ページ1に含まれる内容の要旨
User’s Guide
Shop online at
omega.com
e-mail: info@omega.com
For latest product manuals:
omegamanual.info
D DP P6 63 30 00 00 0A A-I I
D DP P6 63 30 00 00 0B B-I I
Digital P Panel M Meter
LP0694A
ページ2に含まれる内容の要旨
® OMEGAnet Online Service Internet e-mail omega.com info@omega.com Servicing North America: U.S.A.: One Omega Drive, P.O. Box 4047 ISO 9001 Certified Stamford, CT 06907-0047 TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5A1, Canada TEL: (514) 856-6928 FAX: (514) 856-6886 e-mail: info@omega.ca For immediate technical or application assistance: ® U.S.A. and Canada: Sales Service: 1-800-826-6342/1-800-TC-OMEGA ® Customer Service: 1-800-622-2378
ページ3に含まれる内容の要旨
FOUR SELECTABLE D.C. RANGES 200 µ A, 2 mA, 20 mA, 200 mA MINIMUM AND MAXIMUM DISPLAY CAPTURE LCD, REFLECTIVE OR RED/GREEN LED BACKLIGHTING 0.48" (12.2 mm) HIGH DIGITS OPTIONAL SETPOINT OUTPUT MODULES U R C L USUS LISTED LISTED IND. . CONT EQ. OPTIONAL SERIAL COMMUNICATIONS MODULES (RS232 or RS485) 51EB OPERATES FROM 9 TO 28 VDC POWER SOURCE FRONT PANEL OR SOFTWARE PROGRAMMABLE DISPLAY COLOR CHANGE CAPABILITY AT SETPOINT OUTPUT NEMA 4X/IP65 SEALED FRONT BEZEL GENERAL DESCRIPTION The DP6
ページ4に含まれる内容の要旨
GENERAL METER SPECIFICATIONS 1. DISPLAY: 5 digit LCD 0.48" (12.2 mm) high digits Vibration According to IEC 68-2-6: 5 to 500 Hz, in X, Y, Z direction for 1.5 hours, 5 g’s. DP63000A-I: Reflective LCD with full viewing angle DP63000B-I: Transmissive LCD with selectable red or green LED backlight, Shock According to IEC 68-2-27: Operational 30 g, 11 msec in 3 directions Altitude: Up to 2000 meters viewing angle optimized. Display color change capability with output state when using an output module
ページ5に含まれる内容の要旨
OPTIONAL PLUG-IN CARDS ADDING OPTION CARDS The DP63000 meters can be fitted with optional output cards and/or serial DUAL SINKING OUTPUT CARD communications cards. The details for the plug-in cards can be reviewed in the Type: Non-isolated switched DC, N Channel open drain MOSFET specification section below. The plug-in cards, that are sold separately, can be Current Rating: 100 mA max. installed initially or at a later date. V : 0.7 V @ 100 mA DS ON V : 30 VDC WARNING: Disconnect all power to
ページ6に含まれる内容の要旨
3.0 INSTALLING PLUG-IN CARDS The Plug-in cards are separately purchased option cards that perform specific CAUTION: The Plug-in cards and main circuit board contain static functions. The cards plug into the main circuit board of the meter sensitive components. Before handling the cards, discharge static charges from your body by touching a grounded bare metal object. Ideally, handle the cards at a static controlled clean workstation. Also, only handle the cards by the edges. Range Jumpers Dirt,
ページ7に含まれる内容の要旨
4.3 INPUT WIRING Current Signal CAUTION: Power input common is NOT isolated from user input (self powered) common. In order to preserve the safety of the meter application, the power input common must be suitably isolated from hazardous live earth referenced voltage; or input common must be at protective earth ground potential. If not, hazardous voltage may be present at the User Inputs and User Input Common terminals. Appropriate considerations must then be given to the potential of the user i
ページ8に含まれる内容の要旨
5.0 REVIEWING THE FRONT BUTTONS AND DISPLAY BUTTON DISPLAY MODE OPERATION ENTERING PROGRAM MODE PROGRAMMING MODE OPERATION SEL Index display through enabled values Press and hold for 2 seconds to activate Store selected parameter and index to next parameter Advances through the program menu RST Resets values (MIN/MAX) or outputs Increments selected parameter value or selection OPERATING MODE DISPLAY DESIGNATORS “1” - To the right of the display indicates setpoint 1 output activated. MAX - Maximu
ページ9に含まれる内容の要旨
6.1 MODULE 1 - SIGNAL INPUT PARAMETERS (1-INP) PARAMETER MENU SCALING STYLE INPUT RANGE RANGE RANGE SELECTION SELECTION StYLE KEy APLY rAN6E RESOLUTION RESOLUTION 200.00 µ A 20.000 mA 200uA 0.02A KEy 0.02A 2.0000 mA 200.00 mA 0.002A 0.2A If Input Values and corresponding Display Values are known, the Key-in Select the input range that corresponds to the external signal. This selection (KEY) scaling style can be used. This allows scaling without the presence or should be high enough to avoid
ページ10に含まれる内容の要旨
DISPLAY MODE DESCRIPTION USER INPUT FUNCTION Serial transmit of the active parameters Print Request selected in the Print Options menu USrIN Print (Module 5). NO Same as Print Request followed by a Print and Reset P-r5t momentary reset of the assigned value(s). DISPLAY MODE DESCRIPTION Setpoint 1 Reset Resets setpoint 1 output. rSt-1 No Function User Input disabled. NO Setpoint 2 Reset Resets setpoint 2 output. rSt-2 See Programming Mode Access chart Program Mode Lock-out P-Loc (Module 3). Set
ページ11に含まれる内容の要旨
6.3 MODULE 3 - DISPLAY AND FRONT PANEL BUTTON PARAMETERS (3-dSP) PARAMETER MENU DISPLAY UPDATE TIME DISPLAY COLOR (BACKLIGHT UNIT ONLY) dSP-t seconds 0.5 1 2 COLOr 1 rEd 6rn This parameter sets the display update time in seconds. rEd Enter the desired display color, red or green. This parameter is active for backlight units only. FRONT PANEL DISPLAY SELECT ENABLE (SEL) SEL yES NO yES DISPLAY INTENSITY LEVEL (BACKLIGHT UNIT ONLY) The yES selection allows the SELbutton to toggle through the
ページ12に含まれる内容の要旨
6.4 MODULE 4 - SETPOINT OUTPUT PARAMETERS (4-SPt) PARAMETER MENU The Setpoint Output Parameters are only active when an optional output SETPOINT VALUE module is installed in the meter. SPt-n -19999 to 99999 10000 SETPOINT SELECT Enter the desired setpoint value. The decimal point position for the setpoint and hysteresis values follow the selection set in Module 1. NO SP-1 SP-2 SPSEL NO HYSTERESIS VALUE Enter the setpoint (output) to be programmed. The n in the following parameters will ref
ページ13に含まれる内容の要旨
button or user input manual reset, serial reset command or meter power OUTPUT RESET WITH DISPLAY RESET cycle. When the user input or RST button is activated (momentary action), the corresponding “on” output is reset immediately and remains off until the rEn-n NO YES trigger point is crossed again. (Previously latched alarms will be off if power up Display Value is lower than setpoint value.) YES This parameter enables the RST button or user input to reset the output when L-dLY = Latch with del
ページ14に含まれる内容の要旨
PROGRAMMING QUICK OVERVIEW Press and hold SEL button to enter Programming Mode.
ページ15に含まれる内容の要旨
WARRANTY/DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product. If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department w
ページ16に含まれる内容の要旨
Where Do I Find Everything I Need for Process Measurement and Control? OMEGA…Of Course! Shop online at omega.com TEMPERATURE ] Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies ] Wire: Thermocouple, RTD & Thermistor ] Calibrators & Ice Point References ] Recorders, Controllers & Process Monitors ] Infrared Pyrometers PRESSURE, STRAIN AND FORCE ] Transducers & Strain Gages ] Load Cells & Pressure Gages ] Displacement Transducers ] Instrumentation & Accessories FLOW/LEVEL ] Ro