Résumé du contenu de la page N° 1
Contents
HP E1351A/53A FET Multiplexer User’s Manual
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Declaration of Conformity . . . . . . . .
Résumé du contenu de la page N° 2
Measuring Temperature Using Thermocouples (HP E1353A Module only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Setup for Measuring Thermocouple Temperature using an External Multimeter . . 43 Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Chapter 4.Understanding the HP E1351A/53A FET Multiplexer Modules . . . . . . . . . 45 Using This Chapter . . . . .
Résumé du contenu de la page N° 3
:CPON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 :CTYPe? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 :ERRor? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 TRIGger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Résumé du contenu de la page N° 4
Notes 4 HP E1351A/53A FET Multiplexer User’s Manual Contents
Résumé du contenu de la page N° 5
Certification Hewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the factory. Hewlett- Packard further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Tech- nology (formerly National Bureau of Standards), to the extent allowed by that organization’s calibration facility, and to the calibration facilities of other International Standards Organization members. Warranty
Résumé du contenu de la page N° 6
Documentation History All Editions and Updates of this manual and their creation date are listed below. The first Edition of the manual is Edition 1. The Edi- tion number increments by 1 whenever the manual is revised. Updates, which are issued between Editions, contain replacement pages to correct or add additional information to the current Edition of the manual. Whenever a new Edition is created, it will contain all of the Update information for the previous Edition. Each new Edition or U
Résumé du contenu de la page N° 7
Declaration of Conformity according to ISO/IEC Guide 22 and EN 45014 Manufacturer’s Name: Hewlett-Packard Company Loveland Manufacturing Center Manufacturer’s Address: 815 14th Street S.W. Loveland, Colorado 80537 declares, that the product: Product Name: 16-Channel FET Multiplexer Modules Model Number: E1351A, E1353A Product Options: All conforms to the following Product Specifications: Safety: IEC 1010-1 (1990) Incl. Amend 1 (1992)/EN61010-1 (1993) CSA C22.2 #1010.1 (1992) UL 1244 EMC: CISPR
Résumé du contenu de la page N° 8
Notes 8 HP E1351A, E1353A 16-Channel FET Multiplexer Modules User’s Manual
Résumé du contenu de la page N° 9
Please fold and tape for mailing Reader Comment Sheet HP E1351A/53A 16-Channel FET Multiplexer Module User’s Manual Edition 4 You can help us improve our manuals by sharing your comments and suggestions. In appreciation of your time, we will enter you in a quarterly drawing for a Hewlett-Packard Palmtop Personal Computer (U.S. government employees cannot participate in the drawing). Your Name City, State/Province Company Name Country Job Title Zip/Postal Code Address Telephone Number with Area
Résumé du contenu de la page N° 10
10 HP E1351A/53A 16-Channel FET Multiplexer Module User’s Manual
Résumé du contenu de la page N° 11
Chapter 1 Getting Started with the HP E1351A/53A Using This Chapter This chapter describes the HP E1351A 16-Channel FET and the HP E1353A 16-Channel Thermocouple FET Multiplexer Modules, and shows how to program the modules using SCPI (Standard Commands for Programmable Instruments) commands. This chapter contains the following sections: • FET Multiplexer Module Description . . . . . . . . . . . . . . . . . . Page 11 • Multimeter Connection Points. . . . . . . . . . . . . . . . . . . . . . . .
Résumé du contenu de la page N° 12
For high-speed operation (100 K switches/sec.) the scanning list is automatically downloaded into RAM on the multiplexer module. Triggering for channel advance is from the two handshake lines on the digital bus. The scanning operation does not require any intervention from the mainframe CPU. This only applies for switchboxes or scanning voltmeter configurations that have all FET multiplexer modules. For a downloaded scan list in switchboxes, the trigger source must be TRIG:SOUR DBUS or TRIG:S
Résumé du contenu de la page N° 13
Figure 1-1. FET Multiplexer Module Block Diagram Chapter 1 Getting Started with the HP E1351A/53A 13
Résumé du contenu de la page N° 14
Switchbox and Scanning Voltmeter Configurations A VXIbus instrument is a module or group of modules which perform a specified function. For the HP E1300/01 mainframe, the first module in an instrument must have a logical address which is evenly divisible by 8 (16, 24, 112), and the rest of the modules in the instrument are numbered consecutively. The instrument’s secondary address is the whole number equal to the logical address divided by 8 (for a logical address of 16, 17, 18,…23, the second
Résumé du contenu de la page N° 15
120 TRIGGER 70903 !Trigger the switchbox to advance the channel list 130 NEXT I !Increment count 140 END Scanning Voltmeter When the multiplexer(s) is combined with a multimeter to form a single instrument, they become a virtual instrument, a scanning voltmeter. The multiplexer(s) and the multimeter have the same secondary address. The multimeter automatically configures the multiplexer, so the SCAN:MODE, SCAN:PORT and TRIG:SOUR commands are not required. Channel advance is from the digital
Résumé du contenu de la page N° 16
Figure 1-2. Custom Cable for Digital Bus Triggering Programming Language The examples in this manual use the Standard Commands for Programmable Instruments (SCPI) commands, Hewlett-Packard BASIC and an HP 9000 Series 200/300 computer over the HP-IB. Chapter 5 contains information on SCPI command format. Appendix B contains details on the registers for register-based programming. Initial Operation Use the following program to verify initial multiplexer operation by closing a channel and queryin
Résumé du contenu de la page N° 17
Chapter 2 Configuring the HP E1351A/53A FET Multiplexer Modules Using This Chapter This chapter shows how to configure the 16-Channel FET Multiplexer Modules, how to connect external wiring and how to connect multimeters. This chapter contains the following sections: • Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 17 • Multiplexer Card Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . Page 18 • Selecting the Interrupt Line Number . . . . . . . .
Résumé du contenu de la page N° 18
Multiplexer Card Numbers HP plug-in modules installed in an HP VXIbus mainframe are treated as independent instruments each having a unique secondary HP-IB address. An instrument may be composed of a single plug-in module or multiple plug-in modules. The card number identifies the module within a switchbox or scanning voltmeter configuration. The multiplexer module with the lowest logical address is always card number 01. The multiplexer module with the next successive address is 02, and so o
Résumé du contenu de la page N° 19
Figure 2-3. Card Numbers for Scanning Voltmeter The logical addresses noted in Figures 2-1, 2-2, and 2-3 apply to modules installed in an HP 75000 Series B Mainframe (HP Model Number E1300A/E1301A) or in a mainframe with an HP E1405/1406 Command Module. See the HP 75000 Series B Installation and Getting Started Guide or the appropriate HP Command Module Manual for more information on switchboxes and scanning voltmeter configurations, and logical addressing. For uses in other systems or mainfra
Résumé du contenu de la page N° 20
Selecting the Interrupt Line Number The multiplexer module generates interrupts at the end of a downloaded scan list. The interrupt line number switch determines which backplane IRQ line will be used. Different mainframes and resource managers recognize different backplane IRQ lines. The interrupt line number must be set to the line number your system is programmed to recognize. Refer to your mainframe manual. Refer to Figure 2-5 to change the interrupt lines. Valid line numbers are from 1