Résumé du contenu de la page N° 1
QUICKSTART GUIDE
MODEL 3000TA-XL-EU
Trace Oxygen Analyzer
Teledyne Analytical Instruments
Résumé du contenu de la page N° 2
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments
Résumé du contenu de la page N° 3
Model 3000 TA-XL-EU Quickstart Guide GETTING STARTED This Quickstart Guide is designed to get you set up and operating your Teledyne Analytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss. This Quickstart Guide should be used in conjunction with the Instruction Manual that shipped with your instrument. Only necessary features to get you operational are discussed in this guide. Many of the advanced features of th
Résumé du contenu de la page N° 4
Quickstart Guide Trace Oxygen Analyzer Function Keys: Six touch-sensitive membrane switches are used to change the specific function performed by the analyzer: • Analyze Perform analysis for oxygen content of a sample gas. • System Perform system-related tasks (described in detail in chapter 4, Operation.). • Span Span calibrate the analyzer. • Zero Zero calibrate the analyzer. • Alarms Set the alarm setpoints and attributes. • Range Set up the 3 user definable ranges fo
Résumé du contenu de la page N° 5
Model 3000 TA-XL-EU Quickstart Guide Rear Panel (Equipment Interface) At the rear panel you make the gas and electrical connections to the instrument. • Power Connection: Universal AC power source. • Gas Inlet and Outlet: One inlet and one exhaust out. • RS-232 Port : Serial digital concentration signal output and control input. • Equipment Interface Connector: Interconnections for analyzer functions including: • Analog Outputs: 0–1 VDC oxygen concentration plus 0-1 VDC r
Résumé du contenu de la page N° 6
Quickstart Guide Trace Oxygen Analyzer SETUP AND INSTALL Mount the Analyzer The Model 3000TA -EU is for indoor use in a general purpose area. It is NOT for hazardous environments of any type. It is designed for flush panel mounting. Use the four mounting holes— one in each corner of the rigid frame to panel mount the instrument. Make Gas Connections Note: Do not remove plastic caps on VCR fittings, unless you are ready to make connections. The analyzer is equipped
Résumé du contenu de la page N° 7
Model 3000 TA-XL-EU Quickstart Guide Electrical Connections CAUTION: READ ALL SAFETY INFORMATION REGARDING ELECTRICAL HAZARDS ASSOCIATIED WITH THIS INSTRUMENT BEFORE MAKING ELECTRICAL CONNECTIONS. CAUTION: POWER IS APPLIED TO THE INSTRUMENT'S CIRCUITRY AS LONG AS THE INSTRUMENT IS CONNECTED TO THE POWER SOURCE. THE RED SWITCH ON THE FRONT PANEL IS FOR SWITCHING POWER ON OR OFF TO THE DISPLAYS AND OUTPUTS ONLY. • Connect primary input power: Insert the power cord in
Résumé du contenu de la page N° 8
Quickstart Guide Trace Oxygen Analyzer • Install 50-Pin Equipment Interface Connector Connect the 50 pin connector to the mating connector on the rear panel. • Connect the 9-pin D connector to the RS-232 serial port. See Section 3.3.2.3 in the Instruction Manual for information on setting up serial communication for this instrument. Install the Micro-fuel Cell The Model 3000TA-XL-EU can be equipped with either the standard B-2CXL Micro-fuel Cell or as an option, Teledyne’
Résumé du contenu de la page N° 9
Model 3000 TA-XL-EU Quickstart Guide 1. Place the cell on the holder with the screen side facing down. Note: There is a small location hole drilled in the holder. This hole mates with a guide pin on the bottom rear of the cell block. The hole in the cell block holder must align with the guide pin on the cell block. 2. Remove power from the instrument. 3. Remove the old sensor (if installed). Discard in accordance with all Federal, State, Local or other regulations which may apply. 4. Pur
Résumé du contenu de la page N° 10
Quickstart Guide Trace Oxygen Analyzer BASIC OPERATION Once the analyzer has been installed, it can be configured for your application. Prior to using the instrument you must: • Calibrate the instrument • Define the analysis ranges and choose whether to use autoranging or a fixed range. • Set alarm setpoints and modes of alarm operation. Additional configuration parameters are available to tailor the instrument to your particular application. Refer to the Instruction Manual for speci
Résumé du contenu de la page N° 11
Model 3000 TA-XL-EU Quickstart Guide Available Functions The six function buttons are used to control the analyzer. Some functions have submenus that allow additional choices as shown below. Teledyne Analytical Instruments 9
Résumé du contenu de la page N° 12
Quickstart Guide Trace Oxygen Analyzer Any function can be selected at any time by pressing the appropriate button (unless password restrictions apply). Password Protection If a password is assigned, then setting the following system parameters can be done only after the password is entered: span and zero settings, alarm setpoints, analysis range definitions, switching between autoranging and manual override, setting up an auto-cal, and assigning a new password. However, th
Résumé du contenu de la page N° 13
Model 3000 TA-XL-EU Quickstart Guide ◄► arrow keys to scroll back and forth between letters, and the ▲▼ arrow keys to change the letters to the proper password. • Press Enter to enter the password. Change Password? =Yes =No • Press Escape. Calibration of the Analyzer For trace analysis, the analyzer must be purged before calibration. Refer to the Instruction manual for purging procedure. Zero Calibration Make sure an appropriate zero gas is connected to the samp
Résumé du contenu de la page N° 14
Quickstart Guide Trace Oxygen Analyzer The zeroing process will automatically conclude when the output is within the acceptable range for a good zero. Then the analyzer automatically returns to the Analyze mode. Manual Mode Zeroing • Press Zero to enter the Zero function. • Use the ▲▼ keys to toggle between AUTO and MAN zero settling. Stop when MAN appears, blinking, on the display. Zero: Settling: Man To Begin • Press Enter to begin the zero calibration. #### ppm
Résumé du contenu de la page N° 15
Model 3000 TA-XL-EU Quickstart Guide #### ppm Span Slope=#### ppm/s The instrument automatically returns to the Analyze mode at the conclusion of the auto span cycle.. Manual Mode Spanning: • Press Span to start the Span function. Span: Settling:MAN For Next • Use the ▲▼ keys to toggle between AUTO and MAN span settling. Stop when MAN appears, blinking, on the display. Press Enter to move to the next screen. Calib. Holding time Cal hold: 5 mi
Résumé du contenu de la page N° 16
Quickstart Guide Trace Oxygen Analyzer • Press the Alarm button on the front panel to enter the Alarm function. Make sure that AL–1 is blinking. AL—1 AL—2 Choose Alarm • Using the ◄► arrow keys to move the blinking over to AL–1. Then press Enter. AL—1 1000 ppm HI Dft—N Fs—N Ltch—N Five parameters can be changed on this screen: • Value of the alarm setpoint, AL–1 #### ppm (oxygen); value can be set from 0.00 to 10,000.00 ppm • Out-of-range direction, HI or LO • Defe
Résumé du contenu de la page N° 17
Model 3000 TA-XL-EU Quickstart Guide • Use the ◄► arrow keys to blink the range to be set: low (L), medium (M), or high (H). • Use the ▲▼ arrow keys to enter the upper value of the range (all ranges begin at 0 ppm). Repeat for each range you want to set. • Press Enter to accept the values and return to analyze mode. Note: The ranges must be increasing from low to high, for example, if range 1 is set as 0–100 ppm and range 2 is set as 0–1,000 ppm, range 3 cannot be set as 0– 500 ppm sin