ページ1に含まれる内容の要旨
®
GE GE Protimeter Aquant
Sensing User Instructions
Introduction
The GE Protimeter Aquant BLD5760 moisture detector is used to
assess and monitor the relative moisture level of non-conductive
and porous materials: such as plaster, masonry, concrete and
fiberglass (GRP). It is an ideal instrument for making rapid surveys
of solid walls and floors and to pinpoint areas of concern in tiled
shower cubicles, GRP boat hulls, motor homes and caravans etc.
The Aquant BLD5760 has a digital display t
ページ2に含まれる内容の要旨
4. Place the instrument against the surface of the wall, floor or element at the chosen point of measurement as shown. Ensure that your forefinger and thumb do not extend beyond the top of the black plastic grip band and wrap all fingers around the side of the instrument. Note: The Aquant should be held at a nominal 25° to the surface so that both top edge of the instrument and the sensor bulge are in contact with the surface. 5. Read the relative moisture level value from the display and n
ページ3に含まれる内容の要旨
2. Scroll across the display and record the following information: a. part number (“bLd5760”) b. firmware date in the form yy-mm-dd (e.g. “04-05-20”) c. calibration code (e.g. E1-1). Note: Once the scrolling is complete, display shows “0 = 0” meaning that option 0 is set to 0. The and buttons may now be used to change the instrument options and settings, respectively, as detailed in Table 1 below. TABLE 1. User Set-Up Options When And Setting Option No. Is: Number Is: Aquant Set-Up Is:
ページ4に含まれる内容の要旨
3. Press once again to select auto switch off options (2), then press until the display shows 2 = 3. 4. Press to save settings, exit Set-up mode and return to operational modes. 6Error Messages When zeroing if the error message E32 appears on the instrument display, signifying that the instrument reading is beyond the zeroing range of 130. 7 Care and Maintenance When the Aquant is not being used, then do the following: • Store the Aquant in the carry case (provided with the kit) in a stabl