Series FR566G Deluxe Clinical Care Recliner
Series FR566DG Deluxe Clinical Care Recliner with Drop Arm
Series FR566GH Deluxe Clinical Care Recliner
with Heat and Massage
Series FR566DGH Deluxe Clinical Care Recliner with Drop Arm
with Heat and Massage
Assembly and Operating Instructions
IMPORTANT: READ THIS MANUAL BEFORE OPERATING YOUR LUMEX
FR566G / FR566DG / FR566GH / FR566DGH DELUXE CLINICAL CARE RECLINER.
SAVE THIS MANUAL FOR FUTURE USE.
THE MOST CURRENT VERSION OF THIS MANUAL CAN BE FOUN
ページ2に含まれる内容の要旨
CONTENTS INTRODUCTION. ....................................................................................................................................................... 3 INTENDED.USE. ................................................................................................................................................. 3 IMPORT ANT.SAFETY .PRECAUTIONS—PLEASE.READ.BEFORE.USE. ....................................................... 3 W ARNINGS. .....................................
ページ3に含まれる内容の要旨
INTRODUCTION This manual contains assembly, operation, and maintenance instructions for Lumex Clinical Care Recliners Series FR566G, FR566DG, FR566GH and FR566DGH. Read the entire manual carefully before using your recliner, and refer to it during use if you have questions. If you have further questions, please call Graham-Field / Lumex Technical Support at 770-447-1609. INTENDED USE Lumex Clinical Care Recliners are intended to make the care, treatment and recovery of patients easier and mo
ページ4に含まれる内容の要旨
. W ARNING:.This.product.has.a.maximum.weight.capacity.of.350.lb.(158.kg),.EVENL Y . DISTRIBUTED. . W ARNING:.Casters.must.be.locked.at.all.times,.except.during.transport. . W ARNING:.Unauthorized.modification.of.your.recliner.or.the.use.of.non-Lumex. replacement.parts.may.change.the.structure.of.the.recliner.and.could.create.a. hazardous.condition,.which.may.result.in.serious.injury.and.will.void.the.warranty . . W ARNING:.Notice.for.California.Customers-.California.Proposition.65.W ARNING: Thi
ページ5に含まれる内容の要旨
ASSEMBLY UNPACKING 1. Check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field authorized distributor. 2. Remove all packaging materials from recliner base. PRODUCT DESCRIPTION Features, FR566G and FR566GH upholstered back upholstered seat padded push handle upholstered armrest upholstered legrest / skirt upholstered footrest total locking caster fold down side table (total of 3 on recliner; black locking tab) frame rear direc
ページ6に含まれる内容の要旨
Features, FR566DG and FR566DGH upholstered back upholstered seat upholstered armrest padded push handle gas spring upholstered actuating lever legrest / skirt upholstered footrest fold down side table total locking caster frame (total of 3 on recliner; black locking tab) rear directional (swivel) locking caster removable side panel (right rear leg only; green locking tab) Clinical Care Recliner Series FR566DGH FR566DG-INS-LAB-RevC13: Deluxe Clinical Care Recliner Assembly and Operating Instruct
ページ7に含まれる内容の要旨
Upholstered Back Assembly, All Models upholstered back upholstered skirt threaded knobs upholstered back’s kd brackets steel tube mechanism’s bayonet links bracket and link to be even threaded knob through large hole in bayonet link kd bracket bayonet link 1. Lift the skirt on the rear of the upholstered back. To make assembly easier, hold skirt up with a small piece of tape to fully expose the kd (knock-down) brackets that are to be mated to the two bayonet lugs on the recliner mechanism. 2
ページ8に含まれる内容の要旨
OPERATION . W ARNING:.Disconnection.of.the.device:.In.case.of.emergency ,.always.disconnect. the.device.by.removing.the.power.cord.plug.from.the.power.outlet. TENTE® CASTERS, ALL MODELS . W ARNING:.Casters.must.be.locked.at.all.times,.except.during.transport. black actuator tab green actuator tab All Lumex Clinical Care Recliners now feature the latest design, all-plastic casters from TENTE. Specifically designed for use in healthcare environments, the new casters are easier to clean, easier
ページ9に含まれる内容の要旨
SIDE TABLE OPERATION, ALL MODELS To Raise Table side table (disengaged position) Raise the side table all the armrest way up and slide in toward the armrest until tabs engage 3 /8" clevis pin slots in mount bracket. Press attachment side table down lightly on table to ensure (engaged position) locking tabs are engaged. side table mount bracket To Lower Table tabs engage into Lift the side table up slightly slots on side table until tabs disengage slots and mount bracket pull table away fr
ページ10に含まれる内容の要旨
SIDE PANEL OPERATION For easy cleaning and access to the side and underside of the recliner, Lumex Clinical Care Recliners feature easy-to-remove side panels. FR566G and FR566GH To Remove Side Panel: depress panel tabs the two push buttons located side panel on the lower sides of frame. Pull side panel gently away and down from frame. There are tabs on the frame underneath the armrest that hold the side panel in place. To Install Side Panel: insert side panel into recliner frame by loca
ページ11に含まれる内容の要旨
RECLINER OPERATION, ALL MODELS The Lumex Series FR566G / FR566DG / FR566GH / FR566DGH recliners feature three positions for patient comfort. The fourth position is a Trendelenburg (or Shock Position) that must be activated by a caregiver. The Trendelenburg Position CANNOT be activated by a patient while in the recliner. In addition, this recliner features a back that can be repositioned via the Adjustment Lever located on the Push Handle on the back of the recliner. First Position The first
ページ12に含まれる内容の要旨
Upright Position Third Position The third position is the Second Recline Position and is also known as the “Full Recline” Upright Position position. This position is ideal for patient treatment, rest and reading. First Recline Position (TV Position) To Achieve Third Position / Second Recline Position: • Reclining by the patient: from the First Recline Position continue to push against the back by placing your hands on the armrests and continue to recline yourself to the Second Recline Positi
ページ13に含まれる内容の要旨
DROP ARM FUNCTION, FR566DG AND FR566DGH ONLY The Lumex Deluxe Clinical Care Recliner with Drop Arm (Series FR566DG and FR566DGH) features self-storing drop arms that are simple for the caregiver to actuate. By lowering the arm, a safer and easier lateral transfer for the resident and caregiver can be achieved. . W ARNING:. Always.lock.all.recliner.casters.before.attempting.to.transfer.a.patient.. DO NOT.attempt.to.transfer.a.patient.if.you.are.unsure.about.the.physical.abilities. of.the.patien
ページ14に含まれる内容の要旨
LUMEX CLINICAL CARE RECLINERS WITH HEAT AND MASSAGE FEATURE INSTALLATION AND OPERATING GUIDE Series FR566GH and FR566DGH 1 3 Tools Required: /4” Hex Wrench, /8” Hex Wrench, Phillips Screwdriver, masking or packaging tape . W ARNING:.Important!.Read.and.understand.these.instructions.before.assembling. or using.the.Specialty.Healthcare.Seating.Heat.and.Massage.Unit..If.you.do. not.understand.any.part.of.these.warnings,.cautions.or.instructions,.contact.a. healthcare.professional.for.direction.in.
ページ15に含まれる内容の要旨
Connection The Heat and Massage Unit comes pre-installed in Lumex Clinical Care Recliners with the Heat and Massage Option (FR566GH and FR566DGH). During recliner shipment, the Heat and Selector Switch connections are disconnected. Several easy connections are all that is required to make the unit functional. Main Components 1. The selector switch (shown at right), in the switch box mounted on the right side of the recliner underneath the armrest, is a single multi-position switch that
ページ16に含まれる内容の要旨
Heat and Massage Unit Setup Instructions 1. Place the back cushion over the vertical bayonet recliner mechanism lugs of the recliner mechanism. Secure back on each side with the threaded screw knobs enclosed knob with the Recliner Operating Instructions, as shown at right. 2. Raise the back skirt on the lower rear of the back cushion and use masking or packaging tape Knob Secure Back Cushion to to temporarily secure it up and out of the way. Recliner Mechanism This will permit uninterrup
ページ17に含まれる内容の要旨
Heat and Massage Unit Operation Instructions 1. Plug the power supply cord into any grounded outlet. 2. The Heat and Massage Unit is operated using the single multi-position selector switch located on the right side of the recliner underneath the armrest. The Heat and Massage functions can be operated independently. The selector switch faceplate is shown at right. 3. Press the selector switch UP to activate the HEAT FUNCTION. 4. Press the selector switch DOWN to activate the MASSAGE FUNCT
ページ18に含まれる内容の要旨
Heat and Massage Unit Troubleshooting Guide The cables and connections need to be validated. This will require the services of a maintenance person and a clear working area. 1. INSPECTION: Perform the following checks. If a working recliner of the same model is available, use it for comparison. 2. Is the AC cord properly connected to the power supply? Is the cable from the power supply to the control module junction box in place? 3. With the power supply AC cord plugged into a live wall outle
ページ19に含まれる内容の要旨
CARE AND MAINTENANCE . W ARNING:.Inspect.all.components.periodically.and.replace.as.needed. EVERY THREE MONTHS • Check and clean casters. • Tighten bolts in each leg that secure casters into frame. • Check brakes on casters. • Check and clean lock recline mechanism. • Check recliner mechanism for operation in all positions, including Trendelenburg Position, with assistant seated in recliner. UPHOLSTERY MAINTENANCE AND STAIN REMOVAL • Ordinary dirt: wash with warm water and a mild soap or deterg
ページ20に含まれる内容の要旨
The information published in this care guide refers to the performance of PERMABLOK3® products in specific tests conducted under laboratory conditions. Results may vary under actual conditions. This information is not a guarantee and does not relieve the user from the responsibility of the proper and safe use of the product and all cleaning agents. The use of certain agents can be harmful to the surface appearance and lifespan of vinyl. Spradling, its agents, and assigns assume no responsib