Emerson 644 user manual

User manual for the device Emerson 644

Device: Emerson 644
Category: Weather Radio
Manufacturer: Emerson
Size: 2.07 MB
Added : 9/5/2014
Number of pages: 98
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Emerson 644. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Emerson 644.

For your convenience

If looking through the Emerson 644 user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Emerson 644 on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Emerson 644 to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Emerson 644 User manual - Online PDF
Advertisement
« Page 1 of 98 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Emerson 644 but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Emerson 644. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1

Reference Manual
00809-0400-4728, Rev AA
June 2011
Rosemount 644 Temperature Transmitter

with FOUNDATION fieldbus
www.rosemount.com

Summary of the content on the page No. 2

Summary of the content on the page No. 3

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 Table of Contents SECTION 1 Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1 Introduction Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Summary of the content on the page No. 4

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 APPENDIX A Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1 Functional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1 Specifications and Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2 Reference Data Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Summary of the content on the page No. 5

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 Rosemount 644 Temperature Transmitters Rosemount 644 Hardware Revision 9 ™ FOUNDATION Fieldbus Device Revision 2 Device Descriptor Revision 1 NOTICE Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure to thoroughly understand the contents before installing, using, or maintaining this product. The United States has two toll-free assistance numbers and o

Summary of the content on the page No. 6

Summary of the content on the page No. 7

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 Section 1 Introduction Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-2 Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-3 Return of Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-4 SAFETY MESSAGES Instructions and proced

Summary of the content on the page No. 8

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 OVERVIEW Manual This manual is designed to assist in the installation, operation, and maintenance of Rosemount 644 head mount and 644 rail mount. Section 1: Introduction • Transmitter and Manual Overview • Considerations • Return of Material Section 2: Installation • Mounting • Installation • Wiring • Power Supply • Commissioning Section 3: Configuration • Calibration • Hardware Maintenance • Diagnostic Messaging Appendix A: Speci

Summary of the content on the page No. 9

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 Refer to the following literature for a full range of compatible connection heads, sensors, and thermowells provided by Emerson Process Management. • Temperature Sensors and Assemblies Product Data Sheet, Volume 1 (document number 00813-0100-2654) • Temperature Sensors and Assemblies Product Data Sheet, Volume 2 (document number 00813-0200-2654) CONSIDERATIONS General Electrical temperature sensors such as RTDs and thermoco

Summary of the content on the page No. 10

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 Figure 1-1. 644 head mount 60 (108) Transmitter Connection Head Temperature Rise vs. Extension 50 (90) Length 40 (72) 815 °C (1500 °F) Oven Temperature 30 (54) 22 20 (36) 540 °C (1000 °F) Oven Temperature 10 (18) 250 °C (482 °F) Oven Temperature 0 3 4 5 6 7 8 9 3.6 Extension Length (in.) Example The transmitter specification limit is 85 °C. If the ambient temperature is 55 °C and the process temperature to be measured is 800 °

Summary of the content on the page No. 11

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 Section 2 Installation Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-7 Power Supply . . . . . .

Summary of the content on the page No. 12

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 Figure 2-1. Installation Flowchart Bench START Calibration? HERE FIELD INSTALL Does not apply to the 644 Set Failure Mode BASIC SETUP with FOUNDATION fieldbus Switch Set Sensor Type Mount Transmitter Set Number of Wires Wire Transmitter Set Units Power Transmitter Set Range Values FINISHED Set Damping VERIFY Simulate Sensor Input Within Specifications? Refer to Section 3: Configuration 2-2

Summary of the content on the page No. 13

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 MOUNTING Mount the transmitter at a high point in the conduit run to prevent moisture from draining into the transmitter housing. The 644 head mount installs: • In a connection head or universal head mounted directly on a sensor assembly • Apart from a sensor assembly using a universal head • To a DIN rail using an optional mounting clip. The 644 rail mount attaches directly to a wall or to a DIN rail. Mounting a 644H to a DIN

Summary of the content on the page No. 14

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 INSTALLATION Typical European Head Mount Transmitter with DIN Plate Style Sensor Installation 1. Attach the thermowell to the pipe or process container wall. Install and tighten the thermowell before applying process pressure. 2. Assemble the transmitter to the sensor. Push the transmitter mounting screws through the sensor mounting plate and insert the snap rings (optional) into the transmitter mounting screw groove. 3. Wire

Summary of the content on the page No. 15

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 Typical North American Head Mount Transmitter with Threaded Sensor Installation 1. Attach the thermowell to the pipe or process container wall. Install and tighten thermowells before applying process pressure. 2. Attach necessary extension nipples and adapters to the thermowell. Seal the nipple and adapter threads with silicone tape. 3. Screw the sensor into the thermowell. Install drain seals if required for severe environment

Summary of the content on the page No. 16

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 LCD Display Installation The LCD display provides local indication of the transmitter output and abbreviated diagnostic messages governing transmitter operation. Transmitters ordered with the LCD display are shipped with the meter installed. After-market installation of the meter can be performed if the transmitter has a meter connector (transmitter revision 5.5.2 or later). After-market installation requires the meter kit (pa

Summary of the content on the page No. 17

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 WIRING All power to the transmitter is supplied over the signal wiring. Use ordinary copper wire of sufficient size to ensure that the voltage across the transmitter power terminals does not drop below 9 Vdc. If the sensor is installed in a high-voltage environment and a fault condition or installation error occurs, the sensor leads and transmitter terminals could carry lethal voltages. Use extreme caution when making contact

Summary of the content on the page No. 18

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 Figure 2-6. Connecting a FOUNDATION fieldbus Host 6234 ft (1900 m) max (depending upon cable characteristics) System to a Transmitter Loop Terminators Integrated Power Conditioner and Filter Power (Trunk) Supply (The power supply, FOUNDATION filter, first fieldbus terminator, and Configuration configuration Tool tool are typically located in the control room.) Power/ Signal Devices 1 Wiring through 16 Sensor Connecti

Summary of the content on the page No. 19

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 RTD or Ohm Inputs The transmitters will accept a variety of RTD configurations, including 2-wire, 3-wire, or 4-wire. If the transmitter is mounted remotely from a 3-wire or 4-wire RTD, it will operate within specifications, without recalibration, for lead wire resistances of up to 60 ohms per lead (equivalent to 6,000 feet of 20 AWG wire). In this case, the leads between the RTD and transmitter should be shielded. If using on

Summary of the content on the page No. 20

Reference Manual 00809-0400-4728, Rev AA June 2011 Rosemount 644 Lead wire imbalance seen by the transmitter = 0.5  0.5  Basic error== ---------------------------------------------------------------------------------- 1.3 C  0.00385  /  C  100  Error due to amb. temp. var. of  25 °C  0.0039  /  C  25 C  0.5  = ------------------------------------------------------------------------------------------------------ -= 0.1266 C  0.00385  /  C  100  • Pt100 2-wire


Alternative user manuals
# User manual Category Download
1 Emerson 203-5751 User manual Weather Radio 0
2 Emerson 2468CB User manual Weather Radio 1
3 Emerson 2468CD User manual Weather Radio 2
4 Emerson 396P User manual Weather Radio 2
5 Emerson 396PVP User manual Weather Radio 0
6 Emerson PN-51-T56 User manual Weather Radio 0
7 Sony 443 User manual Weather Radio 1
8 3Com WL-548A User manual Weather Radio 0
9 Acu-Rite #75077 User manual Weather Radio 3
10 Acu-Rite 00305TG User manual Weather Radio 1
11 Acu-Rite 00593W User manual Weather Radio 4
12 3Com 6100 User manual Weather Radio 2
13 Sony User manual Weather Radio 2
14 Sony SRF-M85V User manual Weather Radio 102
15 Acu-Rite 00608BPDI User manual Weather Radio 9