Instruction Sheet
PN 51A-499AOZ/rev.H
September 2010
Model 499A OZ
Ozone Sensor
For additional information, please visit our website
at www.emersonprocess.com/raihome/liquid/.
CAUTION CAUTION
Do not exceed pressure and temperature specifications.
SENSOR/PROCESS APPLICATION COMPATIBILITY
Pressure: 65 psig max (549 kPa abs max)
The wetted sensor materials may not be compatible
Temperature: 32 to 122°F (0 to 50°C)
with process com position and operating conditions.
Application compat ibility is en
ページ2に含まれる内容の要旨
MODEL 499A OZ INSTALLATION INSTALLATION Install the sensor in a flowing sample. Keep the sample flow as constant as possible at a value within the limits below. Flow through 1 to 5 gpm (3.8 to 19 L/min) Open channel 1 ft/sec (0.3 m/sec) Low flow cell 2 to 5 gph (7.6 to 19 L/hr) The Model 499AOZ sensor is NOT recommended for submersion into tanks, basins, or ponds. FIGURE 1. Sensor Orientation FIGURE 2. Flow-Through 2 in. Tee FIGURE 3. Flow-Through 1-1/2 in. Tee FIGURE 4. Low Flow Cell (PN 2409
ページ3に含まれる内容の要旨
MODEL 499A OZ WIRING WIRING FIGURE 6. Wiring Sensor with OptimumEMI/RFI or FIGURE 5. Wiring Sensor with Standard Cable to Variopol Cable to Model 54eA Analyzer Model 54eA Analyzer FIGURE 7. Wiring Sensor with Standard Cable to FIGURE 8. Wiring Sensor with Optimum EMI/RFI or Model 1056 and Model 56 Analyzer Variopol Cable to Model 1056 and Model 56 Analyzer 3
ページ4に含まれる内容の要旨
MODEL 499A OZ WIRING FIGURE 9. Wiring Sensor with Standard Cable to FIGURE 10. Wiring Sensor with Optimum EMI/RFI or 5081-A Transmitter. Variopol Cable to 5081-A Transmitter. FIGURE 11. Wiring Sensor with Standard Cable to FIGURE 12. Wiring Sensor with Optimum EMI/RFI or Xmt-A Transmitter. Variopol Cable to Xmt-A Transmitter. 4
ページ5に含まれる内容の要旨
MODEL 499A OZ WIRING FIGURE 13. Wiring Sensor with Standard Cable to FIGURE 14. Wiring Sensor with Optimum EMI/RFI Model 1066 Transmitter. or Variopol Cable to Model 1066 Transmitter. When making connections through a junction box (PN 23550-00), wire point-to-point. NOTE: Use a wire nut and pigtail (included) when connecting several wires to the same terminal. FIGURE 15. Pin Out Diagram for Model 499AOZ-VP Sensor (top view of connector end of sensor) CALIBRATION Full Scale: Because stable dilu
ページ6に含まれる内容の要旨
MODEL 499A OZ MAINTENANCE MAINTENANCE Periodic maintenance and cleaning is required for best performance of the sensor. Generally, the membrane and fill solution should be replaced every four to six months. Sensors installed in harsh or dirty environments require more frequent maintenance. The optimum maintenance frequency is best determined by experience. 7. Hold the sensor at about a 45-degree angle with the cathode end pointing up. Add electrolyte solution WARNING through the fill hole until
ページ7に含まれる内容の要旨
MODEL 499A OZ MAINTENANCE FIGURE 16. Sensor Parts SPARE PARTS 33523-02 Electrolyte Fill Plug 9550094 O-Ring, Viton 2-014 33521-02 Membrane Retainer 23501-11 Ozone Membrane Assembly: includes one membrane assembly and one O-ring 23502-11 Ozone Membrane Kit: includes 3 membrane assemblies and 3 O-rings 9210299 #3 Ozone Sensor Fill Solution, 4 oz (120 mL) 7
ページ8に含まれる内容の要旨
MODEL 499A OZ ORDERING INFORMATION The right people, the right answers, ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE right now. http://www.raihome.com Specifications subject to change without notice. Credit Cards for U.S. Purchases Only. Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 http://www.raihome.com © Rosemount Analytical Inc. 2010 8