BUFFALO FILTER
Smoke Evacuation Systems
Operator’s Manual
®
DELL PlumeSafe
Smoke Evacuation System
ページ2に含まれる内容の要旨
Limited Warranty For a period of one (1) year following the date of delivery, BUFFALO FILTER warrants the Dell PlumeSafe® Evacuation System against any defects in material or workmanship. For a period of one (1) year following the date of delivery. BUFFALO FILTER will repair or replace (at BUFFALO FILTER’S option) the same without charge, provided that routine maintenance as specified in this manual has been performed using replacement parts approved by BUFFALO FILTER. This wa
ページ3に含まれる内容の要旨
Table of Contents & List of Illustrations Section Title Page 1.0 SYSTEM DESCRIPTION 1.1 1.1 Introduction 1.1 1.2 Inspection 1.1 1.3 Operational Information 1.2 1.4 Cautions and Warnings 1.2 2.0 OPERATING INSTRUCTIONS 2.1 2.1 System Controls 2.1 2.2 2.2 ViroSafe Filter Instructions 2.3 Set-up and Operation 2.3 2.4 Specifications 2.4 3.0 MAINTENANCE 3.1 3.1 General Maintenance Information 3.1 3.2 Cleaning 3.1 3.3 Periodic Inspection 3.1 3.4 Troublesh
ページ4に含まれる内容の要旨
System Description Section 1.0 1.1 Introduction The Dell PlumeSafe® Smoke Evacuation Systems is designed to remove smoke and noxious odors produced by surgical ablation during laser procedures. The Dell PlumeSafe Smoke Evacuation Systems have been designed with a precision centrifugal action pump. The ultra-quiet motor is used to draw the surgical smoke from the surgical site through the vacuum tubing and into the ViroSafe filter where the surgical smoke is processed by a
ページ5に含まれる内容の要旨
1.1 ® ® BUFFALO FILTER Dell PlumeSafe Smoke Evacuation System
ページ6に含まれる内容の要旨
1.3 Operational Information The operational information contained in this section is intended for the customer review of regulatory issues. The information pertains to the use of the products both domestically and internationally: 1. The BUFFALO FILTER Dell PlumeSafe Smoke Evacuation System(s) complies with IEC601.1 electrical specifications in the following systems: 100-120V 50/60Hz, 2. The Dell PlumeSafe Smoke Evacuation System(s) has a classification under the UL
ページ7に含まれる内容の要旨
WARNINGS: The BUFFALO FILTER ViroSafe Filters are completely disposable. Please handle these disposable filters according to your local codes or regulations and hospital policy. These filters may be disposed of or incinerated, whichever is appropriate for your institution. • The ViroSafe Filter should be changed according to the prescribed life of the filter. The ViroSafe Filter, used with the Dell PlumeSafe Smoke Evacuation System(s), should not be used for more than the time
ページ8に含まれる内容の要旨
There are no user serviceable components in the Dell PlumeSafe Smoke Evacuation System(s). Refer service to qualified service personnel. Use only with the power cord provided and always plug into a grounded outlet. CAUTIONS: • Do not block either the tubing or the filter. If either becomes occluded or significantly restricted, the motor/blower may overheat and cause the unit to fail. • Using any other filter or accessory not supplied by BUFFALO FILTER may cause damage
ページ9に含まれる内容の要旨
Operating Instructions Section 2.0 2.1 System Controls The electronic system controls on the Dell PlumeSafe Smoke Evacuation System(s) are designed to be easy to understand and simple to use. The membrane control panel contains the motor on/off switch, suction power adjustment, footswitch enabled, filter life reset button, filter life indicator, TURBO and normal modes (refer to page 2.5). Note: Please be sure to read all instructions before installing accessories or operating
ページ10に含まれる内容の要旨
• FILTER LIFE INDICATOR (Membrane Control Panel) The filter life indicator on the membrane control panel provides a visual indication of the status of the life of the filter in use. The filter life indicator for the Dell PlumeSafe Smoke Evacuation System(s) ViroSafe filter is one hour. To reset the filter life: After each filter change, press and hold the RESET button for five seconds to reset the timer for the next filter. Reading the Filter Life Indicator Install a ViroSa
ページ11に含まれる内容の要旨
Filter Removal Instructions: 1. After the ViroSafe Filter has been exhausted and requires changing, turn the smoke evacuation system off and disconnect any accessory tubing attached to the filter. 2. Pull the ViroSafe Filter from the smoke evacuation system and dispose of in accordance with facility biohazard practice and policy. The ViroSafe Filter may be disposed of or incinerated. 3. Clean the unit with appropriate germicide prior to re-use. Follow the indicated instructions
ページ12に含まれる内容の要旨
2.4 Specifications Dell PlumeSafe® Whisper BUFFALO FILTER P/N PSWDELL CONFIGURATION Portable with handles 3-Stage filtration in one casing (pre-filter, Dell ViroSafe Filters 3/8” port Carbon, ULPA ) Filter(s) ULPA Grade 0.12 with > 99.9995% efficiency Particle size, µm Filter life 1 hour Filter life indicator Timed replacement ViroSafe Filter P/N Dell Filter (2Pk) MAX. FLOW SETTING 70 Lpm Variable flow control Yes Motor, hp (kW) 0.33 (0.25) HOSES/TUBING 3/8”
ページ13に含まれる内容の要旨
Dell PlumeSafe Smoke Evacuation System 1. Motor Suction ON Indicator 2. Liter Per Minute Flow Rate Indicator 3. TURBO Suction Mode 4. Normal Suction Mode 5. Proximal evacuation hose connection to filter 6. Filter Receptacle with filter 7. Micro Flow Adjustment 8. Filter Life Indicator 9. Footswitch Enabled LED 10. Motor Suction On/Off 11. Filter Life Reset 12. Footswitch Jack Main Power Switch – I/O Circuit Breakers Dell60UF ULPA One-hour filters (2) 3/8” x 10’ no
ページ14に含まれる内容の要旨
Maintenance Section 3.0 3.1 General Maintenance Information This section contains information for ordinary upkeep of the Dell PlumeSafe Smoke Evacuation System(s). While the system has been designed and manufactured to high industry standards, it is recommended that periodic inspection and performance testing be performed by a qualified Biomedical Technician to ensure continued safe and effective operation. 3.2 Cleaning The exterior of the system may be cleaned by wiping
ページ15に含まれる内容の要旨
Customer Service Section 4.0 4.1 Equipment Return For the quickest response to your service needs, please follow these procedures: Step 1: Write down model and the serial number of the Dell PlumeSafe Smoke Evacuation System. Step 2: Call Customer Service at the toll free or local number listed and describe the problem. Step 3: If the problem cannot be resolved over the phone and the equipment must be returned for repair, you must obtain a “Return Goods Authorization”
ページ16に含まれる内容の要旨
Terms & Warranty Section 5.0 SPECIFICATIONS: to be defective and the installation or use thereof. Specifications are subject to change without notice. No warranty is included for incidental or consequential damages of any nature arising from SHIPMENT OF ORDER: any defect. The warranty above is the only warranty Buffalo Filter will try to accommodate individual made by Buffalo Filter and is expressly in lieu of all customer requests for shipping method. Buffalo Filter other w
ページ17に含まれる内容の要旨
2004 BUFFALO FILTER 595 Commerce Drive Buffalo, New York USA 14228 (716) 835-7000 or (716) 835-3414 fax Division of Medtek Devices, Inc. www.buffalofilter.com Customer Service: 800-343-2324 (U.S. Only) – 716-835-7000