Hanna Instruments HI8519 user manual

User manual for the device Hanna Instruments HI8519

Device: Hanna Instruments HI8519
Category: Microphone
Manufacturer: Hanna Instruments
Size: 0.39 MB
Added : 7/16/2014
Number of pages: 40
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for Hanna Instruments HI8519. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with Hanna Instruments HI8519.

For your convenience

If looking through the Hanna Instruments HI8519 user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual Hanna Instruments HI8519 on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual Hanna Instruments HI8519 to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
Hanna Instruments HI8519 User manual - Online PDF
Advertisement
« Page 1 of 40 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual Hanna Instruments HI8519 but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to Hanna Instruments HI8519. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1

°C pH
HI 8417
BUFFER VALUES 20°C
°C
ON d1 NBS pH 6,88 CAL
t1 STD pH 7,02 pH
d2 NBS pH 4,00
mV
t2 STD pH 4,00 °C
CON
OFF d3 NBS pH 9,22
t3 STD pH 10,05
Instruction Manual
HI8417 - HI8519
HI8520 - HI8521
Microprocessor
Bench-top
pH Meters
These Instruments are in
http://www.hannainst.com Compliance with the CE Directives

Summary of the content on the page No. 2

Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instru- ment. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with CSA, UL and (EN 50081-1 and EN 50082-1) directives. TABLE OF CONTENTS PRELIMINARY EXAMINATION 3 GENERAL DESCRIPTION 3 FUNCTIONAL DESC

Summary of the content on the page No. 3

PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your dealer. Note: Save all packing material until you are sure that the instru- ment functions correctly. All defective items must be returned in the original packing with the supplied accessories. GENERAL DESCRIPTION HI8417, HI8519, HI8520 and HI8521 Hanna Instruments bench- top pH meters are designed for simplicity of u

Summary of the content on the page No. 4

HI8520 pH/°C meter automatically recognizes 3 standard calibration buffer values (4.01, 7.01 and 10.01 pH). HI8520 comes supplied with HI1332B plastic-body combination pH electrode, HI7669/2W temperature probe (with 1 meter/ 3.3' cable), 12VDC adapter (HI710005 or HI710006) and a dust cover. HI8521 pH/mV/°C meter automatically recognizes 3 standard cali- bration buffer values (4.01, 7.01 and 10.01 pH). HI8521 comes supplied with HI1332B plastic-body combination pH electrode, HI7669/2W temperatur

Summary of the content on the page No. 5

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI8417 THE FRONT PANEL OF HI8417 °C pH HI 8417 BUFFER VALUES 20°C °C CAL ON d1 NBS pH 6,88 t1 STD pH 7,02 pH d2 pH 4,00 NBS pH 4,00 mV t2 STD °C CON OFF d3 NBS pH 9,22 pH 10,05 t3 STD ON / OFF To turn the instrument ON and OFF. (Warning: Push the OFF key before unplugging instrument. If unplugged before turned off, the meter's memory will be erased). CAL To enter the calibration mode. CON To confirm the calibration data. °C / °C To set temperature for m

Summary of the content on the page No. 6

THE REAR PANEL OF HI8417 TEMPERATURE DC POWER PROBE 12V to 20V COMBINED RECORDER pH or ORP REFERENCE + RESET 0-2 VOLT ELECTRODES ELECTRODE 4 1 23 5 6 1. BNC socket for pH and ORP electrode. 2. Reference Jack socket for reference electrode. 3. RESET button. 4. Recorder output connections. 5. Temp. Probe socket (for HI7669/2W). 6. DC Power Socket (for HI710005 or HI710006). Electrode and probe connections For combination pH or ORP electrodes (with internal reference) plug the electrode's BNC to th

Summary of the content on the page No. 7

instrument's power supply is disconnected before the meter is switched off. Press the RESET button and restart the entire operation. Power connection Plug the 12VDC adapter (HI710005 or HI710006) into the DC socket (#6). Note: Make sure the main line is protected by a fuse. Note: HI8417 uses a rechargeable battery to retain the pH calibra- tion data. It lasts approximately one month when fully charged. The battery is automatically charged any time the meter is plugged-in (either turned on or off

Summary of the content on the page No. 8

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI8519 THE FRONT PANEL OF HI8519 pH HI 8519 OFFSET SLOPE TEMPERATURE °C OFFSET SLOPE TEMPERATURE °C Temperature Buffer Values (pH) °C °F 4.01 7.01 10.01 0 32 4.01 7.13 10.32 10 50 4.00 7.07 10.18 ON 20 68 4.00 7.03 10.06 pH mV °C 30 86 4.02 7.00 9.96 OFF 40 4.04 6.98 9.88 104 ON/OFF To switch the meter ON and OFF. pH To display pH values. mV To display mV values. °C To display temperature values set by the temper

Summary of the content on the page No. 9

Electrode and probe connections For combination pH or ORP electrodes (with internal reference) plug the electrode's BNC to the socket provided (#1). For an electrode with a separate reference, connect the measuring electrode's BNC to the BNC socket (#1) and the reference electrode's jack to the socket provided (#2). Power connection Plug the 12VDC adapter (HI710005 or HI710006) into the DC socket (#3). Note: Make sure the main line is protected by a fuse. SPECIFICATIONS HI 8519 Range pH 0.00 to

Summary of the content on the page No. 10

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI8520 THE FRONT PANEL OF HI8520 ERROR SYMBOLS Temperature Buffer Values (pH) ON CAL E1: pH out of range °C °F 4.01 7.01 10.01 OFF E2: °C out of range 0 32 4.01 7.13 10.32 °C E4: wrong pH 7 10 50 4.00 7.07 10.18 20 68 4.00 7.03 10.06 E5: wrong pH 4/10 CON RANGE 30 86 4.02 7.00 9.96 40 4.04 6.98 9.88 104 ON / OFF To switch instrument ON and OFF. RANGE To display pH or temperature measurements. CAL/CON CAL to enter the calibration procedure and CON to con

Summary of the content on the page No. 11

off. Press the RESET button and restart the entire operation. Power connection Plug the 12VDC adapter (HI710005 or HI710006) into the socket on the back of the meter (#5). Note: Make sure the main line is protected by a fuse. Note: HI8520 uses a rechargeable battery to retain the pH calibra- tion data. It lasts approximately one month when fully charged. The battery is automatically charged any time the meter is plugged-in (either turned on or off). SPECIFICATIONS HI 8520 Range pH 0.00 to 14.00

Summary of the content on the page No. 12

FUNCTIONAL DESCRIPTION & SPECIFICATIONS OF HI8521 THE FRONT PANEL OF HI8521 ERROR SYMBOLS Temperature Buffer Values (pH) ON CAL E1: pH out of range °C °F 4.01 7.01 10.01 OFF E2: °C out of range 0 32 4.01 7.13 10.32 °C E3: mV out of range 10 50 4.00 7.07 10.18 20 68 4.00 7.03 10.06 E4: wrong pH 7 CON RANGE 30 86 4.02 7.00 9.96 40 4.04 6.98 9.88 E5: wrong pH 4/10 104 ON / OFF To switch instrument ON and OFF. RANGE To display pH / ORP / Temperature measurements. CAL/CON CAL to enter the calibration

Summary of the content on the page No. 13

instrument's power supply is disconnected before the meter is switched off. Press the RESET button and restart the entire operation. Power connection Plug the 12VDC adapter (HI710005 or HI710006) into the socket on the back of the meter (#5). Note: Make sure your main line is protected by a fuse. Note: HI8521 uses a rechargeable battery to retain the pH calibra- tion data. It lasts approximately one month when fully charged. The battery is automatically charged any time the meter is plugged-in (

Summary of the content on the page No. 14

OPERATIONAL GUIDE TAKING PH MEASUREMENTS Connect the instrument to the 12VDC adapter (HI710005 or HI710006). Plug the adapter into the mains. Make sure that the instrument has been calibrated for pH before taking pH measurements (see page 18). For HI8519 only: ON • Connect the pH electrode to the meter and switch OFF the instrument on by pressing the ON/OFF key. • Immerse the electrode tip (4 cm/1½") into the sample and shake briefly. • Take the temperature of the solution with a ChecktempC or a

Summary of the content on the page No. 15

4 cm (1½") • In HI8417 the temperature reading is also displayed on the left LCD. °C pH Note: the electrode should be submerged approximately 4 cm (1½") into the solution. If the temperature probe is used, it should be located as close to the pH electrode as possible. If measurements are taken in different samples successively, it is recommended that the electrode is rinsed thoroughly for better condi- tioning and to eliminate cross-contamination of the sample. For the rinsing process, it is re

Summary of the content on the page No. 16

4 cm (1½") • When the temperature is set manually, ºC "°C" symbol blinks. With the tempera- ture probe connected, the "°C" symbol is displayed continuously. • Use the UP or DOWN arrow keys to °C °C manually adjust the display reading to the value of the reference ther- mometer. TAKING TEMPERATURE MEASUREMENTS (NOT FOR HI 8519) • Taking a temperature measurement is very easy. • Turn the instrument ON and press the RANGE key RANGE (for HI8520 and HI8521 only) to get into tem- perature mode. • Make

Summary of the content on the page No. 17

For HI8417 and HI8521 • Connect the ORP electrode to the meter and submerge the tip (4 cm / 1½") into the sample to be tested. Note: ORP measurements are taken without temperature compensa- tion. pH pH RANGE • Press the pH/mV key (HI8417) or mV mV RANGE key (HI8521) once to enter the mV mode. • Allow a few minutes for the reading to stabilize. • The display will indicate the absolute mV mV value (positive or negative). • When values are less than ±400 mV, tenths of mV are dis- played; for values

Summary of the content on the page No. 18

4 cm (1½") PH CALIBRATION IMPORTANT The instrument's pH range should be re-calibrated: • When the meter is new. • Whenever the pH electrode is replaced. • At least once a month. • After use in aggressive chemicals. • After cleaning procedure and changing the reference electrolyte. • For greatest accuracy. INITIAL PREPARATION Pour small quantities (up to 4cm / 1½" level) of pH 7.01 (HI7007 or HI8007) and pH4.01 (HI7004 or HI8004) or pH 10.01 (HI7010 or HI8010) solution into clean beakers. If HI 7

Summary of the content on the page No. 19

Procedure: ON • Switch the instrument on after connecting the pH electrode and the temperature probe. • Immerse the pH electrode and the ATC probe into a neutral pH buffer solution (pH 7.01 for technical buffers, or pH 6.86 for NBS buffers). Shake briefly and wait 1-2 minutes for thermal equilibrium. CAL • Press the CAL key. The symbol "d1" will appear on the display. For calibration with technical buffers (pH 7.01), press CAL key again and "t1" will appear on the display. (Note: the "pH" symbol

Summary of the content on the page No. 20

• The meter automatically recognizes the buffer value and prompts for the second at "d2" or "t2" (= pH 4.01, acid) or "d3" or "t3" (pH 9.18 and pH10.01 respectively, alkaline). CON • Press the CON key to confirm the pH value. • The display will indicate the temperature compen- sated buffer value e.g. if the temperature is at 25°C, the display will show "pH 4.01" or "10.01" or "pH 9.18", depending on the solution being used. If the temperature is at 20°C, the display will show respectively "pH 4.


Alternative user manuals
# User manual Category Download
1 Hanna Instruments HI8417 User manual Microphone 23
2 Hanna Instruments HI8520 User manual Microphone 13
3 Hanna Instruments HI8521 User manual Microphone 5
4 Hanna Instruments Microprocessor Bench-top pH Meters User manual Microphone 2
5 Edelbrock 15541 User manual Microphone 0
6 Sony DWA-01D User manual Microphone 0
7 Sony ECM DM5P User manual Microphone 6
8 Sony ECM Z37C User manual Microphone 29
9 Sony DWT-B01 User manual Microphone 27
10 Sony ECM DS70P User manual Microphone 38
11 Sony ECM 719 User manual Microphone 276
12 Sony ECM MS908C User manual Microphone 53
13 Sony CMU-BR100 User manual Microphone 32
14 Sony ECM R300 User manual Microphone 3
15 Sony ECM 360 User manual Microphone 2