION INSTRUMENTATION
9480 CARROLL PARK DRIVE
SAN DIEGO, CA 92121-5201
USA
Toll-free Phone: 800-324-5190
Phone: 858-657-9800
Fax: 858-657-9816
Email: api-sales@teledyne.com
Website: http://www.teledyne-api.com/
Copyright 2011 07337B DCN6324
Teledyne Advanced Pollution Instrumentation December 07, 2011
Summary of the content on the page No. 2
Summary of the content on the page No. 3
IMPORTANT SAFETY INFORMATION Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a safety alert symbol and placed throughout this manual and inside the instrument. The symbols with messages are defined as follows: WARNING: Electrical Shock Hazard HAZARD: Strong oxidizer GENERAL WARNING/CAUTION: Read the accompanying message for
Summary of the content on the page No. 4
CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. 1tée par un pictogramme d’alerte de sécurité; ces pictogrammes se retrouvent dans ce manuel et à l’intérieur des instruments. Les symboles correspondent aux consignes suivantes : AVERTISSEMENT : Risque de choc électrique DANGER : Oxydant puissant AVERTISSEMENT GÉNÉRAL / MISE EN GARDE : Lire la cons
Summary of the content on the page No. 5
ABOUT THIS MANUAL This addendum is to be used in conjunction with the T200 Operation Manual, PN 06858. REVISION HISTORY This section provides information regarding changes to this manual. 2011 December 07, Model T265 Addendum, Rev DCN Description of Changes B 6324 Specs update: Zero Noise, LDL, Zero and Span Drifts 2011 November 08, T265 Addendum, PN07337A, DCN6120, Initial Release 07337B DCN6324 iii
Summary of the content on the page No. 6
This page intentionally left blank. 07337B DCN6324 iv
Summary of the content on the page No. 7
TABLE OF CONTENTS IMPORTANT SAFETY INFORMATION ....................................................................................................................... I ABOUT THIS MANUAL................................................................................................................................................. III REVISION HISTORY ..............................................................................................................................................
Summary of the content on the page No. 8
This page intentionally left blank. vi 07337B DCN6324
Summary of the content on the page No. 9
1. INTRODUCTION This section presents a brief overview of the Model T265; supplemental information may be found in the Model T200 Operation Manual. 1.1 OVERVIEW The Model T265 is a close derivative of the Model T200 Chemiluminescence NOx Analyzer. This addendum provides an overview of the instrument with details of the features and functions that are specific to the Model T265; it is intended as a supplement to the Model T200 operation manual (Teledyne API part number 06858). Note: it
Summary of the content on the page No. 10
Introduction Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual This page intentionally left blank. 8 07337B DCN6324
Summary of the content on the page No. 11
2. SPECIFICATIONS, APPROVALS, & CERTIFICATIONS Table 2-1 presents the specifications for the Model T265 analyzer. This is followed by EPA and Safety and Emissions Compliance information. 9 07337B DCN6324
Summary of the content on the page No. 12
Specifications, Approvals, & Certifications Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual Table 2-1. Model T265 Specifications PARAMETER SPECIFICATION Ranges 0-100 ppb to 0-2,000 ppb full scale, user selectable. Measurement Mode Single Range or AutoRange 3 3 Measurement Units ppb, ppm, µg/m mg/m 1 Zero Noise <0.15 ppb (RMS) 1 Span Noise <0.5% of reading (RMS) above 100 ppb Lower Detectable Limit <0.3 ppb Zero Drift (24 hours) <0.5 ppb Span Drift (24 ho
Summary of the content on the page No. 13
Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual Specifications, Approvals, & Certifications 2.1 EPA DESIGNATION Teledyne API’s Model T265 Chemiluminescence Ozone Analyzer received EPA approval for Designation EQOA-0611-199. The designation will be considered valid when operated under the following conditions: on any full scale range between 0-100 ppb and 0-1000 ppb with any range mode (Single, Dual, or AutoRange) at any ambient temperature in the range o
Summary of the content on the page No. 14
Specifications, Approvals, & Certifications Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual This page intentionally left blank. 12 07337B DCN6324
Summary of the content on the page No. 15
3. GETTING STARTED This section addresses the procedures for unpacking the instrument and inspecting for damage, and introduces the instrument layout, then presents the procedures for getting started: making electrical and pneumatic connections, and conducting an initial calibration check. Please see the Model T200 Operation Manual for more information. 3.1 UNPACKING CAUTION – GENERAL SAFETY HAZARD To avoid personal injury, always use two persons to lift and carry the Model T265.
Summary of the content on the page No. 16
Getting Started Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual CAUTION Remove dust plugs from pneumatic fittings on the rear panel of the analyzer, prior to applying power to the analyzer. Note It is recommended that you store shipping containers/materials for future use if/when the instrument should be returned to the factory for repair and/or calibration service. See Warranty section in this manual and shipping procedures on our Website at http://www.teled
Summary of the content on the page No. 17
Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual Getting Started 3.2 INSTRUMENT LAYOUT Instrument layout includes front panel and display, rear panel connectors, and internal chassis layout. 3.2.1 Front Panel Figure 3-1 illustrates the front panel of the Model T265 analyzer. Figure 3-1. Front Panel 15 07337B DCN6324
Summary of the content on the page No. 18
Getting Started Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual 3.2.2 Rear Panel Figure 3-2 illustrates the layout of the Model T265 analyzer’s rear panel. Figure 3-2. Rear Panel 16 07337B DCN6324
Summary of the content on the page No. 19
Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual Getting Started 3.2.3 Internal Chassis Layout Figure 3-3 illustrates the location of the internal chassis components. Figure 3-3. Internal Chassis Layout 17 07337B DCN6324
Summary of the content on the page No. 20
Getting Started Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual 3.3 CONNECTIONS AND SETUP This section presents connection information for setting up the instrument to operate. 3.3.1 Electrical and Pneumatic Connections 1. Refer to Figure 3-2 for the rear panel electrical and pneumatic connectors. 2. Mount the NOx scrubber with base down (arrow labeled “FLOW” pointing down) onto the rear panel. 3. Connect tubing from NOx scrubber base fitting to rear panel fit