Table of contents 1.0 Introduction 1.1 Safety Information 1.2 Unpacking 2.0 Operation Procedure 2.1 Unit Drawing 2.2 Description of Buttons and Switches 2.3 Read Mode Procedure 2.4 Source Mode Procedure 3.0 Operator Maintenance 4.0 Specifications 5.0 Calibration Procedure
ページ3に含まれる内容の要旨
1.0 Introduction ® The OMEGA CL3512A calibrator/thermometer is two meters in one. The CL3512A simulates type J/K/T/E thermocouple signals. Each signal is adjustable by using the coarse and fine dials. The CL3512A can also be used as a dual type J/K/T/E thermocouple input thermometer. Features include a large 3½ digit display with backlighting and display selections of HOLD, °C/°F, and0.1/1°. The source mode of the CL3512A simulates the thermocouple output to check the operation of a th
ページ4に含まれる内容の要旨
2.0 Operation Procedure 2.1 Drawing of Unit Actual Dimensions: 195mm(H) x 92mm(W) x 53mm(D) 3 2.2 Descriptions of Buttons and Switches / Buttons Temperature is displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F). When the thermometer is turned on, it is set to the temperature scale that was in use when the thermometer was last turned off. To change the temperature scale, press the °C or °F key. Button Press “ ” button to toggle the on and off backlight.
ページ5に含まれる内容の要旨
/ / Input Selections The input selection indicates which input is selected; T1 thermocouple, T2 thermocouple or the difference between the two thermocouples (T1-T2). When the thermometer is turned on, it is set to the temperature input that was in use when the thermometer was last turned off. Selector Switch The circular selector switch is used to turn the unit off () or to select Reading Mode (Input) or Simulation Mode (Output). There are nine selector switch positions under output
ページ6に含まれる内容の要旨
2.4 Source Mode Procedure (Calibration Simulation) 1. Turn CL3512A on by turning selector switch to any one of the output selector positions. 2. Press T1 button. 3. Select thermocouple type (J, K, T or E) using type button. Select °C or °F using °C or °F buttons. 4. Plug proper type of thermocouple wire into either one of the two output connectors. (SMP male connector required) 5. Attach the other end of the thermocouple wire to the instrument being tested/calibrated. 6. Move the out
ページ7に含まれる内容の要旨
3.0 Operator Maintenance WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the thermometer before removing the cover. Battery Replacement Power is supplied by a 9 volt “transistor” battery. The “ ” appears on the LCD display when replacement is needed. To replace the battery, remove the two screws from the back of the meter and lift off the battery cover. Remove the battery from battery contacts. 9 4.0 Specifications ELECTRICAL Temperature Sca
ページ8に含まれる内容の要旨
GENERAL Display: 3½ digit liquid crystal display (LCD) with a maximum reading of1999. Polarity: Automatic, positive implied, negative polarity indication. Overrange: -OL is displayed. Zero: Automatic. ” is displayed when the battery voltage Low battery indication: The “ drops below the operating level. Measurement rate: 1 times/second. Accuracy: Stated accuracy at 23°C±5°C, <75% R.H. Dimensions: 195mm(H) x 92mm(W) x 53mm(D). Weight: approx. 9 oz. (250g) including battery. Input Conn
ページ9に含まれる内容の要旨
13
ページ10に含まれる内容の要旨
WARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA's customers receive maximum coverage on each product. If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA's Customer Servic