S8 Lightweight™ II
POSITIVE AIRWAY PRESSURE DEVICE
User Guide
English
ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia • ResMed Corp (US
Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA • ResMed (UK) Ltd (EU Authorized
Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK • ResMed Offices Australia, Austria, Belgium,
Brazil, China, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Malaysia, Mexico, Neth
Résumé du contenu de la page N° 2
Contents Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Contraindications 1 Adverse Effects 1 The S8 Lightweight II System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Masks 2 Humidifiers 2 ResScan Data Card 2 Using the S8 Lightweight II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Set Up the S8 Lightweight II 3 Start Treatment 4 Stop Treatment 4 How to Use the Control Panel 4 How to
Résumé du contenu de la page N° 3
Introduction The S8 Lightweight II continuous positive airway pressure (CPAP) system is indicated for the treatment of obstructive sleep apnoea (OSA) in patients > 30 kg. The S8 Lightweight II CPAP system is intended for home and hospital use. Contraindications The S8 Lightweight II is not a life support device and may stop operating during power failure or certain fault conditions. It should not be used by patients who are dependent on continuous therapy. Positive airway pressure therapy ma
Résumé du contenu de la page N° 4
The S8 Lightweight II System Your S8 Lightweight II comprises: • S8 Lightweight II positive airway pressure device • 2 m air tubing • Power cord • Carry bag. Carry handle Data Card slot Screen Keypad Air inlet DC Air outlet AC Optional components include: • ResScan™ Data Card • 3 m air tubing • DC-12 converter. Masks For information on using masks, see your mask manual. Recommended masks are available on www.resmed.com on the Products page under
Résumé du contenu de la page N° 5
Using the S8 Lightweight II WARNING ! • Make sure the power cord and plug are in good condition and the equipment is not damaged. • Only ResMed air tubing should be used with the device. A different type of air tubing may alter the pressure you actually receive, reducing the effectiveness of your treatment. • Blocking the hose and/or air inlet of the device while in operation could lead to overheating of the device. CAUTION ! • Be careful not to place the device where it can be bumped or wh
Résumé du contenu de la page N° 6
Start Treatment LCD screen Keypad Start/Stop key 1 Make sure the power is on. The product name is displayed briefly on the LCD screen, then the standby (Ramp) screen appears. The key and LCD backlights also turn on. 2 Lie down and arrange the air tubing so that it is free to move if you turn in your sleep. CAUTION ! Do not leave long lengths of air tubing around the top of your bed. They could twist around your head or neck while you are sleeping. 3 Alter the ramp time if required (see “How t
Résumé du contenu de la page N° 7
The S8 Lightweight II keypad has the following keys: Key Function Starts or stops treatment. Start/Stop Up Allows you to increase settings options and scroll through the S8 Lightweight II menu. Down Allows you to decrease settings options and scroll through the S8 Lightweight II menu. Performs the function indicated by the guiding text displayed Left above it on the LCD screen. Guiding text includes menu, change, and apply. Performs the function indicated by the guiding text displayed Right
Résumé du contenu de la page N° 8
How to Change the Settings on your S8 Lightweight II You can change settings (such as Ramp Time) on the S8 Lightweight II by using the keypad and LCD. When the S8 Lightweight II is in standby mode you can display a series of screens (a menu) on the LCD screen in order to view and change the settings for a particular function. S8 ESCAPE II PATIENT MENU S8 CLASSIC S8 CLASSIC RAMP 10min Menu EPR LEVEL: 1 Change Exit Only displayed if Exit cl
Résumé du contenu de la page N° 9
1. On the standby (RAMP) screen, press . 2. Press until you see EPR LEVEL, then press (Change). 3. Press or until you see the setting that you require. 4. Press (Apply) to save the setting. How to change the language setting 1. On the standby (Ramp) screen, press (Menu). 2. Press until you see LANG (Language), then press (Change). 3. Press or until you see the language that you require. 4. Press (Apply) to save the new language. Messages on the S8 Lightweight II LCD Your clinicia
Résumé du contenu de la page N° 10
Using the Data Card to Collect data If your clinician needs to review your treatment, they will ask you to use the Data Card to copy data from your S8 Lightweight II, and to return the card to them. The Data Card will be provided in a protective folder. Data that is copied to a Data Card is still stored and available on the S8 Lightweight II. Copy data onto the Data Card 1. Switch on the S8 Lightweight II and wait until you see the standby (Ramp) screen. 2. Hold the Data Card with the arrow
Résumé du contenu de la page N° 11
CAUTION ! • Do not hang the air tubing in direct sunlight as it may harden over time and eventually crack. • Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturising or antibacterial soaps or scented oils to clean the air tubing or the S8 Lightweight II. These solutions may cause hardening and reduce the life of the product. Daily Disconnect the air tubing and hang it in a clean, dry place until next use. Weekly 1. Remove the air tubing from the S8 Lightweight II and t
Résumé du contenu de la page N° 12
Servicing CAUTION ! • Do not attempt to open the S8 Lightweight II case. There are no user serviceable parts inside. • Inspection and repair should only be performed by an authorised agent. Under no circumstances should you attempt to service or repair the S8 Lightweight II yourself. This device is intended to provide safe and reliable operation when operated and maintained in accordance with the instructions provided by ResMed. To ensure that your device continues to provide optimum perfo
Résumé du contenu de la page N° 13
Anything special I should do before putting the mask on? Before wearing your mask, wash your face to remove excess facial oils. This will allow a better fit and prolong the life of the mask cushion. Do not use moisturiser when wearing a mask. My mouth feels dry during treatment—what can I do? You may experience dryness of the nose, mouth, and/or throat during the course of treatment, especially during winter. In many cases, a humidifier may resolve this discomfort. Contact your clinician for
Résumé du contenu de la page N° 14
Insufficient air delivered from S8 Lightweight II Ramp Time is in use. Wait for air pressure to build up or change ramp time. Air filter is dirty. Replace air filter. Air tubing is kinked or Straighten or replace tubing. punctured. Air tubing is not connected Check air tubing. properly. Mask and headgear are not Adjust position of mask and headgear. positioned correctly. Cap(s) are missing from Replace cap(s). access port on mask. Pressure required for See your clinician to adjust the pressure
Résumé du contenu de la page N° 15
Technical Specifications . Operating pressure range 4 to 20 cm H O 2 Dimensions (H x W x D) 112 mm x 164 mm x 145 mm Weight 1.3 kg Power Supply • Input range for S8 Lightweight II with HumidAire 4i:100–240V, 50/60Hz; 110V, 400Hz; 2.5A < 140VA (110W) (maximum power consumption). Instantaneous peak power consumption <340VA. • Refer to the DC-12 converter instructions for DC ratings. Only use the ResMed DC-12 converter for DC input. • Typical power consumption at 20 cm H 0 is 34VA or 20W. 2 Air
Résumé du contenu de la page N° 16
Symbols which appear on the device Attention, consult accompanying documents; Follow instructions for use; Drip proof; Type CF equipment; Dangerous voltage; Class II equipment; Start/Stop; Manufacturer; Environmental information WEEE 2002/96/EC is a European Directive that requires the proper disposal of electrical and electronic equipment. This device should be disposed of separately, not as unsorted municipal waste. To dispose of your device, you should use appropriate collecti
Résumé du contenu de la page N° 17
• In the event of power failure or machine malfunction, remove the mask. • The S8 Lightweight II can be set to deliver pressures up to 20 cm H O. In the 2 unlikely event of certain fault conditions, pressures up to 30 cm H O are 2 possible. • If oxygen is used with this device, the oxygen flow must be turned off when the device is not operating. If the oxygen has been left on, turn off the device, then wait 30 minutes before turning on the device again. Explanation: When the CPAP device is
Résumé du contenu de la page N° 18
Limited Warranty ResMed warrants that your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer. This warranty is not transferable. Product Warranty Period ResMed humidifiers, ResControl™, ResLink™, ResTraxx™ 1 Year ResMed flow generators 2 Years Accessories, mask systems (including mask frame, cushion, 90 Days headgear and tubing). Excludes single-use devices. Note: Some models are not available
Résumé du contenu de la page N° 19
Limited Warranty 17
Résumé du contenu de la page N° 20
18