Xantrex Technology XPL 56-1 user manual

User manual for the device Xantrex Technology XPL 56-1

Device: Xantrex Technology XPL 56-1
Category: Power Supply
Manufacturer: Xantrex Technology
Size: 0.64 MB
Added : 6/1/2013
Number of pages: 36
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Xantrex Technology XPL 56-1. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Xantrex Technology XPL 56-1.

For your convenience

If looking through the Xantrex Technology XPL 56-1 user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Xantrex Technology XPL 56-1 on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Xantrex Technology XPL 56-1 to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Xantrex Technology XPL 56-1 User manual - Online PDF
Advertisement
« Page 1 of 36 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Xantrex Technology XPL 56-1 but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Xantrex Technology XPL 56-1. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1


XPL 18-3
XPL 30-1
XPL 30-2
XPL 30-2D
XPL 30-2T
XPL 56-1

Operating Manual
XPL Series Linear
DC Power Supply

Summary of the content on the page No. 2

About Xantrex Xantrex Technology Inc. is a world-leading supplier of advanced power electronics and controls with products from 50 watt mobile units to one MW utility-scale systems for wind, solar, batteries, fuel cells, microturbines, and backup power applications in both grid-connected and stand-alone systems. Xantrex products include inverters, battery chargers, programmable power supplies, and variable speed drives that convert, supply, control, clean, and distribute electrical power

Summary of the content on the page No. 3

Introduction The XPL series provides medium levels of power (up to 125 watts) in a compact and attractive case style using linear regulation. High resolution controls allow precise setting of voltage and current levels which are indicated on accurate and highly legible digital meters. Excellent line and load regulation is matched by very low noise and good transient response. Improved power efficiency ensures that the units require no fan cooling. The XPL 30-2D has two independent iso

Summary of the content on the page No. 4

Specification MAIN OUTPUTS Voltage Range: 0V to 18V minimum (XPL 18-3) 0V to 30V minimum (XPL 30-1, XPL 30-2, XPL 30-2D, XPL 30-2T) 0 to 56V minimum (XPL 56-1) Current Range: 0A to 1A minimum (XPL 30-1) 0A to 1.1A minimum (XPL 56-1) 0A to 2A minimum (XPL 30-2, XPL 30-2D, XPL 30-2T) 0A to 3.3A minimum (XPL 18-3) Voltage Setting: By coarse and fine controls. Current Setting: By single logarithmic control. Operating Mode: Constant voltage or constant current with automatic cross-over.

Summary of the content on the page No. 5

GENERAL AC Input: 230V AC or 115VAC ± 10%, 50/60Hz. Installation Category II. Power Consumption: XPL 30-1: 85VA max; XPL 30-2/XPL 18-3/XPL 56-1: 160VA max; XPL 30-2D/XPL 30-2T: 320VA max. Operating Range: +5ºC to +40ºC, 20% to 80% RH. Storage Range: −40ºC to + 70ºC. Environmental: Indoor use at altitudes up to 2000m, Pollution Degree 2. Safety: Complies with EN61010-1. EMC: Complies with EN61326. Size: Singles: 140 x 160 x 295 mm (WxHxD). Dual & Triple: 260 x 160 x 295 mm (WxH

Summary of the content on the page No. 6

EMC This instrument has been designed to meet the requirements of the EMC Directive 89/336/EEC. Compliance was demonstrated by meeting the test limits of the following standards: Emissions EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and Laboratory Use. Test limits used were: a) Radiated: Class B b) Conducted: Class B c) Harmonics: EN61000-3-2 (2000) Class A; the instrument is Class A by product category. Immunity EN61326 (1998) EMC

Summary of the content on the page No. 7

Safety This power supply is a Safety Class I instrument according to IEC classification and has been designed to meet the requirements of EN61010-1 (Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use). It is an Installation Category II instrument intended for operation from a normal single phase supply. This instrument has been tested in accordance with EN61010-1 and has been supplied in a safe condition. This instruction manual contains some inform

Summary of the content on the page No. 8

Installation Mains Operating Voltage Check that the instrument operating voltage marked on the rear panel is suitable for the local supply. Should it be necessary to change the operating voltage, proceed as follows: 1. Ensure that the instrument is disconnected from the AC supply. 2. Remove the screws holding the case upper and handle and lift off the case upper. 3. Change the transformer connections following the diagram below: 230V Operation 115V Operation

Summary of the content on the page No. 9

Operation - Main Outputs Setting Up the Output With the POWER switch on (l) and the output OFF the output voltage and current limit can be accurately preset using the VOLTAGE and CURRENT controls; the left-hand meter shows the set voltage and the right-hand meter shows the set maximum current. When the output switch is switched ON, the lamp lights; the left-hand meter now shows the actual voltage and the right-hand meter the actual load current. Constant Voltage The output volta

Summary of the content on the page No. 10

Operation – Auxiliary Output Output Voltage The output voltage can be set between 1.5V and 5.0V using the front panel rotary adjustment beside the centre terminals. With the right-hand main output OFF, pressing the SHOW AUX PRESET button will display the AUX output voltage and current limit on the right-hand meters. Current Limit The output will go into current limit above 2A; the UNREG lamp lights in this condition. Protection The output has intrinsic short-circuit protection and is

Summary of the content on the page No. 11

Calibration Allow at least 5 minutes warm-up before commencing calibration. Access to Calibration Adjustments All adjustments are on the front panel control board; all the trimmers are adjusted from the reverse side of the board through holes in the board. To gain access to the board it is necessary to remove the top cover. WARNING! When the instrument is connected to its supply the removal of covers is likely to expose li

Summary of the content on the page No. 12

Current Calibration Switch output OFF. Set output voltage to nominally 2V. Set current control to minimum. Connect the DMM (set to Amps) and load in series across the output. Switch output ON. Adjust VR8/108 (offset compensation of current control error amp) for a reading of 0·003A ± ·001A on the DMM. Check that the CC LED is ON. Increase voltage controls and current control to maximum. Adjust load until the DMM reads 2·00A [3·00, 1·00] ± 0·02A. Adjust VR7/107 (measured output current)

Summary of the content on the page No. 13

Sécurité Cet instrument est de Classe de sécurité 1 suivant la classification IEC et il a été construit pour satisfaire aux impératifs EN61010-1 (impératifs de sécurité pour le matériel électrique en vue de mesure, commande et utilisation en laboratoire). Il s'agit d'un instrument d'installation Catégorie II devant être exploité depuis une alimentation monophasée habituelle. Cet instrument a été soumis à des essais conformément à EN61010-1 et il a été fourni en tout état de sécurité. C

Summary of the content on the page No. 14

Installation Tension d’utilisation secteur Vérifier que la tension opérationnelle de l'instrument indiquée sur le panneau arrière est appropriée pour l'alimentation locale. S'il s'avère nécessaire de modifier la tension opérationnelle, procéder de la manière décrite ci-dessous: 1. S'assurer que l'instrument est débranché de l'alimentation c.a. 2. Enlever les vis de retenue de la partie supérieure du boîtier et de la manette et soulever la partie supérieure du boîtier. 3. Changer les c

Summary of the content on the page No. 15

Fonctionnement - Sorties principales Réglage de la sortie Lorsque l’interrupteur POWER (alimentation) est en position (l) et la sortie est en position OFF (éteinte), il est possible de régler avec précision la limite de tension et de courant de sortie au moyen des commandes VOLTAGE (Tension) et CURRENT (Courant); l’appareil de mesure gauche indique la tension réglée et l’appareil droit le courant maximum réglé. Lorsque le commutateur de sortie est en position ON (marche), le témoin

Summary of the content on the page No. 16

Fonctionnement – Sortie Auxiliaire Tension de sortie La tension de sortie peut être réglée entre 1.5V et 5.0V en utilisant la commande rotative du panneau avant près des bornes centrales. Avec l'afficheur de l'état de la sortie principale droite, un appui sur le bouton SHOW AUX PRESET (montrer les réglages auxiliaire) provoquera l'affichage de la tension et de la limite du courant de la sortie auxiliaire sur l'afficheur droit. Limite de courant La sortie passe à la limite de courant a

Summary of the content on the page No. 17

Maintenance Le Constructeur ou ses agents à l'étranger répareront tout bloc qui tombe en panne. Si le propriétaire de l'appareil décide d'effectuer lui-même la maintenance, ceci doit uniquement être effectué par un personnel spécialisé qui doit se référer au manuel d’entretien que l'on peut se procurer directement auprès du Constructeur ou de ses agents à l'étranger. Nettoyage S'il faut nettoyer le bloc d'alimentation, utiliser un chiffon légèrement imbibé d'eau ou d'un détergent doux.

Summary of the content on the page No. 18

Sicherheit Dieses Gerät wurde nach der Sicherheitsklasse (Schutzart) I der IEC-Klassifikation und gemäß den europäischen Vorschriften EN61010-1 (Sicherheitsvorschriften für elektrische Mess-, Steuer-, Regel- und Laboranlagen) entwickelt. Es handelt sich um ein Gerät der Installationskategorie II, das für den Betrieb von einer normalen einphasigen Versorgung vorgesehen ist. Das Gerät wurde gemäß den Vorschriften EN61010-1 geprüft und wurde in sicherem Zustand geliefert. Die vorliegende

Summary of the content on the page No. 19

Installation Netzbetriebsspannung Sicherstellen, daß die auf der Geräterückwand angegebene Betriebsspannung mit der Versorgungsspannung am Ort übereinstimmt. Falls es erforderlich ist, die Betriebsspannung zu ändern, wie folgt vorgehen: 1. Sicherstellen, daß das Gerät vom Wechselstromnetz getrennt ist. 2. Schrauben entfernen, mit denen die obere Gehäusehälfte und der Handgriff befestigt sind und obere Gehäusehälfte abheben. 3. Transformatoranschlüsse entsprechend dem nachstehenden Dia

Summary of the content on the page No. 20

Betrieb - Hauptausgänge Einstellung des Ausgangs Bei eingeschaltetem POWER (Netzschalter I) und ausgeschaltetem Ausgang läßt sich die Ausgangsspannung und Strombegrenzung mit Hilfe der Knöpfe VOLTAGE (Spannung) und CURRENT (Strom) genau voreinstellen. Die linke Anzeige zeigt die eingestellte Spannung und die rechte den eingestellten Maximalstrom an. Ist der Ausgangsschalter aktiviert, leuchtet die ON (Ein)-Leuchte auf. Jetzt wird auf der linken Anzeige die Istspannung und an der


Alternative user manuals
# User manual Category Download
1 Xantrex Technology 230 VAC/50 User manual Power Supply 2
2 Xantrex Technology 40 A (XADC-40) User manual Power Supply 6
3 Xantrex Technology 18V 10A User manual Power Supply 1
4 Xantrex Technology 42V 10A User manual Power Supply 5
5 Xantrex Technology 500 User manual Power Supply 0
6 Xantrex Technology 300 PLUS User manual Power Supply 12
7 Xantrex Technology 60 A (XADC-60) User manual Power Supply 2
8 Xantrex Technology 600HD User manual Power Supply 0
9 Xantrex Technology 400R User manual Power Supply 1
10 Xantrex Technology ACCB User manual Power Supply 0
11 Xantrex Technology 700 PLUS User manual Power Supply 0
12 Xantrex Technology 80 A (XADC-80) User manual Power Supply 4
13 Xantrex Technology ACCB-L User manual Power Supply 0
14 Xantrex Technology ACCB-L-L1 User manual Power Supply 0
15 Xantrex Technology APG-XT User manual Power Supply 0
16 Sony AC U50A User manual Power Supply 5
17 Sony AC LS1 User manual Power Supply 2
18 Sony NP-FM30 User manual Power Supply 35
19 Sony NP FX101 User manual Power Supply 1
20 Sony NH-AAA-DA User manual Power Supply 2