ページ1に含まれる内容の要旨
ふ
SK-1000WP
SK-2000WP
SK-5000WP
SK-5001WP
SK-10KWP
SK-20KWP
SK-30KWP
1WM PD4000461B
- 0 -
ページ2に含まれる内容の要旨
SAFETY PRECAUTIONS All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Institute: Product Safety Signs and Labels). The meanings are as follows: A potentially hazardous situation which, if not WARNING avoided, could result in death or serious injury. A potentially hazardous situation which, if not CAUTION avoided, may result in minor or moderate injury. This manual is subject to change without notice at any time to
ページ3に含まれる内容の要旨
PARTS DESCRIPTION Weighing Platform Spirit level ON/OFF key Turns the scale power on/off. UNITS key (only for (k)g- lb model) Changes weighing units, (k)g or lb. RE-ZERO key Display Sets display to zero or subtracts weight of a container. Four leveling feet Battery cover Batteries are not included. Battery cover lock Battery compartment AC adapter socket Battery cover lock DISPLAY STABLE indicator Weighing Indicates w
ページ4に含まれる内容の要旨
PREPARATION Installing Batteries / Connecting the AC Adapter Installing Batteries CAUTION When replacing batteries, be careful not to apply too much force to the weighing platform. An excessive force could damage the scale. Unlock and remove the battery cover. (See illustration at the right) Insert six new batteries (R20P / LR20 / D size) into the battery compartment, taking extreme care to that the plus and minus polarities are observed. Be sure to install and lock the battery
ページ5に含まれる内容の要旨
WEIGHING Before Weighing Automatic Power-Off Function. If the scale is left ON and the stable indicator is displayed, the automatic power-off function turns the power off after approximately five minutes (factory setting). To disable or enable the function, press the [ON/OFF] key while pressing the [RE-ZERO] key. The scale stores this setting. To verify this, press both the [RE-ZERO] and [UNITS] keys while in the weighing mode. (See "CALIBRATION" about the [UNITS] key.) poff0 Func
ページ6に含まれる内容の要旨
CALIBRATION When to Calibrate Calibration may be required when the scale is initially installed, if it is moved a substantial distance, or in accordance with local regulations. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur. The Position of the [UNITS] key for the Calibration The (k)g model has no [UNITS] key, but the key for the calibration is located under
ページ7に含まれる内容の要旨
Calibration by Gravity Compensation If you have calibration weights at the capacity of the scale, you can calibrate the scale with the weights instead of compensating for the gravity acceleration. 2 The scale is calibrated at 9.798 m/sec at the factory. If the gravity acceleration at your location is not the same, calibrate the scale by compensating for the gravity acceleration. (See “The Value of Gravity at Various Locations” and “World Map” and find your location value.) 1. E
ページ8に含まれる内容の要旨
SPECIFICATIONS Model SK-1000WP SK-2000WP SK-5000WP SK-5001WP SK-10KWP SK-20KWP SK-30KWP 1000 g 2000 g 5000 g 5000 g 10 kg 20 kg 30 kg Capacity 2.2 lb 4.4 lb 11 lb --- 22 lb 44 lb 66 lb 0.5 g 1 g 2 g 1 g 0.005 kg 0.01 kg 0.02 kg Resolution 0.001 lb 0.002 lb 0.005 lb --- 0.01 lb 0.02 lb 0.05 lb Non-linearity ±1g ±2g ±4g ±4g ±0.01 kg ±0.02 kg ±0.04 kg Repeatability 0.5g 1g 2g 2g 0.005 kg 0.01 kg 0.02 kg Span drift ±0.015% / °C TYP (5°C~35°C / 41°F~95°F) Operating -10°C~40°C / 14°F~104°F, Les
ページ9に含まれる内容の要旨
World Map - 8 -
ページ10に含まれる内容の要旨
3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013 JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148 A&D ENGINEERING, INC. 1756 Automation Parkway, San Jose, California 95131 U.S.A. Telephone: [1] (408) 263-5333 Fax: [1] (408)263-0119 A&D INSTRUMENTS LIMITED Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D INSTRUMENTS LIMITED Große Straße