SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others. Each safety message is associated with a safety alert symbol. These symbols are found in the manual and inside the instrument. The definition of these symbols is described below: General Warning/Caution: Refer to the instructions f
ページ3に含まれる内容の要旨
TABLE OF CONTENTS SAFETY MESSAGES....................................................................................................................................II TABLE OF CONTENTS ...............................................................................................................................III FIGURES ........................................................................................................................................................V TABLES...........
ページ4に含まれる内容の要旨
5.1 CALIBRATION ...........................................................................................................................................47 5.2 ADJUSTING UV LAMP AND DETECTOR ......................................................................................................47 5.3 CHANGING THE ANALOG OUTPUT RANGE..................................................................................................49 5.3.1 Voltage output..........................................
ページ5に含まれる内容の要旨
FIGURES FIGURE 1.1 - REAR PANEL PNEUMATIC CONNECTIONS............................................................13 FIGURE 1.2 – NEMA 4X ENCLOSURE DIMENSIONS.......................................................................16 FIGURE 1.3 – NEMA 4X AC LINE POWER CONNECTION .............................................................17 FIGURE 1.4 – NEMA 4X FRONT PANEL..............................................................................................18 FIGURE 1.5 – NEMA 4X PNEU
ページ6に含まれる内容の要旨
TABLES TABLE 1.1 DIGITAL OUTPUTS..............................................................................................................21 TABLE 1.2 FINAL TEST AND CALIBRATION VALUES ...................................................................28 TABLE 2.1 SYSTEM MODES...................................................................................................................33 TABLE 2.2 TEST MEASUREMENTS....................................................................
ページ7に含まれる内容の要旨
1.0 INTRODUCTION 1.1 Preface Teledyne API is pleased that you have purchased the Model 450. We offer a full one year warranty (see Section 1.2) and we at Teledyne API will be pleased to provide you with any support required so that you may utilize our equipment to the fullest extent. The Teledyne API Model 450 keyboard/operator interface with its "talking keys" makes the Teledyne API a very user-friendly system. We hope you will not experience any problems with the Teledyne API Model
ページ8に含まれる内容の要旨
1.2 WARRANTY POLICY ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API) 02024b Prior to shipment, Teledyne API equipment is thoroughly inspected and tested. Should equipment failure occur, Teledyne API assures its customers that prompt service and support will be available. COVERAGE After the warranty period and throughout the equipment lifetime, Teledyne API stands ready to provide on-site or in-plant service at reasonable rates similar to those of other manufacturers in the industry
ページ9に含まれる内容の要旨
TERMS AND CONDITIONS All units or components returned to Teledyne API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid. 1.3 Principle of Operation The detection of ozone molecules is based on absorption of 254 nm UV light due to an internal electronic resonance of the O molecule. The Model 450 uses a mercury lamp 3 constructed so that a large majority of the l
ページ10に含まれる内容の要旨
The intensity of light is converted into a voltage by the detector/preamp module. The voltage is converted into a number by a voltage-to-frequency (V/F) converter capable of 80,000 count resolution. The digitized signal, along with the other variables, are used by the CPU to compute the concentration using the above formula. Every 6 seconds the M450 completes a measurement cycle consisting of a 2 second wait period for the sample tube to flush, followed by a 1 second measurement of the
ページ11に含まれる内容の要旨
1.4 Specifications Ranges User selectable to any full scale range from 1ppm to 1000 ppm 3 3 Measurement Units ppbv, ppmv, ppmw, pphm, µg/m , mg/ m Zero Noise < .0015 ppm (rms) Span Noise < .5% of reading (rms) (above 0.1 ppm) Lower Detectable Limit < .003 ppm (rms) Linearity Better than 1% full scale Response Time (95%) <30 sec Minimum Stream Duration 1.0 min (Cycle Time)* Sample Flow Rate 1-2.5 L/min o Temperature Range 5-45 C Humidity Range 10-90% RH, Non-Condensing Dimensions
ページ12に含まれる内容の要旨
1.5 Installation (19” Rack Mount Enclosure) The Model 450 is shipped with the following standard equipment: 1. Power cord. 2. Instruction manual. CAUTION To avoid personal injury, always use two persons to lift and carry the Model 450. Upon receiving the Model 450 please do the following: 1. Verify no apparent shipping damage. (If damage has occurred please advise shipper first, then Teledyne API.) 2. Remove all red colored shipping screws from the inside and bottom o
ページ13に含まれる内容の要旨
CAUTION Verify that the instrument is set up for proper line voltage and frequency. 6. Turn on the M450 by switching the switch on the lower right corner of the front panel. The front panel display should light with a sequence of messages, -API - M450 - software version number, then a normal display as shown in Figure 2.2. 7. Allow about 20 minutes for the UV lamp temperature to come up to it’s setpoint then press the TST> button on the front keyboard to scroll through the TEST value
ページ14に含まれる内容の要旨
1.6 Installation (NEMA 4X, Wall Mount Enclosure) The Model 450 is shipped with the following standard equipment: 1. Instruction manual. CAUTION To avoid personal injury, always use two persons to lift and carry the Model 450. Upon receiving the Model 450 please do the following: 1. Verify no apparent shipping damage. (If damage has occurred please advise shipper first, then Teledyne API.) 2. Mount the NEMA 4X enclosure to a vertical surface using four 5/16” bolts. Mountin
ページ15に含まれる内容の要旨
10. Allow about 20 minutes for the UV lamp temperature to come up to it’s setpoint then press the TST> button on the front keyboard to scroll through the TEST values. Compare these values to those noted during the final factory checkout listed in Table 1.2. The values observed should closely match the Table 1.2 values. P/N 02395D1 Teledyne API Model 450 O Monitor Instruction Manual - Page 15 3
ページ16に含まれる内容の要旨
FIGURE 1.2 – NEMA 4X ENCLOSURE DIMENSIONS P/N 02395D1 Teledyne API Model 450 O Monitor Instruction Manual - Page 16 3
ページ17に含まれる内容の要旨
FIGURE 1.3 – NEMA 4X AC LINE POWER CONNECTION P/N 02395D1 Teledyne API Model 450 O Monitor Instruction Manual - Page 17 3
ページ18に含まれる内容の要旨
FIGURE 1.4 – NEMA 4X FRONT PANEL P/N 02395D1 Teledyne API Model 450 O Monitor Instruction Manual - Page 18 3
ページ19に含まれる内容の要旨
FIGURE 1.5 – NEMA 4X PNEUMATIC CONNECTIONS P/N 02395D1 Teledyne API Model 450 O Monitor Instruction Manual - Page 19 3
ページ20に含まれる内容の要旨
FIGURE 1.6 - ELECTRICAL SIGNAL CONNECTIONS 1.7 Electrical and pneumatic connections 1.7.1 Analog Output The analog output is a two pin connector on the rear panel (See Figure 1.6). The analog output can be configured for voltage or current output. The standard is a 0-5 volt output. 0-100mV, 0-1V, 0-10V, and 4-20 mA outputs are also available. See Section 5.3 for information on setting other output ranges. 1.7.2 Relay Outputs Three form C relay outputs are provided on the rea