AQ1000 / AQ2000
ALTERNATING PRESSURE
BUBBLE PAD AND PUMP SYSTEM
USER MANUAL
Important: Do not operate the Mattress System
without first reading and understanding this manual!
Save this manual for future use.
The latest version of this manual can be found online
at www.grahamfield.com.
AQ1000-INS-LAB-RevD13
Summary of the content on the page No. 2
CONTENTS INTRODUCTION ...................................................................................... 3 INTENDED USE OF THIS DEVICE ........................................................ 3 CONTRAINDICA TION ............................................................... ....... 3 DANGER ................................................................................................. 4 W ARNING ........................................................................................
Summary of the content on the page No. 3
INTRODUCTION Use this manual for setup and operation of the Lumex AQ1000 / AQ2000 Alternating Pressure Pad and Pump System. Read all instructions before using the pad and pump system. Save this manual for future reference. INTENDED USE OF THIS DEVICE The intended use of the Alternating Pressure Pad and Pump System, in either a homecare or long-term care setting, is the treatment of pressure ulcers while optimizing user comfort, and pain management as prescribed by a physician. Contraindic
Summary of the content on the page No. 4
CAUTION: Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor or moderate personal injury . s NOTICE: Indicates a potential hazard or unsafe practice that, if not avoided, could result in product or property damage. Info: Provides application recommendations or other useful information to ensure that you get the most from your product. DANGER To reduce the risk of electrocution: DANGER: Always unplug this product immediately after using.
Summary of the content on the page No. 5
W ARNING: Never operate this product if a) it has a d amaged cord or plug. b) it is not w orking properly . c) it has been dropped or damaged. d) it has been dropped into water . Return the product to your Graham-Field authorized distributor . W ARNING: Keep the cord away from heated surfaces. W ARNING: Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the openings may be blocked. Keep the air opening free of li
Summary of the content on the page No. 6
EQUIPMENT SYMBOLS Attention! Consult accompanying documents (this manual) “BF”symbol: Indicates that this product is in accordance with the degree of protection against electric shock for type BF equipment Grounding terminal Class II (for UL only) Disposal of Electrical & Electronic Equipment (WEEE): Do not treat this product as household waste. For more detailed information with regard to returning and recycling this product, please consult your local city office, household waste disposal
Summary of the content on the page No. 7
PRODUCT DESCRIPTION BUBBLE PAD AND PUMP SYSTEM air tubes bubble pad pump bubble pad and pump system PUMP front panel air tube connectors wire hangers fuses (220V only) pump (AQ2000 shown) AQ1000-INS-LAB-RevD13 7
Summary of the content on the page No. 8
FRONT PANEL AQ1000 ON/OFF switch air tube connectors AQ1000 pump front panel AQ2000 ON/OFF switch pressure control dial AQ2000 pump front panel Front panel pressure control dial (AQ2000 only): The pressure control dial controls the air pressure output. To increase pressure, turn the dial clockwise. To decrease pressure, turn the dial counter-clockwise. 8 AQ1000-INS-LAB-RevD13
Summary of the content on the page No. 9
INSTALLATION UNPACKING 1. Before unpacking, check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field authorized distributor. 2. Remove all loose packing from the carton. 3. Carefully remove all components from the carton. 4. Inspect all components. If damage is evident, please notify the carrier and your Graham-Field authorized distributor. PUMP AND BUBBLE PAD INSTALLATION 1. Remove all sheets, blankets, and pillows from t
Summary of the content on the page No. 10
5. Connect one end of the air tubes to the bubble pad: slide the ends of the air tubes over the connectors on the top of the foot end of the bubble pad. 6. Connect the other end of the air tubes to the pump connectors; ensure that the ends of the air tubes are firmly fitted over the bubble pad and pump connectors before continuing. s NOTICE: Ensure that the air tubes are not kinked or tucked under the bubble pad or mattress. W ARNING: Locate all cords so that they will not be stepped o
Summary of the content on the page No. 11
OPERATION 1. Turn the ON/OFF switch to the ON position. The pump will begin to deliver air into the bubble pad. 2. Replace all sheets, blankets, and pillows on the bed. Info: if the pressure level seems low, check for any leakage in the air tubes or bubble pad. Replace any damaged tubes. If necessary, contact your Graham-Field authorized distributor for repair or replacement. AQ2000 ONLY: PRESSURE ADJUSTMENT Adjust the pressure control dial for comfort. Info: For maximum effectiveness, patie
Summary of the content on the page No. 12
MAINTENANCE CLEANING Perform the following cleaning procedures before use and as necessary thereafter. Pump Wipe the pump with a clean, damp cloth and mild detergent, and keep it away from dust. Air dry. s NOTICE: Do not use phenolic products or corrosive or powdered cleansers to clean the pump. s NOTICE: Do not immerse or soak the pump. Bubble pad Wipe the bubble pad with a clean, damp cloth and mild detergent. The bubble pad may also be cleaned using a 10% solution of sodium hypochlorit
Summary of the content on the page No. 13
Fuse replacement (220V system only) tool needed: small screwdriver 1. If you suspect a blown fuse, disconnect the plug from the wall outlet immediately. fuse fuse cover fuse location 2. Use a small screwdriver to remove the cover of the fuse holder (fuse locations are shown in picture above). 3. Insert a new fuse of the correct rating (T1A/250V). 4. Replace the fuse holder cover. Ensure that fuse holder cover is securely installed. STORAGE 1. Fold the bubble pad loosely. 2. Store bubble pad a
Summary of the content on the page No. 14
SPECIFICATIONS Model AQ1000 (Standard) AQ2000 (Deluxe) Power Supply 120V: 120V AC, 60 Hz, 1A 220V: 230VAC, 50 Hz, 7W 120V: N/A Fuse Rating 220V: T1A, 250V Pressure 1.9 psi (13.10 kPa) 1.0 - 2.0 psi (6.89-13.79 kPa) 120V: 4.5 LPM 120V: 4-5 LPM Air Output 220V: 5.0 LPM 220V: 5-6 LPM Cycle Time 120V: 5 min. 220V: 6 min. Power Cord 10 ft. (304.8 cm) Length Pump Unit Size 8.5 x 4.25 x 3.62 in. 9.75 x 5.12 x 3.88 in. (21.59 x 10.80 x 9.19 cm) (24.77 x 13.00 x 9.86 cm) Pump Unit Weight 2.9 lb. (6.38 kg
Summary of the content on the page No. 15
WARRANTY GF Health Products, Inc. warrants the Lumex AQ1000 / AQ2000 Alternating Pressure Pad and Pump System for defects in workmanship and materials for the terms listed below. Pump: one year Pad: 90 days If a product is deemed to be under warranty, GF Health Products, Inc. shall provide, at its option, (1) replacement of any defective part or product or (2) a credit of the original selling price made to GF Health Products, Inc.’s initial customer. The warranty does not include any la
Summary of the content on the page No. 16
GF Health Products, Inc. 2935 Northeast Parkway Atlanta, Georgia 30360 tel: 770-368-4700 fax: 770-368-2386 www.grahamfield.com © 2008 GF Health Products, Inc.