ページ1に含まれる内容の要旨
GCX Mounting Assembly Operation/Installation Manual
®
Dräger Model Cato with Siemens SC9000 Patient Monitor
Patient Monitor – Left Side on Pivot Arm / Power Supply on Right Side
The purpose of these instructions is to:
1. Describe attachment of GCX mounting hardware to the Cato anesthesia machine.
2. Describe attachment of the Siemens monitor to the GCX mounting hardware.
WARNING: USE OF MOUNTING HARDWARE AND MONITORING COMPONENTS OTHER THAN THOS
ページ2に含まれる内容の要旨
Install M Series Arm 1. The Slide has (2) ¼-20 x ¾” SHSS set screws that, when tightened, secure the Arm at the desired location within the Channel. 2. While supporting the bottom of the Arm with one hand, guide the Slide (at the rear of the arm) into the top of the Channel. 3. Once the Arm is at the desired height, tighten both set screws using the supplied 1/8” (3 mm) hex key. Slide Pick & Go Base into Mounting Plate 1) Align the mounting adapte
ページ3に含まれる内容の要旨
INSTRUCTIONS FOR USE – M Series Pivot Arm Pivot 1. To pivot the Arm (near the wall), simply push on the side of the Arm or mounted device. 2. To adjust the pivot tension at the Slide, use a ½” (or 13mm) socket driver to tighten or loosen the Pivot Tension Bolt to obtain the desired tension. Pivot Arm Swivel Tension Bolt Pivot Tension Bolt Swivel 1) To rotate the device simply push or pull the corners of the device. 2) To adjust the swivel ten
ページ4に含まれる内容の要旨
Cleaning 1. The mounting assembly may be cleaned with most mild, non-abrasive solutions commonly used in the hospital environment (e.g. diluted bleach, ammonia, or alcohol solutions). 2. The surface finish will be permanently damaged by strong chemicals and solvents such as acetone and trichloroethylene. 3. Steel wool or other abrasive material should never be used. 4. Damage caused by the use of unapproved substances or processes will not be warranted. To verify compatibility