Table of contents page 1. Outline..............................................................................................................................................................................................4 2. Basic Specification...........................................................................................................................................................................5 2.1 Format of Transmission...............................................
ページ4に含まれる内容の要旨
1. Outline 1) The following data is exchanged between the AU2700/5400 and the host computer system. (1) Test Requisition Information (T.R.I.) (2) Test Results 2) The data transmission is conducted in either of the following mode: (1) Class A Transmission is without synchronizing with analysis. This mode does not utilize the ACK/NAK protocol. (2) Class B Transmission is synchronizing with analysis. This mode does utilize the ACK/NAK protocol. 3) The Following format and protocol can be selected b
ページ5に含まれる内容の要旨
(2.1) Format of Transmission Item Contents Transmission RS-232C line Synchronization Synchronization Data transmission Half-Duplex mode Bit/sec 4800bps, 9600bps Configuration start bit 1 bit data bit 7 or 8 bit parity bit none, odd or even stop bit 1 or 2 bit total 9-12 bit These can be selected by [PARAMETER] [ONLINE] [PROTOCOL] menu. Form of Class A Transmission are conducted from the sender to Transmission the receiver in a certain time of interval. Class B
ページ6に含まれる内容の要旨
(2.3) Text format (1) Basic configuration as follows: 1) 2) 3) 4) 5) 6) Name Digit Contents Remarks 1) Text start code 1,2 01h-1Fh Data start code Normally 02h 2) Text 2 R# T.R.I. inquiry AU2700/5400 Classification RB Start of T.R.I. inquiry → Host R_ * T.R.I. inquiry of normal sample RH T.R.I. inquiry of rerun sample RE End of T.R.I. inquiry S# T.R.I. Host S_ * T.R.I. of normal sample → AU2700/5400 SH T.R.I. of rerun sample SE Sto
ページ7に含まれる内容の要旨
(2) Blocking A. Definition of Term 1) 2) 3) 4) 5) 6) Term Definition Text length Is the total number of bytes from 1) to 6). [PARAMETER] [ONLINE] [PROTOCOL] menu. Max block It is the Max length of text in one record. length [PARAMETER] The length of block can be selected in online parameter menu. [ONLINE] (256, 512, 1024) [PROTOCOL] menu. - 7 -
ページ8に含まれる内容の要旨
B. Kind of text and blocking Kind Kind of text Method of Blocking Fixed length • Start of T.R.I. inquiry Non-blocking Text • T.R.I. inquiry text • End of T.R.I. transmission text. • Stop of T.R.I. inquiry • Start of transfer result transmission text. • End of result transfer transmission text. Variable • T.R.I. text Using Max block length length text • Rerun T.R.I. text blocking < text length • Test result text Block first block • Rerun res
ページ9に含まれる内容の要旨
3. Transmission Text Format (3.1) Common and changeable item The following can be selected by [PARAMETER] [ONLINE] [PROTOCOL] menu. Text start code 1) Unit No. 3) Common Item Text end code 5) BCC 6) Sample Type Dummy(in Test Result transmission Text) Block Identification No. Rack No./Cup Position Unit No. on the T.R.I. text (with out Rerun) ANL unit No. and Cuvette side on Test Result transmission Text (without RB,ACAL,QC) Space Code before the Text Classification (S,R,D text) Reagent Lot No
ページ10に含まれる内容の要旨
(3) Repeat run inquiry test requisition. Cup 1) R H 3) Rack No. Posi- Sample no. Sample ID 5) 6) tion Sample type (4) End text to inquiry test requisition. 1) R E 3)5)6) - 10 -
ページ11に含まれる内容の要旨
(3.3) T.R.I. text (1) Normal sample test requisition information text. cup 1) S _ Rack No. posi- Sample no. Sample ID Dummy tion _ _ _ _ Sample type Block Identification No. s Patient Patient Patient e age month Information 1 Information 2 Information 3 x Patient Patient Patient Information 4 Information 5 Information 6 T
ページ12に含まれる内容の要旨
(3) End of T.R.I. Text transmission. 1) S E 5) 6) (3.4) Test result transmission text (1) Start of test result transmission text. 1) D B 3)5)6) - 12 -
ページ13に含まれる内容の要旨
(2) Normal sample test result text. Cup 1) D _ 3) Rack no. Posi- Sample no. Sample ID Dummy tion _ _ _ _ Sample type Block Identification No. S Patient Patient Patient e Age Month Information 1 Information 2 Information 3 x Patient Patient Patient Information 4 Information 5 Information 6
ページ14に含まれる内容の要旨
(3) Stat(fast) test result text (AU2700 only). Cup P n n n 1) d _ 3) Rack no. Posi- Sample no. Sample ID Dummy _ _ _ _ tion _ _ _ _ 1-22 Sample type Block Identification No. S Patient Patient Patient e Age Month Information 1 Information 2 Information 3 x Patient Patient Patient Information 4 Information 5 Inform
ページ15に含まれる内容の要旨
(4) Emergency(fast) test result text (AU2700 only). Cup E n n n 1) d _ 3) Rack no. Posi- Sample no. Sample ID Dummy _ _ _ _ tion _ _ _ _ 1-22 Sample type Block Identification No. S Patient Patient Patient e Age Month Information 1 Information 2 Information 3 x Patient Patient Patient Information 4 Information 5 Inf
ページ16に含まれる内容の要旨
(5) Repeat run result text -1 . Cup Original 1) D H 3) Rack no. Posi- Sample no. Sample ID Sample no. tion Sample type Block Identification No. Blocking Test R1 Bottle R2 Bottle code R1 Lot No. No. R2 Lot No. No. Cuvette side ANL unit No. Blocking Flags Test result 5)
ページ17に含まれる内容の要旨
(6) Repeat run result text -2 Stat fast sample (AU2700 Only) . Cup P n n n Original 1) d H 3) Rack no. Posi- Sample no. Sample ID Sample No. _ _ _ _ tion 1-22 Sample type Block Identification No. Blocking Test R1 Bottle R2 Bottle code R1 Lot No. No. R2 Lot No. No. Cuvette side ANL unit No.
ページ18に含まれる内容の要旨
(7) Reagent blank result text. Dummy 1) D R 3) _ _ _ _ _ _ Sample no. _ _ _ _ _ _ _ _ _ Sample type This dummy is same ID digits Block Identification No. and assigned area is decided associated with the rack
ページ19に含まれる内容の要旨
(8) Acal (Calibration) result text. Dum 1) D A 3) _ _ _ _ _ _ Sample no. Calibrator ID my _ _ This dummy spaces is assigned in case rack No. exists Calibrator No. (1-80) and assigned area is decided associated with the rack No. Block Identification No. ANL unit No. Cuvette side Blocking Test R1 Bottle R2 Bottle Code R1 Lot No. No. R2 Lot No. No.
ページ20に含まれる内容の要旨
(9) QC (Quality Control) result text. Dum 1) D Q 3) _ _ _ _ _ _ Sample no. Control ID my _ _ This dummy No. is assigned in case rack No. exists Control No. and assigned area is decided associated with the rack No. Block Identification No.