VXI TVS600 user manual

User manual for the device VXI TVS600

Device: VXI TVS600
Category: Electronic Accessory
Manufacturer: VXI
Size: 0.06 MB
Added : 4/20/2014
Number of pages: 24
Print the manual

Download

How to use this site?

Our goal is to provide you with a quick access to the content of the user manual for VXI TVS600. Using the online preview, you can quickly view the contents and go to the page where you will find the solution to your problem with VXI TVS600.

For your convenience

If looking through the VXI TVS600 user manual directly on this website is not convenient for you, there are two possible solutions:

  • Full Screen Viewing - to easily view the user manual (without downloading it to your computer), you can use full-screen viewing mode. To start viewing the user manual VXI TVS600 on full screen, use the button Fullscreen.
  • Downloading to your computer - You can also download the user manual VXI TVS600 to your computer and keep it in your files. However, if you do not want to take up too much of your disk space, you can always download it in the future from ManualsBase.
VXI TVS600 User manual - Online PDF
Advertisement
« Page 1 of 24 »
Advertisement
Print version

Many people prefer to read the documents not on the screen, but in the printed version. The option to print the manual has also been provided, and you can use it by clicking the link above - Print the manual. You do not have to print the entire manual VXI TVS600 but the selected pages only. paper.

Summaries

Below you will find previews of the content of the user manuals presented on the following pages to VXI TVS600. If you want to quickly view the content of pages found on the following pages of the manual, you can use them.

Abstracts of contents
Summary of the content on the page No. 1

Reference
TVS600 & TVS600A Series
Waveform Analyzers
070-9918-00
This document supports firmware version 1.5 and
above.

Summary of the content on the page No. 2

Copyright  Tektronix, Inc. All rights reserved. Licensed software products are owned by Tektronix or its suppliers and are protected by United States copyright laws and international treaty provisions. Use, duplication, or disclosure by the Government is subject to restrictions as set forth in subparagraph (c)(1)(ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.227-7013, or subparagraphs (c)(1) and (2) of the Commercial Computer Software – Restricted Rights clause a

Summary of the content on the page No. 3

Quick Reference to TVS600 & TVS600A Commands This manual provides an abbreviated reference to the commands of TVS600 and TVS600A Series Waveform Analyzers. Use this reference to find the correct syntax for the commands. The commands are grouped by subsystem node and listed in alphabetical order. In general, the term TVS600A includes the TVS600A and the TVS600 models. Some commands apply only to TVS600A models; these commands appear in bold type. Syntax Commands ending with “?” are queries only.

Summary of the content on the page No. 4

Quick Reference to TVS600 & TVS600A Commands Commands can also include the argument-type enumerated; for this type, the alternative argument forms are listed as follows: Form: Example: Sometimes the forms are combined, for example: which means: (one of CHAN 1, CHAN 2 , CHAN 3, or CHAN 4) OR AATS OR (one of CALC1, CALC2, CALC3, or CALC4) For detailed descriptions of the commands an

Summary of the content on the page No. 5

Quick Reference to TVS600 & TVS600A Commands Commands Some commands in the following list apply only to TVS600A models; these commands appear in bold type. AADVance Commands in the [SENSe:]AADVance subsystem control how auto-advance acquisition records are acquired and transferred to a VXIbus controller. AADVance [:STATe][?] :COUNt[?] :RECord :COUNt[?] :STARt[?] ABORT This command in the ABORt subsystem operates in combination with the ARM and TRIGger subsystems to st

Summary of the content on the page No. 6

Quick Reference to TVS600 & TVS600A Commands AVERage Commands in the [SENSe:]AVERage subsystem control averaging (on incoming waveforms) in the SENSe block, which occurs prior to processing of the waveforms by the CALCulate block. AVERage [:STATe][?] :COUNt[?] :TYPE[?] CALCulage Commands in the CALCulate subsystem process and perform measurements on acquisition records. These calculations are typically performed immediately after the waveform

Summary of the content on the page No. 7

Quick Reference to TVS600 & TVS600A Commands :HPASs[?] :LPASs[?] :SPAN[?] :SREJection[?] :STARt[?] :STATe[?] :STOP[?] :TWIDth[?] :FORMat[?] :IMMediate[?] :INTegral :STATe[?] :PATH[?] :EXPRession[?] :SMOothing [:STATe][?] :POINts[?] :TRANsform :FREQuency :ST

Summary of the content on the page No. 8

Quick Reference to TVS600 & TVS600A Commands :EDGE[?] :GATE[?] :METHod[?] ABSolute | RELative :STARt [:ABSolute][?] :RELative[?] :STOP [:ABSolute][?] :RELative[?] :HIGH[?] :HMEThod[?] :LOW[?] :LMEThod[?] :HREFerence [:ABSolute][?] :RELative[?] :LREFerence [:ABSolute][?] :RELative[?] :MREFerence [:ABSolute][?] :HYSTeresis[?] :RELative[?] :RM

Summary of the content on the page No. 9

Quick Reference to TVS600 & TVS600A Commands CALibration Commands in the CALibration subsystem run the waveform analyzer self-calibration functions. CALibration [:ALL][?] :RESults [:CODE]? :VERBose? :PROBe[n][?] :RESults? DATA? Commands in the [SENSe:]FUNCtion and [SENSe:]DATA subsystems control sense functions and provide a means of accessing the data produced by the sense functions. DATA? :PREamble? FORMat Commands in the FORMat subsystem set the f

Summary of the content on the page No. 10

Quick Reference to TVS600 & TVS600A Commands :DINTerchange[?] FUNCtion Commands in the [SENSe:]FUNCtion and [SENSe:]DATA subsystems control sense functions and provide a means of accessing the data produced by the sense functions. FUNCtion [:ON][?] :ALL :COUNt? :OFF[?] :ALL :COUNt? :CONCurrent[?] :STATe? , INITiate Commands in the INITiate subsystem operate in combination with the ARM

Summary of the content on the page No. 11

Quick Reference to TVS600 & TVS600A Commands :FILTer [:LPASs] [:STATe][?] :FREQuency[?] :IMPedance[?] :PROBe :ATTenuation? :IDENtification? :OFFSet? :PROTection :STATe[?] MEMory Commands in the MEMory subsystem store and retrieve instrument settings. The waveform analyzer can store the current instrument state in any of ten nonvolatile memory locations or in a file on your controller. MEMory :DATA[?] :NSTates? :STATe :CATalog? :DEFine? <

Summary of the content on the page No. 12

Quick Reference to TVS600 & TVS600A Commands OUTPut Commands in the OUTPut subsystem route signals to the VXIbus trigger lines and enable the probe compensation signals and reference signals. OUTPut :ECLTrg [:STATE][?] :POLarity[?] :SOURce[?] :PCOMpensate [:STATE][?] FUNCtion[?] :REFerence [:STATE][?] :FUNCtion[?] :TTLTrg [:STATE][?] :POLarity[?]

Summary of the content on the page No. 13

Quick Reference to TVS600 & TVS600A Commands [SENSe:] Commands in the [SENse:] system appear in the references that follow: [SENSe:] see AADVance see AVERage see DATA see FUNCtion see ROSCillator see SWEep see VOLTage STATus Commands in the STATus subsystem can determine the state of the waveform analyzer and control what events can interrupt the system controller. For an overview of the status and event reporting system, refer to your user manual. STATus :OPERation [:EVENt]? :CONDition?

Summary of the content on the page No. 14

Quick Reference to TVS600 & TVS600A Commands :QUEStionable [:EVENt]? :CONDition? :ENABle[?] | :NTRansition[?] | :PTRansition[?] | :QENable :NTRansition[?] | :PTRansition[?] | :SESR :QENable[?] | SWEep Commands in the [SENSe:]SWEep subsystem control the acquisition time base for all VOLTage[n] acquisitions. SWEep :OFFSet :POINts[?] :TIME[?] :OREFerence :LOCation[?] :POINts[?] :TIME[?] :TINTerva

Summary of the content on the page No. 15

Quick Reference to TVS600 & TVS600A Commands SYSTem Commands in the SYSTem subsystem program utility functions and return version information about the waveform analyzer. SYSTem :AUToset :VOLTage :SWEep :TRIGger :BDATe[?] , , :CDATe[?] , , :COMMunicate :SERial :BAUD[?] :CONTrol :DCD[?] :RTS[?] :ECHO[?] :ERESponse[?] :LBUFfer[?] :PACE[?] :P

Summary of the content on the page No. 16

Quick Reference to TVS600 & TVS600A Commands :ERRor [:NEXT]? :ALL? :CODE [:NEXT]? :ALL? :COUNt? :PROTect[?] :SECurity :IMMediate :SET[?] :VERSion? TEST Commands in the TEST subsystem execute the internal self-tests of the waveform analyzer module and return the pass or fail status. TEST [:ALL][?] :RESults [:CODE]? :VERBose? :STOP 14 TVS600 & TVS600A Series Waveform Analyzers Reference

Summary of the content on the page No. 17

Quick Reference to TVS600 & TVS600A Commands TRACe? Commands in the Trace subsystem transfer acquisition records or measurement results and their data preambles from the waveform analyzer to the VXIbus controller or to internal REF storage locations. (The data preamble includes scaling and length informa- tion for the associated data record.) See the FORMAT subsystem for setting the format for data transfers. The query does not return data until pending acquisitions or calculations complete. Hen

Summary of the content on the page No. 18

Quick Reference to TVS600 & TVS600A Commands TRIGger[:A] Commands in the TRIGger[:A] subsystem operate in combination with the ARM, INITiate, TRIGger:B and ABORt subsystems to trigger acquisitions. The defined alias for the SCPI trigger :SEQuence[1] is A. TRIGger [:A] :ATRigger [:STATe][?] :COUPling[?]AC | DC : :DEFine? :DELay[?] :FILTer [:LPASs] [:LPASs] [:STATe][?] :HPASs [:STATe][?] :NREJect[?] [:STATe][?] :HOLDoff :TIME[?] :L

Summary of the content on the page No. 19

Quick Reference to TVS600 & TVS600A Commands :SOURce[?] :TYPE[?] TRIGger[:A]:PULSe Commands in the TRIGger:PULSe subsystem provide the capability to trigger when a pulse occurs that is outside specified parameters. TRIGger [:A] :PULSe :CLASs[?] :GLITch :POLarity[?] :QUALify[?] :WIDTh[?] :SOURce[?

Summary of the content on the page No. 20

Quick Reference to TVS600 & TVS600A Commands TRIGger[:A]:LOGic Commands in the TRIGger:LOGic subsystem provide the capability to trigger on logic patterns, clocked or not clocked. TVS600A models only. TRIGger [:A] :LOGic :CLASs[?] :CONDition[?] :FUNCtion[?] :PATTern :QUALify[?] :WIDTh[?] :STATe :SLOPe[?] :THReshold[n][?] TRIGger[:A]:SHOLdtime Commands in the TRIGger:SHOLdt


Alternative user manuals
# User manual Category Download
1 VXI TVS600A User manual Electronic Accessory 0
2 Sony In-car Cradle User manual Electronic Accessory 0
3 Sony 7011458 User manual Electronic Accessory 1
4 Agilent Technologies U1251A User manual Electronic Accessory 33
5 Samsung HL-P5067W User manual Electronic Accessory 0
6 Samsung HL-P6167W User manual Electronic Accessory 1
7 Samsung HL-P5667W User manual Electronic Accessory 0
8 Behringer EURODESK MX9000 User manual Electronic Accessory 57
9 Black Box AC504A User manual Electronic Accessory 1
10 Black Box ACX1000-MS User manual Electronic Accessory 3
11 Black Box ACX1000-CM User manual Electronic Accessory 0
12 Black Box ICS02A User manual Electronic Accessory 0
13 Black Box ACX1000 User manual Electronic Accessory 0
14 Black Box ACX1000-MM User manual Electronic Accessory 0
15 Black Box LBNC300A User manual Electronic Accessory 0