Operating Instructions
VACUUM PYCNOMETER
Model: 45-9300 (AP-132)
Distributor:
ELE International, Inc. ELE International, Ltd.
PO Box 608 Chartmoor Road Chartwell Business Park
Loveland, CO 80539 Leighton Buzzard, Bedfordshire
USA LU7 4WG England
Phone: (970) 663-9780 Phone: + (0) 1525 249200
Fax: (970) 663-9781 Fax: + (0) 1525 249249
E-Mail: soiltest@eleusa.com E-Mail: ele@eleint.co.uk
Website: www.eleusa.com Website: www.eleint.co.uk
ページ2に含まれる内容の要旨
International TABLE OF CONTENTS I. GENERAL INFORMATION................................................................................................ 1 II. RELATED USER DOCUMENTATION............................................................................. 1 III. CALIBRATION................................................................................................................. 1 IV. OPERATION...........................................................................
ページ3に含まれる内容の要旨
International VACUUM PYCNOMETER MODEL 45-9300 (AP-132) I. GENERAL INFORMATION The 45-9300 (AP-132) Vacuum Pycnometer is designed to be used in determining the theoretical maximum specific gravity of bituminous paving mixtures, using the procedure described in ASTM standard test method D 2041 for type F containers. This large capacity (approximately 10 liters) unit can easily handle samples up to 13.2 lbs (6 kg), minimizing segregation effects. Constructed of lightw
ページ4に含まれる内容の要旨
International IV. OPERATION A. Sample Preparation Follow the procedure outlined in ASTM D 2041 for separating the sample particles, oven-drying the sample, and cooling it to room temperature. B. Operation 1. Weigh the empty pycnometer without the lower water inlet hose and without the gauge and hose assembly on the top. Designate this weight as "B." 2. Place the prepared sample in the pycnometer and weigh the sample and pycnometer together, designating this weight as "C."
ページ5に含まれる内容の要旨
International 8. Open the water inlet valve to fill the pycnometer to the very top without leaving any air space. Be careful not to turn off the vacuum. Close the water inlet valve. 9. Using the quick-disconnect, detach the gauge-hose assembly from the top of the pycnometer. 10. If there is any air trapped in the pycnometer, reconnect the aspirator and vacuum gauge (using the quick-connect) and let more water in. 11. Close the water inlet valve. 12. Disconnect the water inlet
ページ6に含まれる内容の要旨
International V. CALCULATIONS If the test temperature is within +3° or -5°F of 77°F (+ 1.7° or - 2.8°C of 25°C), the following equation can be used to determine the maximum specific gravity of the sample, with an error of 0.001 points or less due to thermal effects: Where: A = Mass of the dry sample in air (grams) D = Mass of container when filled with water at 77°F (25° C) (grams) E = Mass of container filled with both water and sample at 77°F (25°C) (grams) Note:
ページ7に含まれる内容の要旨
International A. Pycnometer Not Getting Vacuum 1. Check 'O' ring to make sure it is clean and in good shape. If not, contact ELE/Soiltest for a replacement 'O' ring. 2. Check to make sure the water inlet valve is on tight. 3. Tighten the stopcock nut (at bottom of stopcock) of the water inlet valve. B. Air Bubbles Entering from the Water Inlet Valve 1. Check to make sure the water inlet valve is on tight. 2. Tighten the stopcock nut (at bottom of stopcock) of the water i
ページ8に含まれる内容の要旨
International 45-9301 (AP-131) Filtering Kit. Functions as water trap for vacuum pump. 45-9320 (AP-134) Manometer VIII. SPECIFICATIONS Capacity Approximately 10 liters; 6 kg (13.2 lbs) sample weight Construction Lightweight polycarbonate, with transparent upper half Gauge 2" (50.8 mm) diameter; 30" x 0.5 " Hg Aspirator Plastic; 3/8" pipe thread (included) Connections Water inlet valve; quick disconnect for vacuum gauge and hose Dimensions 10-3/4" (273 mm) o.
ページ9に含まれる内容の要旨
International FIGURE 1 . 7 Distributor: ELE International, Ltd. ELE International, Inc. Chartmoor Road Chartwell Business PO Box 608 Park Loveland, CO 80539 Leighton Buzzard, Bedfordshire USA LU7 4WG England Phone: (970) 663-9780 Phone: + (0) 1525 249200 Fax: (970) 663-9781 Fax: + (0) 1525 249249 E-Mail: soiltest@eleusa.com E-Mail: ele@eleint.co.uk Website: www.eleusa.com Website: www.eleint.co.uk
ページ10に含まれる内容の要旨
International GRAPHS 8 Distributor: ELE International, Ltd. ELE International, Inc. Chartmoor Road Chartwell Business PO Box 608 Park Loveland, CO 80539 Leighton Buzzard, Bedfordshire USA LU7 4WG England Phone: (970) 663-9780 Phone: + (0) 1525 249200 Fax: (970) 663-9781 Fax: + (0) 1525 249249 E-Mail: soiltest@eleusa.com E-Mail: ele@eleint.co.uk Website: www.eleusa.com Website: www.eleint.co.uk