Introduction Intended Use Intended Use ® For In Vitro Diagnostic use only. The VITROS ECi/ECiQ Immunodiagnostic System perform random access, batch, and STAT immunodiagnostic assays on human fluid specimens. The system uses chemiluminescence detection technology to provide accurate and reliable results for heterogeneous assays. All reactions for a single quantitative, semi-quantitative, and qualitative measurement take place within a coated well. Note: The availability of the ECi/ECiQ Immu
ページ3に含まれる内容の要旨
Introduction Installation and Site Specifications Site Drawing The site drawing in figure 1-1 provides the physical dimensions of the system. The system weighs approximately 800 pounds (366 kilograms). N M L D C H F A E B K G I J Figure 1-1. Physical Dimensions The system and printer dimensions are described below: Reference Dimension System Dimensions: A 111.8 cm (44 in.) B 162.6 cm (64 in.) C 130.2 cm (51.25 in.) D 179.1 cm (70.5 in.) E 73.7 cm (29 in.) F 61 cm (24 in.) G 59.7 cm (23.5 in.) -
ページ4に含まれる内容の要旨
Introduction Installation and Site Specifications Reference Dimension UPS Dimensions: I 49.4 cm (19.4 in.) J 43.2 cm (17.0 in.) K 8.9 cm (3.5 in.) Printer Stand for Epson LQ/Model 570e and Epson LX/Model 300* Printer Dimensions: L 73.2 cm (28.8 in.) M 61 cm (24 in.) N 87.1 cm (34.3 in.) Printer Stand with Texas Instruments Omni 800/Model 830e Printer Dimensions: L 73.2 cm (28.8 in.) M 61 cm (24 in.) N 91.4 cm (36 in.) *Standard printer for the system. Printers are ordered in accordance with regi
ページ5に含まれる内容の要旨
Introduction Installation and Site Specifications e-Connectivity™ e-Connectivity™ requires a Virtual Private Network (VPN) and Modem. The VPN and Modem require two electrical outlets and non-dedicated power lines can be used. They can also be plugged into a single power strip which can share the same receptacle as the ECi/ECiQ System. Power cords are supplied with the VPN and Modem. Line Voltage and Frequency for VPN and Modem used world wide: VPN Modem Line Voltage: 100 - 240 volt AC Line V
ページ6に含まれる内容の要旨
Introduction Installation and Site Specifications Uninterruptible Power Supply If you experience frequent power fluctuations, you may want to use an uninterruptible power supply. Contact your technical support representative for more information. Entrance to Room The entrance door opening should be a minimum of 76.2 cm (30 in.) wide. Telephone Recommendations You may also find it convenient to have a telephone near the system to communicate with technical support personnel during troubleshooti
ページ7に含まれる内容の要旨
Introduction Installation and Site Specifications Environmental Specifications The environmental limits for normal operation of the system are defined below and shown in figure 1-2. BTU Output: 4100 BTUs per hour Operating Temperature: 15–30°C (59–86°F) Site Relative Humidity: 15–75% RH noncondensing Altitude: -0.1524/2.439 km (-500/8000 ft.) 30.0°C (86°F) 26.0°C (78°F) Temperature 15.0°C (59°F) 15% 50% 75% Relative Humidity Figure 1-2. Environmental Requirements System Relocation The s