Agilent Technologies 66332A user manual

User manual for the device Agilent Technologies 66332A

Device: Agilent Technologies 66332A
Category: Power Supply
Manufacturer: Agilent Technologies
Size: 1.77 MB
Added : 12/30/2013
Number of pages: 83
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Agilent Technologies 66332A. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Agilent Technologies 66332A.

For your convenience

If looking through the Agilent Technologies 66332A user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Agilent Technologies 66332A on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Agilent Technologies 66332A to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Agilent Technologies 66332A User manual - Online PDF
Advertisement
« Page 1 of 83 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Agilent Technologies 66332A but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Agilent Technologies 66332A. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1





Service Manual
Agilent Model 66332A
Dynamic Measurement DC Source
and Agilent Model 6632B, 6633B, 6634B
System DC Power Supply


















s1

Agilent Part No. 5962-8119 Printed in Malaysia
Microfiche No 6962-8120 October, 2003

Summary of the content on the page No. 2

Warranty Information CERTIFICATION Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Bureau of Standards, to the extent allowed by the Bureau's calibration facility, and to the calibration facilities of other International Standards Organization members. WARRANTY This Agilent Technologies hardware product is wa

Summary of the content on the page No. 3

Safety Summary The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer's failure to comply with these requirements. WARNING Servicing instructions are for use by service-trained personnel. To avoid dangerous el

Summary of the content on the page No. 4

Safety Symbol Definitions Symbol Description Direct current Alternating current Both direct and alternating current Three-phase alternating current Earth (ground) terminal Protective earth (ground) terminal Frame or chassis terminal Terminal is at earth potential (Used for measurement and control circuits designed to be operated with one terminal at earth potential.) Terminal for Neutral conductor on permanently installed equipment Terminal for Line conductor on permanently installed equipment O

Summary of the content on the page No. 5

Notice The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability, and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. This document contains proprie

Summary of the content on the page No. 6

Table of Contents Warranty Information 2 Safety Summary 3 Notice 4 Printing History 5 Instrument Identification 5 Table of Contents 6 INTRODUCTION 9 Organization 9 Safety Considerations 9 Related Documents 9 Revisions 10 Manual Revisions 10 Firmware Revisions 10 Electrostatic Discharge 10 VERIFICATION AND PERFORMANCE TESTS 11 Introduction 11 Test Equipment Required 11 Measurement Techniques 12 Setup for Most Tests 12 Electronic Load 13 Current-Monitoring Resistor 13 Operation Verification Tests

Summary of the content on the page No. 7

Power-on Self-test Failures 37 CV/CC Status Annunciators Troubleshooting 38 Bias and Reference Supplies 38 J307 Voltage Measurements 39 Manual Fan Speed Control 40 Disabling Protection Features 40 Post-repair Calibration 41 Inhibit Calibration Switch 41 Calibration Password 41 Initialization 42 ROM Upgrade 42 Identifying the Firmware 42 Upgrade Procedure 42 Disassembly Procedures 43 List of Required Tools 43 Cover, Removal and Replacement 44 A2 Interface Board, Removal and Replacement 44 Front

Summary of the content on the page No. 8

Summary of the content on the page No. 9

1 Introduction Organization This manual contains information for troubleshooting and repairing to the component level the Agilent Model 66332A Dynamic Measurement DC Source and the Agilent Model 6632B, 6633B, 6634B System DC Power Supplies. Hereafter all models will be referred to as the dc power supply. This manual is organized as follows: Chapter 1 Organization Performance tests Chapter 2 Chapter 3 Troubleshooting procedures Chapter 4 Principles of operation on a block-diagram level Chapter 5

Summary of the content on the page No. 10

1 - Introduction Revisions Manual Revisions If changes have been made to your power supply since the publication of this manual, a yellow Manual Change sheet may be supplied with the manual. It defines the differences between your power supply and the unit described in this manual. The yellow change sheet may also contain information for correcting errors in the manual. Note that because not all changes to the product require changes to the manual, there may be no update information requir

Summary of the content on the page No. 11

2 Verification and Performance Tests Introduction This document contains test procedures to verify that the dc power supply is operating normally and is within published specifications. There are three types of tests as follows: Built-in Self Tests These tests, run automatically when the power supply is turned on, check most of the digital circuits and the programming and readback DACs. Operation Verification These tests verify that the power supply is probably operating normally but do

Summary of the content on the page No. 12

2 - Verification and Performance Tests Resistor 1 ohm, 50 W Ohmite L50J1R0 (substitute for electronic 3 ohm, 100 W (Agilent 66332A/6632B) Ohmite RLS5R0 (adjustable) load if load is too noisy 24 ohm, 100 W (Agilent 6633B) Ohmite RLS25R (adjustable) for CC PARD test) 99 ohm, 100 W (Agilent 6634B) Ohmite RLS100 (adjustable) 1k ohm, 5%, 3W (all models) Agilent 0813-0001 Oscilloscope Sensitivity: 1 mV Agilent 54504A or equivalent Bandwidth Limit: 20 MHz Probe: 1:1 with RF tip RMS Voltme

Summary of the content on the page No. 13

Verification and Performance Tests - 2 Electronic Load Many of the test procedures require the use of a variable load capable of dissipating the required power. If a variable resistor is used, switches should be used to either; connect, disconnect, or short the load resistor. For most tests, an electronic load can be used. The electronic load is considerably easier to use than load resistors, but it may not be fast enough to test transient recovery time and may be too noisy for the noise (

Summary of the content on the page No. 14

2 - Verification and Performance Tests Table 2-2. Programming Ratings Model Voltage Rating Full Scale Rating Current Rating Full Scale Rating Agilent 20 V 20.020 V 5 A 5.0045 A 66332A/6632B 50 V 50.045 V 2 A 2.002 A Agilent 6633B 100 V 100.1 V 1 A 1.001 A Agilent 6634B Constant Voltage (CV) Tests CV Setup If more than one meter or if a meter and an oscilloscope are used, connect each to the terminals by a separate pair of leads to avoid mutual coupling effects. For constant voltage dc tests, con

Summary of the content on the page No. 15

Verification and Performance Tests - 2 e. Open the load and again record the DVM voltage reading. The difference between the DVM readings in steps (d) and (e) is the load effect voltage, and should not exceed the value listed in the performance test record chart for the appropriate model under CV LOAD EFFECT. CV Source Effect This test measures the change in output voltage that results from a change in ac line voltage from the minimum to maximum value within the line voltage specifications. a. T

Summary of the content on the page No. 16

2 - Verification and Performance Tests Transient Recovery Time This test measures the time for the output voltage to recover to within the specified value following a 50% change in the load current. Loading tttt Transient t v v t Unloading Transient Figure 2-2. Transient Waveform a. Turn off the supply and connect the output as in Figure 2-1a with the oscilloscope across the +S and the -S terminals. b. Turn on the supply and program the output voltage to the full-scale value and the cu

Summary of the content on the page No. 17

Verification and Performance Tests - 2 e. Program the output current to full-scale . f. Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to amps and record this value (Iout). Also, record the current reading that appears on the front panel display. The readings should be within the limits specified in the performance test record card for the appropriate model under CC PROGRAMMING @ FULL-SCALE. Current Sink (-CC) Operation This tes

Summary of the content on the page No. 18

2 - Verification and Performance Tests c. Set up voltmeter to take measurements in the statistical mode as follows: Press Shift key, f0, Shift key, N Press ^ (up arrow) until MATH function is selected, then press >. Press ^ (up arrow until STAT function is selected then press (ENTER). d. Set up voltmeter to read the average of the measurements as follows: Press Shift key, f1, Shift key, N. Press down arrow until RMATH function is selected, then press >. Press ^ (up arrow) until MEAN function is

Summary of the content on the page No. 19

Verification and Performance Tests - 2 CC Noise (PARD) Periodic and random deviations (PARD) in the output combine to produce a residual ac current, as well, as an ac voltage superimposed on the dc output. Constant current (CC) PARD is specified as the rms output current in a frequency range 20 Hz to 20 Mhz with the supply in CC operation. a. Turn off the supply and connect the load, monitoring resistor, and rms voltmeter across the monitoring resistor as shown in Figure 2-1a. The Current Monito

Summary of the content on the page No. 20

2 - Verification and Performance Tests Performance Test Record Forms Model Agilent 66332A OR Agilent6632B Report No _______________ Date __________________ Test Description Minimum Results* Maximum Specs. Specs. Constant Voltage Tests Voltage Programming and Readback Low Voltage (0V) Vout - 10 mV __________ + 10 mV Front Panel Display Readback Vout - 3 mV __________ Vout + 3 mV High Voltage (Full Scale) Vout 19.980 V __________ 20.020 V Front Panel Display Readback Vou


Alternative user manuals
# User manual Category Download
1 Agilent Technologies 6032A User manual Power Supply 16
2 Agilent Technologies 6012B User manual Power Supply 4
3 Agilent Technologies 6038A 6010A User manual Power Supply 3
4 Agilent Technologies 6023A, 6028A User manual Power Supply 5
5 Agilent Technologies 6035A User manual Power Supply 8
6 Agilent Technologies 6011A User manual Power Supply 1
7 Agilent Technologies 6023A User manual Power Supply 0
8 Agilent Technologies 6611C User manual Power Supply 19
9 Agilent Technologies 6031A User manual Power Supply 1
10 Agilent Technologies 6625A User manual Power Supply 0
11 Agilent Technologies 6632B User manual Power Supply 22
12 Agilent Technologies 6634B User manual Power Supply 5
13 Agilent Technologies 6629A User manual Power Supply 0
14 Agilent Technologies 6614C User manual Power Supply 7
15 Agilent Technologies 6628A User manual Power Supply 1
16 Sony AC U50A User manual Power Supply 5
17 Sony AC LS1 User manual Power Supply 2
18 Sony NP-FM30 User manual Power Supply 35
19 Sony NP FX101 User manual Power Supply 1
20 Sony NH-AAA-DA User manual Power Supply 2