ページ1に含まれる内容の要旨
LF654 /LF620
Field Intelligent Device – Mount Anywhere Series - Flanged
LF654 /LF622
Electromagnetic Flowmeter
1/2" to 18" (15 to 450 mm)
Introduction
The electromagnetic flowmeter uses Faraday’s Law of
electromagnetic induction to measure the process flow.
The device consists of two units: a detector, through
which the fluid to be measured flows and in which
low-level signals proportional to flow rates are
obtained; and a converter, which supplies excitation
current to the
ページ2に含まれる内容の要旨
TIC-LF654L Drinking water approvals: Fluid conductivity: PU and HR liner: NSF certified 3µS/cm minimum Fluid temperature: Model LF654 Detector -4 to 248 °F (-20 to +120 °C): Mounting style: Flange connection type PFA lined flowmeter - Fluid pressure: -4 to 140 °F ( 20 to +60 °C): - - - 15 psi or 1.0 bar ( 0.1 MPa) to the nominal Polyurethane lined flowmeter - pressure of the connection flange. 23 to 176 °F ( 5to +80 °C): Note: Before shipping from the man
ページ3に含まれる内容の要旨
TIC-LF654L Pulse width: 0.5 to 500ms (but less than half of Model LF620 and LF622 converters the period for 100% flow rate) Input signals Note: The same and simultaneous pulse is not available Analog signal — the voltage signal from detector, between DO1 and DO2.) proportional to process flow rate (for LF622 • Multi-range selection outputs (Note 1) separate type converter). • High, High high, Low, and/or Low low alarm Digital input DI outputs (Note 2) Signal type: 20 to 30Vdc
ページ4に含まれる内容の要旨
TIC-LF654L LCD display: Surge protection: Arresters are installed in the power supply and a Full dot-matrix 128×128 dot LCD display current signal output circuit to help protect the meter (back–light provided) from lightning and improve personnel safety. A parameter change will rotate the display. Case: Aluminum alloy (equal to IP 67) Parameter settings — Parameters can be set as follows: Coating: • IR Switches: Three key switches are provided Acrylic resin-baked coating, pearl–
ページ5に含まれる内容の要旨
TIC-LF654L Installation Dimensions Unit: inch (mm) Note1: Eye bolts are provided at the top for flowmeters sized 8" (200mm) or above, and further, a roll-prevention base is provided for flowmeters sized 10" (250mm) or larger. Note2: Cable glands are not provided for LF654 of cFMus approved type. Refer to the cable connection port at converter. ASME B 16.5 class 150 dimensions: JIS B 2220 10K dimensions: Meter size L1 L2 L3 No. of Weight Meter size L1 L2 L3 No. of Weight (inch) (
ページ6に含まれる内容の要旨
TIC-LF654L Note1: Eye bolts are provided at the top for flowmeters sized 8" (200mm) or above, and further, a roll-prevention base is provided for flowmeters sized 10" (250mm) or larger. Note2: Cable glands are not provided for LF654 of cFMus approved type. Refer to the cable connection port at converter. ASME B 16.5 class 150 dimensions: JIS B 2220 10K dimensions: Unit: inch (mm) Meter size L1 L2 L3 No. of Weight Meter size L1 L2 L3 No. of Weight (inch) (inch) (inch) (inch) bolts (l
ページ7に含まれる内容の要旨
TIC-LF654L Note1: Eye bolts are provided at the top for flowmeters sized 8”(200mm) or above, and further, a roll-prevention base is provided for flowmeters 10”(250mm) or larger. Note2: Cable glands are not provided for LF654 of cFMus approved type. Refer to the cable connection d JIS B 2220 10K dimensions: ASME B 16.5 class 150 dimensions: Meter size L1 L2 L3 No. of Weight Meter size L1 (L2) L3 No. of Weight (inch) (inch) (in
ページ8に含まれる内容の要旨
TIC-LF654L Note1: Eye bolts are provided at the top for flowmeters sized 8" (200mm) or above, and further, a roll-prevention base is provided for flowmeters sized ASME B 16.5 class 150 dimensions: JIS B 2220 10K dimensions: Meter size L1 L2 L3 No. of Weight Meter size L1 (L2) L3 No. of Weight (inch) (inch) (inch) (inch) bolts (lbs) (mm) (mm) (mm) (mm) bolts (kg) 1/2 7.90 6.77 8.54 4 approx. 9.0 15 200 170 218 4 approx. 4.5 7.90 1 7.01
ページ9に含まれる内容の要旨
TIC-LF654L Unit: inch (mm) Weight: Approx. 7 lb (3.5 kg) Note: Cable glands are not provided for LF622F cFMus approved type. Refer to the part Cable connection port at detector. Note: 1 inch = 25.4 mm Figure 7. Separate type converter LF622 and LF622F External Connections ● Combined type LF650/LF620 flowmeter and LF654/LF620F flowmeters Instrument panel : Ordered separately Grounding with 100
ページ10に含まれる内容の要旨
TIC-LF654L ● Separate type LF650/LF622 flowmeter and LF654/LF622F flowmeter Instrument panel : Ordered separately Grounding with 100 or less 2 wire 5.5mm or more ground resistance Power switch Grounding with 100 or less (External double-pole power switch) ground resistance Power supply Thick walled steel conduit Current output (4 20mAdc) Signal cable or PROFIBUS (2-wire shielded hard-rubber sheathed cable) Digital input (20 30Vdc) Excitation cable or Modbus (3-wire shield
ページ11に含まれる内容の要旨
TIC-LF654L (6) The electromagnetic flowmeter is not equipped with Wiring Precautions terminating resistors. Use the terminating resistor (1) Explosion proof type flowmeters are not provided unit for PROFIBUS-PA or junction box, if with cable glands. necessary. Refer to the part Cable connection port at detector (7) Only one PROFIBUS-PA cable goes through a and converter. cable gland of the Electromagnetic Flowmeter. (2) Connect the grounding wire (IV wire 5.5mm² or Please use t
ページ12に含まれる内容の要旨
TIC-LF654L Table 3. Flow Rate and Flow velocity Ordering Information 1. When ordering the LF654 series flowmeters, refer (SI unit) 3 to Tables 4 and 5 (Type Specification Codes). Unit: m /h Flow rate An entry must be made for each of the columns in Size 3.0 12 each of these tables. 0.1 m/s 0.3 m/s 1.0 m/s (mm) m/s m/s 2. Fluid characteristics: 15 0.06362 0.1908 0.6361 1.908 7.634 (1) Type of fluid to be measured and its characteristics 25 0.1767 0.5301 1.767 5.301 21.21
ページ13に含まれる内容の要旨
TIC-LF654L Table 4. Specification Code (Flange type detector LF654 Series) Model Specification Code Detector category Description 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Gr.-A Gr.-B L F 6 5 Flange type electromagnetic flowmeter detector Usage 4 cFMus Class 1 Division 2 Hazardous Location √ √ Meter size D ½"(15mm) √ E 1"(25mm) √ S 1¼"(32mm) √ F 1½"(40mm) √ G 2"(50mm) √ W 2½"(65mm)
ページ14に含まれる内容の要旨
TIC-LF654L Table 5. Specification Code for converters Model Specification Code Contents LF620 LF622 1 2 3 4 5 6 7 8 9 1011 12 13 14 type type L F 6 2 Electromagnetic flowmeter converter 0 Combined (Integral) type – 2 Separate (Remote) type – Purpose F cFMus class I, Division 2 approved Shape A Standard type with case Converter mounting fitting A None – C Panel, Acce